Virtual reality-based Harms tangent screen test for strabismus measurement.

Yaroslava Wenner, Johann Schneider, Timo Drisser, Yu Yi Yang, Thilo Demeter, Maria Fronius, Birgit Lorenz, Thomas Kohnen, Michael Müller, Jochen Triesch
{"title":"Virtual reality-based Harms tangent screen test for strabismus measurement.","authors":"Yaroslava Wenner, Johann Schneider, Timo Drisser, Yu Yi Yang, Thilo Demeter, Maria Fronius, Birgit Lorenz, Thomas Kohnen, Michael Müller, Jochen Triesch","doi":"10.1007/s00417-024-06724-2","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Our study presents a virtual reality-based tangent screen test (VTS) to measure subjective ocular deviations including torsion in nine directions of gaze. The test was compared to the analogous Harms tangent screen test (HTS).</p><p><strong>Methods: </strong>We used an Oculus Go<sup>®</sup> controller and head-mounted-display with rotation sensors to measure patient's head orientation for the VTS. The software was developed with Unity<sup>®</sup>. We assessed subjective squint angles of adult patients with eye motility disorders of different origin in nine positions of gaze by means of HTS and VTS. We calculated mean difference, 95% limits of agreement and intraclass correlations (ICC) for horizontal, vertical and torsional deviations.</p><p><strong>Results: </strong>We included 85 patients. Measured horizontal and vertical deviations showed very good agreement between the two methods in all gaze directions (horizontal and vertical ICC: 0.93-0.98). Agreement was lower for torsional deviations (ICC: 0.79-0.93). The mean difference in primary position was 1.5° (95% limits of agreement -3.5° to 6.5°) for horizontal, 0° (-2,7° to 2,7°) for vertical, and 0.3° (-4.0° to 3.0°) for torsional deviations. The average examination time was 6 min with the VTS compared to 15 min with the HTS.</p><p><strong>Conclusions: </strong>Vertical and horizontal deviations showed good agreement between both tests measuring a slightly higher esodeviation with VTS probably due to an increased vergence demand. Measurement of torsional misalignment would benefit from a controller with more holding stability, such as a PC-mouse with central wheel, facilitating fine adjustments. VTS allowed automatic documentation, required less time and was easier to use than HTS.</p><p><strong>Key messages: </strong>What is known The incidence of neurogenic diplopia increases in the ageing population. The Harms tangent screen test (HTS) reliably measures subjective ocular deviations in patients with diplopia. The HTS requires staff well-trained in orthoptics limiting its use. What is new A novel virtual reality-based tangent screen test (VTS) enables measurement of horizontal, vertical and torsional ocular misalignment and shows a good comparability to the HTS. VTS uses commercially available head-mounted-display and controller, and specifically developed software, requires less time than HTS and allows automatic documentation. VTS administration does not require orthoptic training and can eventually be implemented in a clinical setting to provide a wider availability of subjective squint angle measurements.</p>","PeriodicalId":12795,"journal":{"name":"Graefe’s Archive for Clinical and Experimental Ophthalmology","volume":" ","pages":""},"PeriodicalIF":2.4000,"publicationDate":"2025-01-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Graefe’s Archive for Clinical and Experimental Ophthalmology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00417-024-06724-2","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: Our study presents a virtual reality-based tangent screen test (VTS) to measure subjective ocular deviations including torsion in nine directions of gaze. The test was compared to the analogous Harms tangent screen test (HTS).

Methods: We used an Oculus Go® controller and head-mounted-display with rotation sensors to measure patient's head orientation for the VTS. The software was developed with Unity®. We assessed subjective squint angles of adult patients with eye motility disorders of different origin in nine positions of gaze by means of HTS and VTS. We calculated mean difference, 95% limits of agreement and intraclass correlations (ICC) for horizontal, vertical and torsional deviations.

Results: We included 85 patients. Measured horizontal and vertical deviations showed very good agreement between the two methods in all gaze directions (horizontal and vertical ICC: 0.93-0.98). Agreement was lower for torsional deviations (ICC: 0.79-0.93). The mean difference in primary position was 1.5° (95% limits of agreement -3.5° to 6.5°) for horizontal, 0° (-2,7° to 2,7°) for vertical, and 0.3° (-4.0° to 3.0°) for torsional deviations. The average examination time was 6 min with the VTS compared to 15 min with the HTS.

