Outpatient parenteral antibiotic therapy with elastomeric pumps, what is the flowrate accuracy?

IF 1.7 3区 医学 Q3 PERIPHERAL VASCULAR DISEASE Journal of Vascular Access Pub Date : 2025-11-01 Epub Date: 2025-01-29 DOI:10.1177/11297298251314615
Arthur Muller, Yann Rétory, Laura Négrier, Benoit Lemarié, Anne Pfeifer, Alexandra Schmidt, Christian Dupont, Clara Virbel-Fleischman
{"title":"Outpatient parenteral antibiotic therapy with elastomeric pumps, what is the flowrate accuracy?","authors":"Arthur Muller, Yann Rétory, Laura Négrier, Benoit Lemarié, Anne Pfeifer, Alexandra Schmidt, Christian Dupont, Clara Virbel-Fleischman","doi":"10.1177/11297298251314615","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Outpatient Parenteral Antibiotic Therapy (OPAT) services have expanded throughout the world thanks to elastomeric pumps (EP). Their flowrate is known to vary by ±15%. Treatment effectiveness and the organization of care at home may be impacted, especially with piperacillin/tazobactam infusion (TAZ). An in vitro study was conducted, following the guidelines of EP and drug manufacturers, to assess the impact of three parameters on the accuracy of the infusion flowrate: composition, volume, and temperature of the infused solution.</p><p><strong>Methods: </strong>A test bench was created to evaluate the accuracy of flowrate infusion when changing one parameter at a time (addition of TAZ in the infused solution, minimal filling volume and flow regulator at ambient temperature) compared to the standard conditions provided by the manufacturer. Two types of EP were used and connected to a weighing scale. Instantaneous and average mass flowrates over the total infusion time were determined from the reservoir mass loss. These values  were compared with the nominal flowrates announced by the manufacturers to evaluate the performance of EP under the different conditions. The time the flowrate was within the expected ±15% around the nominal flowrate was calculated as the time in range (TIR).</p><p><strong>Results: </strong>Deviations of the measured mean flowrate from the nominal flowrate range from -37% to 18% depending on the condition. TIR went from 0% to 80% of the total infusion duration. The test bench showed a reliability of more than 96%.</p><p><strong>Conclusions: </strong>The test bench highlights differences in performance. These results raise several questions about the sensitivity of EPs for TAZ infusion in OPAT, the norm's requirements and the need to prescribe a medical device whose limits are well known, as well as therapeutic education or the course of care for treatment administration.</p>","PeriodicalId":56113,"journal":{"name":"Journal of Vascular Access","volume":" ","pages":"1990-1995"},"PeriodicalIF":1.7000,"publicationDate":"2025-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Vascular Access","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/11297298251314615","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/1/29 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Outpatient Parenteral Antibiotic Therapy (OPAT) services have expanded throughout the world thanks to elastomeric pumps (EP). Their flowrate is known to vary by ±15%. Treatment effectiveness and the organization of care at home may be impacted, especially with piperacillin/tazobactam infusion (TAZ). An in vitro study was conducted, following the guidelines of EP and drug manufacturers, to assess the impact of three parameters on the accuracy of the infusion flowrate: composition, volume, and temperature of the infused solution.

Methods: A test bench was created to evaluate the accuracy of flowrate infusion when changing one parameter at a time (addition of TAZ in the infused solution, minimal filling volume and flow regulator at ambient temperature) compared to the standard conditions provided by the manufacturer. Two types of EP were used and connected to a weighing scale. Instantaneous and average mass flowrates over the total infusion time were determined from the reservoir mass loss. These values  were compared with the nominal flowrates announced by the manufacturers to evaluate the performance of EP under the different conditions. The time the flowrate was within the expected ±15% around the nominal flowrate was calculated as the time in range (TIR).

Results: Deviations of the measured mean flowrate from the nominal flowrate range from -37% to 18% depending on the condition. TIR went from 0% to 80% of the total infusion duration. The test bench showed a reliability of more than 96%.

Conclusions: The test bench highlights differences in performance. These results raise several questions about the sensitivity of EPs for TAZ infusion in OPAT, the norm's requirements and the need to prescribe a medical device whose limits are well known, as well as therapeutic education or the course of care for treatment administration.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
门诊用弹性泵进行肠外抗生素治疗,流量准确性如何?
背景:得益于弹性泵(EP),门诊肠道外抗生素治疗(OPAT)服务已在全球范围内扩展。已知其流量变化为±15%。治疗效果和家庭护理的组织可能会受到影响,特别是使用哌拉西林/他唑巴坦输注(TAZ)。在体外实验中,我们按照EP和药品制造商的指导方针,评估了输液液的组成、体积和温度这三个参数对输液流速准确性的影响。方法:建立一个试验台,与制造商提供的标准条件相比,每次改变一个参数(注射溶液中TAZ的添加,最小填充量和环境温度下的流量调节器)来评估流速注射的准确性。使用了两种EP并将其连接到称重秤上。瞬时和平均质量流量在总注入时间内由水库的质量损失确定。将这些值与制造商公布的标称流量进行比较,以评估EP在不同条件下的性能。流量在公称流量的±15%范围内的时间被计算为范围内时间(TIR)。结果:测量的平均流量与标称流量的偏差范围从-37%到18%,具体取决于条件。TIR从0%到80%的总输注时间。试验台的可靠性达到96%以上。结论:测试台突出了性能上的差异。这些结果提出了几个问题,关于OPAT中输注TAZ的EPs的敏感性,规范的要求和处方限制众所周知的医疗器械的必要性,以及治疗教育或治疗管理的护理过程。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Journal of Vascular Access
Journal of Vascular Access 医学-外周血管病
CiteScore
3.40
自引率
31.60%
发文量
181
审稿时长
6-12 weeks
期刊介绍: The Journal of Vascular Access (JVA) is issued six times per year; it considers the publication of original manuscripts dealing with clinical and laboratory investigations in the fast growing field of vascular access. In addition reviews, case reports and clinical trials are welcome, as well as papers dedicated to more practical aspects covering new devices and techniques. All contributions, coming from all over the world, undergo the peer-review process. The Journal of Vascular Access is divided into independent sections, each led by Editors of the highest scientific level: • Dialysis • Oncology • Interventional radiology • Nutrition • Nursing • Intensive care Correspondence related to published papers is also welcome.
期刊最新文献
Port REMoval Outcomes (PREMO) study: A pilot study for functional, microbial, radiological, and macroscopic assessment of totally implantable venous access devices. Ultrasound-guided arterial catheterization with acoustic shadowing technique: A randomized controlled trial. Arteriovenous fistula after distal transradial access: A case report and literature review. Evaluating needle-free connectors associated backflow in Midline and peripherally inserted central catheters: A top bench study. Short-term tunneling of centrally inserted central venous catheters without additional kits: A practical, safe, and emergency-applicable technique.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1