Determination of carbon dioxide by gas chromatography using an electron capture detector for the analysis of greenhouse gases: A comparison and validation with the standard method
Joan Noguerol Arias, August Bonmatí, Francesc X. Prenafeta-Boldú, Míriam Cerrillo
{"title":"Determination of carbon dioxide by gas chromatography using an electron capture detector for the analysis of greenhouse gases: A comparison and validation with the standard method","authors":"Joan Noguerol Arias, August Bonmatí, Francesc X. Prenafeta-Boldú, Míriam Cerrillo","doi":"10.1016/j.chroma.2025.465750","DOIUrl":null,"url":null,"abstract":"<div><div>This study reports the development of a novel analytical method for the determination of carbon dioxide (CO<sub>2</sub>) along with the other main greenhouse gases (GHG)‒methane (CH<sub>4</sub>) and nitrous oxide (N<sub>2</sub>O)‒ from environmental samples by gas chromatography (GC) using an electron capture detector (ECD). This new method was compared and validated with the standard GC method for the determination of CO<sub>2</sub> using a thermal conductivity detector (TCD). The main performance parameters considered for the validation of an analytical method were evaluated: selectivity/specificity, linearity/working range, precision, trueness, limit of detection (LOD) and limit of quantitation (LOQ). This comparison was carried out using gas standards, reference materials and environmental samples containing GHG. Both the ECD- and TCD-based chromatographic analytical methods displayed a similar precision (3.1–3.4 %) and accuracy (101–106 %) for CO<sub>2</sub> analysis. Although the LOQ for the ECD detector is higher than that of the TCD detector (300 vs. 99 µmol mol⁻¹), it is still sufficient for the analysis of most GHG samples. Once validated, the new analytical method was used for the simultaneous determination of CO<sub>2</sub>, CH<sub>4</sub> and N<sub>2</sub>O in gaseous samples obtained from a wide range of conditions and agricultural environments. To this end, a GC instrument was equipped with a flame ionization detector (FID) for the determination of CH<sub>4</sub> and an ECD for CO<sub>2</sub> and N<sub>2</sub>O, connected in series by a system of valves. In this way, it was possible to accurately measure the main GHG present in the atmosphere and in gaseous samples, simplifying the required laboratory equipment and reducing the associated labor and costs.</div></div>","PeriodicalId":347,"journal":{"name":"Journal of Chromatography A","volume":"1745 ","pages":"Article 465750"},"PeriodicalIF":3.8000,"publicationDate":"2025-02-03","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Chromatography A","FirstCategoryId":"1","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0021967325000998","RegionNum":2,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"BIOCHEMICAL RESEARCH METHODS","Score":null,"Total":0}
引用次数: 0
Abstract
This study reports the development of a novel analytical method for the determination of carbon dioxide (CO2) along with the other main greenhouse gases (GHG)‒methane (CH4) and nitrous oxide (N2O)‒ from environmental samples by gas chromatography (GC) using an electron capture detector (ECD). This new method was compared and validated with the standard GC method for the determination of CO2 using a thermal conductivity detector (TCD). The main performance parameters considered for the validation of an analytical method were evaluated: selectivity/specificity, linearity/working range, precision, trueness, limit of detection (LOD) and limit of quantitation (LOQ). This comparison was carried out using gas standards, reference materials and environmental samples containing GHG. Both the ECD- and TCD-based chromatographic analytical methods displayed a similar precision (3.1–3.4 %) and accuracy (101–106 %) for CO2 analysis. Although the LOQ for the ECD detector is higher than that of the TCD detector (300 vs. 99 µmol mol⁻¹), it is still sufficient for the analysis of most GHG samples. Once validated, the new analytical method was used for the simultaneous determination of CO2, CH4 and N2O in gaseous samples obtained from a wide range of conditions and agricultural environments. To this end, a GC instrument was equipped with a flame ionization detector (FID) for the determination of CH4 and an ECD for CO2 and N2O, connected in series by a system of valves. In this way, it was possible to accurately measure the main GHG present in the atmosphere and in gaseous samples, simplifying the required laboratory equipment and reducing the associated labor and costs.
期刊介绍:
The Journal of Chromatography A provides a forum for the publication of original research and critical reviews on all aspects of fundamental and applied separation science. The scope of the journal includes chromatography and related techniques, electromigration techniques (e.g. electrophoresis, electrochromatography), hyphenated and other multi-dimensional techniques, sample preparation, and detection methods such as mass spectrometry. Contributions consist mainly of research papers dealing with the theory of separation methods, instrumental developments and analytical and preparative applications of general interest.