Multicenter validation of a galactomannan chemiluminescence immunoassay for the diagnosis of pulmonary aspergillosis on serum of patients with hematological disease.
Hanne Lamberink, Sammy Huygens, Robina Aerts, Katrien Lagrou, Karin van Dijk, Diana Langerak, Ine Moors, Jerina Boelens, Marijke Reynders, Johan Maertens, Alexander Schauwvlieghe, Mireille van Westreenen, Ga-Lai M Chong, Paul E Verweij, Jochem B Buil, Bart J A Rijnders
{"title":"Multicenter validation of a galactomannan chemiluminescence immunoassay for the diagnosis of pulmonary aspergillosis on serum of patients with hematological disease.","authors":"Hanne Lamberink, Sammy Huygens, Robina Aerts, Katrien Lagrou, Karin van Dijk, Diana Langerak, Ine Moors, Jerina Boelens, Marijke Reynders, Johan Maertens, Alexander Schauwvlieghe, Mireille van Westreenen, Ga-Lai M Chong, Paul E Verweij, Jochem B Buil, Bart J A Rijnders","doi":"10.1128/jcm.01053-24","DOIUrl":null,"url":null,"abstract":"<p><p>An accurate diagnosis of invasive aspergillosis (IA) in patients with underlying hematological malignancies relies heavily on galactomannan detection. In this study, we compared the VirCLIA chemiluminescence immunoassay (CLIA) with the frequently used Platelia enzyme-linked immunosorbent assay (ELISA) on serum from hematology patients with suspected IA. Patients were categorized according to EORTC/MSGERC 2020 definitions into proven/probable IA and possible/no IA. The first cohort included 161 patients at four centers, and the VirCLIA manufacturer's cutoff of 0.200 was evaluated. Next, the optimal cutoff was determined using the Youden's index. In a second independent cohort of 189 patients from four centers, this optimal cutoff was evaluated again. In the first cohort, sensitivities and specificities for probable/proven IA were 21.1% and 100.0% for ELISA (1.0 cutoff) and 36.6% and 95.6% (0.5 cutoff), compared to 11.3% and 97.8% for CLIA (0.200 cutoff). In the second cohort, the sensitivities of ELISA and CLIA were comparable (ELISA ≥ 1.0: 33.3%, CLIA ≥ 0.200: 38.1%). The area under the ROC curve was lower for CLIA than for ELISA in the first cohort (65.0% vs 78.7%, <i>P</i> = 0.005) but comparable in the second cohort (79.5% vs 81.3%, <i>P</i> = 0.649). Youden's index identified 0.100 as the optimal CLIA cutoff with sensitivities of 35.2% and 61.9% in cohorts 1 and 2, respectively, at slightly reduced specificities of 85.6% and 90.5%. While the sensitivity of both assays was low to moderate at best, in patients with a high pre-test probability, we suggest 0.100 as the cutoff for the VirCLIA assay.IMPORTANCEThis study demonstrates a comparable performance of the novel chemiluminescence immunoassay (CLIA) and the conventionally used enzyme-linked immunosorbent assay for galactomannan serum testing in hematological patients at high risk for invasive aspergillosis. In patients with a high pre-test probability, a lower CLIA cutoff of 0.100 is preferred.</p>","PeriodicalId":15511,"journal":{"name":"Journal of Clinical Microbiology","volume":" ","pages":"e0105324"},"PeriodicalIF":6.1000,"publicationDate":"2025-02-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11837514/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Clinical Microbiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1128/jcm.01053-24","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/1/21 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"MICROBIOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
An accurate diagnosis of invasive aspergillosis (IA) in patients with underlying hematological malignancies relies heavily on galactomannan detection. In this study, we compared the VirCLIA chemiluminescence immunoassay (CLIA) with the frequently used Platelia enzyme-linked immunosorbent assay (ELISA) on serum from hematology patients with suspected IA. Patients were categorized according to EORTC/MSGERC 2020 definitions into proven/probable IA and possible/no IA. The first cohort included 161 patients at four centers, and the VirCLIA manufacturer's cutoff of 0.200 was evaluated. Next, the optimal cutoff was determined using the Youden's index. In a second independent cohort of 189 patients from four centers, this optimal cutoff was evaluated again. In the first cohort, sensitivities and specificities for probable/proven IA were 21.1% and 100.0% for ELISA (1.0 cutoff) and 36.6% and 95.6% (0.5 cutoff), compared to 11.3% and 97.8% for CLIA (0.200 cutoff). In the second cohort, the sensitivities of ELISA and CLIA were comparable (ELISA ≥ 1.0: 33.3%, CLIA ≥ 0.200: 38.1%). The area under the ROC curve was lower for CLIA than for ELISA in the first cohort (65.0% vs 78.7%, P = 0.005) but comparable in the second cohort (79.5% vs 81.3%, P = 0.649). Youden's index identified 0.100 as the optimal CLIA cutoff with sensitivities of 35.2% and 61.9% in cohorts 1 and 2, respectively, at slightly reduced specificities of 85.6% and 90.5%. While the sensitivity of both assays was low to moderate at best, in patients with a high pre-test probability, we suggest 0.100 as the cutoff for the VirCLIA assay.IMPORTANCEThis study demonstrates a comparable performance of the novel chemiluminescence immunoassay (CLIA) and the conventionally used enzyme-linked immunosorbent assay for galactomannan serum testing in hematological patients at high risk for invasive aspergillosis. In patients with a high pre-test probability, a lower CLIA cutoff of 0.100 is preferred.
侵袭性曲霉病(IA)的准确诊断在潜在的血液恶性肿瘤患者很大程度上依赖于半乳甘露聚糖检测。在本研究中,我们比较了VirCLIA化学发光免疫分析法(CLIA)与常用的血小板酶联免疫吸附法(ELISA)对疑似IA血液学患者血清的检测结果。根据EORTC/MSGERC 2020定义,将患者分为已证实/可能IA和可能/非IA。第一组包括四个中心的161名患者,并评估了VirCLIA制造商的临界值0.200。其次,使用约登指数确定最佳截止点。在来自四个中心的189名患者的第二个独立队列中,再次评估了这一最佳截止值。在第一个队列中,ELISA对可能/已证实的IA的敏感性和特异性分别为21.1%和100.0%(1.0临界值),36.6%和95.6%(0.5临界值),而CLIA为11.3%和97.8%(0.200临界值)。在第二队列中,ELISA和CLIA的敏感性具有可比性(ELISA≥1.0:33.3%,CLIA≥0.200:38.1%)。CLIA的ROC曲线下面积在第一队列中低于ELISA (65.0% vs 78.7%, P = 0.005),但在第二队列中相当(79.5% vs 81.3%, P = 0.649)。约登指数确定0.100为最佳CLIA截止点,在队列1和队列2中,敏感性分别为35.2%和61.9%,特异性略有降低,分别为85.6%和90.5%。虽然这两种检测方法的灵敏度最高为低至中等,但在检测前概率较高的患者中,我们建议将0.100作为VirCLIA检测的截止值。该研究证明了新型化学发光免疫分析法(CLIA)和常规使用的酶联免疫吸附法在侵袭性曲霉病高危血液学患者半乳甘露聚糖血清检测中的可比性。对于检测前概率较高的患者,最好将CLIA截止值降低至0.100。
期刊介绍:
The Journal of Clinical Microbiology® disseminates the latest research concerning the laboratory diagnosis of human and animal infections, along with the laboratory's role in epidemiology and the management of infectious diseases.