Conclusions: Vertical and horizontal deviations showed good agreement between both tests measuring a slightly higher esodeviation with VTS probably due to an increased vergence demand. Measurement of torsional misalignment would benefit from a controller with more holding stability, such as a PC-mouse with central wheel, facilitating fine adjustments. VTS allowed automatic documentation, required less time and was easier to use than HTS.

Key messages: What is known The incidence of neurogenic diplopia increases in the ageing population. The Harms tangent screen test (HTS) reliably measures subjective ocular deviations in patients with diplopia. The HTS requires staff well-trained in orthoptics limiting its use. What is new A novel virtual reality-based tangent screen test (VTS) enables measurement of horizontal, vertical and torsional ocular misalignment and shows a good comparability to the HTS. VTS uses commercially available head-mounted-display and controller, and specifically developed software, requires less time than HTS and allows automatic documentation. VTS administration does not require orthoptic training and can eventually be implemented in a clinical setting to provide a wider availability of subjective squint angle measurements.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
基于虚拟现实的斜视测量哈姆斯切线屏幕测试。
目的:我们的研究提出了一种基于虚拟现实的切线屏幕测试(VTS)来测量主观眼偏差,包括九个凝视方向的扭转。将该试验与类似的Harms切线筛选试验(HTS)进行比较。方法:我们使用Oculus Go®控制器和带有旋转传感器的头戴式显示器来测量VTS患者的头部方向。该软件是用Unity®开发的。采用HTS和VTS对不同来源的成人眼动障碍患者在9个注视位置的主观斜视角度进行评估。我们计算了水平、垂直和扭转偏差的平均差、95%一致性限和类内相关性(ICC)。结果:纳入85例患者。测量的水平和垂直偏差显示两种方法在所有凝视方向上的一致性非常好(水平和垂直ICC: 0.93-0.98)。扭转偏差的一致性较低(ICC: 0.79-0.93)。在水平方向上,初始位置的平均差异为1.5°(95%的一致性限制为-3.5°至6.5°),在垂直方向上为0°(-2,7°至2,7°),在扭转方向上为0.3°(-4.0°至3.0°)。VTS组的平均检查时间为6分钟,而HTS组为15分钟。结论:垂直和水平偏差在两种测试之间显示出良好的一致性,测量了VTS稍高的内偏差,可能是由于增加的收敛需求。扭转偏差的测量将受益于一个更稳定的控制器,如带有中心轮的pc鼠标,便于精细调整。VTS允许自动文档,所需时间更少,比HTS更容易使用。神经源性复视的发病率随着人口老龄化而增加。哈姆斯切线筛检(HTS)可靠地测量复视患者的主观视力偏差。HTS需要训练有素的工作人员来限制其使用。一种新颖的基于虚拟现实的切线屏幕测试(VTS)可以测量水平、垂直和扭转眼不对准,并显示出与HTS的良好可比性。VTS使用市售的头戴式显示器和控制器,以及专门开发的软件,比HTS需要更少的时间,并允许自动记录。VTS管理不需要正视训练,最终可以在临床环境中实施,以提供更广泛的主观斜视角度测量。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
5.40
自引率
7.40%
发文量
398
审稿时长
3 months
期刊介绍: Graefe''s Archive for Clinical and Experimental Ophthalmology is a distinguished international journal that presents original clinical reports and clini-cally relevant experimental studies. Founded in 1854 by Albrecht von Graefe to serve as a source of useful clinical information and a stimulus for discussion, the journal has published articles by leading ophthalmologists and vision research scientists for more than a century. With peer review by an international Editorial Board and prompt English-language publication, Graefe''s Archive provides rapid dissemination of clinical and clinically related experimental information.
期刊最新文献
Reexamine the link between retinal layer thickness and cognitive function after correction of axial length: the Beijing Eye Study 2011. Comparison of the clinical outcomes after manual capsulorhexis on one eye and precision pulse capsulotomy on the other eye. Reply on comments on the paper "Descemet's membrane transplantation for the treatment of recurrent high myopic macular hole associated with retinal detachment". The rAAV2-ND1 gene therapy for Leber hereditary optic neuropathy. Adverse Events Associated with Devices for Incisional Glaucoma Surgery Performed with Implants as Reported to the FDA MAUDE Database.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1