Novel Software Program to Improve Biometry Measurements Obtained by an Optical Low Coherence Reflectometry Biometer.

IF 2.6 3区 医学 Q2 OPHTHALMOLOGY Journal of cataract and refractive surgery Pub Date : 2025-02-11 DOI:10.1097/j.jcrs.0000000000001630
David L Cooke, Tim L Cooke, Raj Patel, Michael T Heath, Mason Gardiner, David A Murphy, Kamran M Riaz
{"title":"Novel Software Program to Improve Biometry Measurements Obtained by an Optical Low Coherence Reflectometry Biometer.","authors":"David L Cooke, Tim L Cooke, Raj Patel, Michael T Heath, Mason Gardiner, David A Murphy, Kamran M Riaz","doi":"10.1097/j.jcrs.0000000000001630","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To evaluate a novel software program (SpikeFinder) that improves quality of measurements obtained by optical low coherence reflectometry (OLCR) biometry (Lenstar).</p><p><strong>Setting: </strong>Tertiary care academic center.</p><p><strong>Design: </strong>Retrospective case series.</p><p><strong>Method: </strong>ology: After applying inclusion criteria, 12,988 eyes scanned with OLCR biometry and swept-source OCT (SS-OCT) biometry (IOLMaster700) were analyzed. Measurements, including axial length (AL), anterior chamber depth (ACD), and lens thickness (LT), were compared. SpikeFinder was applied to eyes with the most prominent ACD and LT measurement differences. Statistical testing was performed using Wilcoxon (two variables) and Kruskal-Wallis ANOVA tests (multiple variables) with Tukey post-hoc adjustments. Heteroscedastic testing (WHWK Statistics Package) was used to assess differences after software application.</p><p><strong>Results: </strong>99.1% of AL measurements were within 0.01 mm between the two biometers. SS-OCT LT was significantly longer than the OLCR LT: 33% differed by ≥ 0.50 mm (p<0.001). Correspondingly, SS-OCT ACD was significantly shorter than OLCR ACD: 14% differed by ≥ 0.50 mm (p<0.001). SS-OCT biometer - OLCR biometer values for ACD and LT before applying SpikeFinder were -0.64 ± 0.31 mm and 1.15 ± 0.38 mm, respectively (p<0.001). After applying SpikeFinder to a subgroup of 337 eyes, these differences were -0.02 ± 0.03 mm and 0.01 ± 0.11 mm, respectively. SpikeFinder ACD and LT values were similar to SS-OCT biometer-measured ACD (p=0.872) and LT (p=0.987) values.</p><p><strong>Conclusion: </strong>The SpikeFinder program can help improve the accuracy of ACD and LT measurements obtained by the OLCR biometer, which may enhance the performance of formulas that utilize these measurements.</p>","PeriodicalId":15214,"journal":{"name":"Journal of cataract and refractive surgery","volume":" ","pages":""},"PeriodicalIF":2.6000,"publicationDate":"2025-02-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of cataract and refractive surgery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/j.jcrs.0000000000001630","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: To evaluate a novel software program (SpikeFinder) that improves quality of measurements obtained by optical low coherence reflectometry (OLCR) biometry (Lenstar).

Setting: Tertiary care academic center.

Design: Retrospective case series.

Method: ology: After applying inclusion criteria, 12,988 eyes scanned with OLCR biometry and swept-source OCT (SS-OCT) biometry (IOLMaster700) were analyzed. Measurements, including axial length (AL), anterior chamber depth (ACD), and lens thickness (LT), were compared. SpikeFinder was applied to eyes with the most prominent ACD and LT measurement differences. Statistical testing was performed using Wilcoxon (two variables) and Kruskal-Wallis ANOVA tests (multiple variables) with Tukey post-hoc adjustments. Heteroscedastic testing (WHWK Statistics Package) was used to assess differences after software application.

Results: 99.1% of AL measurements were within 0.01 mm between the two biometers. SS-OCT LT was significantly longer than the OLCR LT: 33% differed by ≥ 0.50 mm (p<0.001). Correspondingly, SS-OCT ACD was significantly shorter than OLCR ACD: 14% differed by ≥ 0.50 mm (p<0.001). SS-OCT biometer - OLCR biometer values for ACD and LT before applying SpikeFinder were -0.64 ± 0.31 mm and 1.15 ± 0.38 mm, respectively (p<0.001). After applying SpikeFinder to a subgroup of 337 eyes, these differences were -0.02 ± 0.03 mm and 0.01 ± 0.11 mm, respectively. SpikeFinder ACD and LT values were similar to SS-OCT biometer-measured ACD (p=0.872) and LT (p=0.987) values.

Conclusion: The SpikeFinder program can help improve the accuracy of ACD and LT measurements obtained by the OLCR biometer, which may enhance the performance of formulas that utilize these measurements.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
5.60
自引率
14.30%
发文量
259
审稿时长
8.5 weeks
期刊介绍: The Journal of Cataract & Refractive Surgery (JCRS), a preeminent peer-reviewed monthly ophthalmology publication, is the official journal of the American Society of Cataract and Refractive Surgery (ASCRS) and the European Society of Cataract and Refractive Surgeons (ESCRS). JCRS publishes high quality articles on all aspects of anterior segment surgery. In addition to original clinical studies, the journal features a consultation section, practical techniques, important cases, and reviews as well as basic science articles.
期刊最新文献
Effect of a Topical Antibiotic and Povidone-Iodine Versus Povidone Iodine Alone on Conjunctival Flora: A Systematic Review and Meta-Analysis. In vivo capsular bag size in children with congenital cataract: Implications for placement of Intraocular Lens. Late refractive change after cataract extraction and toric intraocular lens implantation. Novel Software Program to Improve Biometry Measurements Obtained by an Optical Low Coherence Reflectometry Biometer. Visual Outcomes following implantation of a non-diffractive extended depth of focus Intraocular Lens in patients with early dry macular degeneration.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1