Enhancing meniscal repair with tough adhesive puncture sealing (TAPS) suture: A proof-of-concept study on bovine cadaveric knees

IF 2 Q2 ORTHOPEDICS Journal of Experimental Orthopaedics Pub Date : 2024-12-30 DOI:10.1002/jeo2.70129
Stephanie Lamer, David Mazy, Marie-Lyne Nault
{"title":"Enhancing meniscal repair with tough adhesive puncture sealing (TAPS) suture: A proof-of-concept study on bovine cadaveric knees","authors":"Stephanie Lamer,&nbsp;David Mazy,&nbsp;Marie-Lyne Nault","doi":"10.1002/jeo2.70129","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Purpose</h3>\n \n <p>The objective was to use cyclic tensile loading to compare the gap formation at suture site of three different suture materials to repair bovine radial meniscal tears: (1) polyglactin sutures, (2) tough adhesive puncture sealing (TAPS) sutures and (3) ultra-high molecular weight polyethylene (UHMWPE) sutures.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>Twelve ex vivo bovine knees were dissected to retrieve the menisci. Complete radial tears were performed on 24 menisci, which were then separated into three groups and repaired using either pristine 2–0 polyglactin sutures, TAPS sutures (2–0 polyglactin sutures coated with adhesive chitosan/alginate hydrogels) or 2-0 UHMWPE sutures with a single stitch. The repaired menisci were clamped onto an Instron machine and underwent cyclic loading between 5 and 25 N at a frequency of 0.16 Hz. Gap formation between the edges of the tear was measured after 500 cycles using an electronic caliper, when the meniscus was still on the Instron without any load applied.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>Mean gap formation was 5.22 mm (±1.70) for the 2–0 polyglactin sutures, 2.48 mm (±0.25) for the TAPS sutures, and 4.85 mm (±1.55) for the 2–0 UHMWPE sutures. The gap was significantly smaller in the TAPS sutures group compared to the two others because of better force dispersion, decreasing tissue damage by suture indentation and potentially leading to better meniscal healing.</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>From a biomechanical standpoint, coated sutures held the edges of radial meniscal tears closer together compared to conventional sutures. This technology has the potential to reduce tissue damage and improve the success rate of meniscal repairs.</p>\n </section>\n \n <section>\n \n <h3> Level of Evidence</h3>\n \n <p>controlled laboratory study.</p>\n </section>\n </div>","PeriodicalId":36909,"journal":{"name":"Journal of Experimental Orthopaedics","volume":"12 1","pages":""},"PeriodicalIF":2.0000,"publicationDate":"2024-12-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11683771/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Experimental Orthopaedics","FirstCategoryId":"1085","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/jeo2.70129","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose

The objective was to use cyclic tensile loading to compare the gap formation at suture site of three different suture materials to repair bovine radial meniscal tears: (1) polyglactin sutures, (2) tough adhesive puncture sealing (TAPS) sutures and (3) ultra-high molecular weight polyethylene (UHMWPE) sutures.

Methods

Twelve ex vivo bovine knees were dissected to retrieve the menisci. Complete radial tears were performed on 24 menisci, which were then separated into three groups and repaired using either pristine 2–0 polyglactin sutures, TAPS sutures (2–0 polyglactin sutures coated with adhesive chitosan/alginate hydrogels) or 2-0 UHMWPE sutures with a single stitch. The repaired menisci were clamped onto an Instron machine and underwent cyclic loading between 5 and 25 N at a frequency of 0.16 Hz. Gap formation between the edges of the tear was measured after 500 cycles using an electronic caliper, when the meniscus was still on the Instron without any load applied.

Results

Mean gap formation was 5.22 mm (±1.70) for the 2–0 polyglactin sutures, 2.48 mm (±0.25) for the TAPS sutures, and 4.85 mm (±1.55) for the 2–0 UHMWPE sutures. The gap was significantly smaller in the TAPS sutures group compared to the two others because of better force dispersion, decreasing tissue damage by suture indentation and potentially leading to better meniscal healing.

Conclusions

From a biomechanical standpoint, coated sutures held the edges of radial meniscal tears closer together compared to conventional sutures. This technology has the potential to reduce tissue damage and improve the success rate of meniscal repairs.

Level of Evidence

controlled laboratory study.

Abstract Image

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
加强半月板修复与坚韧的胶粘剂穿刺密封(TAPS)缝合:牛尸体膝盖的概念验证研究。
目的:利用循环拉伸载荷比较三种不同缝线材料(1)聚乳酸缝线、(2)坚韧胶粘剂穿刺密封(TAPS)缝线和(3)超高分子量聚乙烯(UHMWPE)缝线在修复牛桡骨半月板撕裂时缝线部位的间隙形成情况。方法:解剖12只离体牛膝,取出半月板。对24例半月板进行完全桡骨撕裂,然后将其分为三组,分别使用原始的2-0聚乳酸缝线、TAPS缝线(2-0聚乳酸缝线涂有粘接壳聚糖/海藻酸盐水凝胶)或2-0 UHMWPE单针缝线进行修复。修复后的半月板夹在Instron机器上,以0.16 Hz的频率进行5 - 25 N的循环加载。在500次循环后,使用电子卡尺测量撕裂边缘之间的间隙形成,当时半月板仍在Instron上,没有施加任何负载。结果:2-0聚乳酸缝线的平均间隙形成为5.22 mm(±1.70),TAPS缝线的平均间隙形成为2.48 mm(±0.25),2-0超高分子量聚乙烯缝线的平均间隙形成为4.85 mm(±1.55)。由于更好的力分散,TAPS缝合组的间隙明显小于其他两组,减少了缝合压痕造成的组织损伤,并可能导致更好的半月板愈合。结论:从生物力学的角度来看,与传统缝线相比,涂层缝线使桡骨半月板撕裂边缘更紧密地结合在一起。该技术具有减少组织损伤和提高半月板修复成功率的潜力。证据水平:实验室对照研究。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Journal of Experimental Orthopaedics
Journal of Experimental Orthopaedics Medicine-Orthopedics and Sports Medicine
CiteScore
3.20
自引率
5.60%
发文量
114
审稿时长
13 weeks
期刊最新文献
Calcaneo-stop for paediatric idiopathic flexible flatfoot: High functional results and return to sport in 644 feet at mid-term follow-up Reproducibility of a new device for robotic-assisted TKA surgery The central fibre areas in the tibial footprint of the posterior cruciate ligament show the highest contribution to restriction of a posterior drawer force—A biomechanical robotic investigation The short version of the ALR-RSI scale is a valid and reproducible scale to evaluate psychological readiness to return to sport after ankle lateral reconstruction Which treatment strategy for irreparable rotator cuff tears is most cost-effective? A Markov model-based cost-utility analysis comparing superior capsular reconstruction, lower trapezius tendon transfer, subacromial balloon spacer implantation and reverse shoulder arthroplasty
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1