Feasibility of MRI-Guided Cardiac Catheterization, Angioplasty, and Stenting in a Commercial Wide-bore 0.55T Scanner.

IF 4.2 1区 医学 Q1 CARDIAC & CARDIOVASCULAR SYSTEMS Journal of Cardiovascular Magnetic Resonance Pub Date : 2025-02-10 DOI:10.1016/j.jocmr.2025.101858
Aimee K Armstrong, Yixuan Liu, John M Kelly, Ramkumar Krishnamurthy, Jason Swinning, Yingmin Liu, Matthew Joseph, Ning Jin, Jianing Pang, Florian Maier, Axel J Krafft, Orville Bramwell, Nathan Ooms, Jesse Roll, Joshua Krieger, David C Gross, Lucien de Mos, Paul Borm, Orlando P Simonetti
{"title":"Feasibility of MRI-Guided Cardiac Catheterization, Angioplasty, and Stenting in a Commercial Wide-bore 0.55T Scanner.","authors":"Aimee K Armstrong, Yixuan Liu, John M Kelly, Ramkumar Krishnamurthy, Jason Swinning, Yingmin Liu, Matthew Joseph, Ning Jin, Jianing Pang, Florian Maier, Axel J Krafft, Orville Bramwell, Nathan Ooms, Jesse Roll, Joshua Krieger, David C Gross, Lucien de Mos, Paul Borm, Orlando P Simonetti","doi":"10.1016/j.jocmr.2025.101858","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Low-field (0.55T) magnetic resonance imaging (MRI) may allow MRI-guided interventions using available catheterization equipment, as radiofrequency-induced heating of interventional devices is reduced at low field.</p><p><strong>Objective: </strong>The purpose of this study was to test the feasibility of real-time MRI-guided right and left heart catheterization (R&LHC) as well as angioplasty and stenting of the inferior vena cava using a commercially available 0.55T MRI system (MAGNETOM Free.Max, Siemens Healthineers AG, Erlangen, Germany) with 80cm patient bore and maximum gradient amplitude and slew rate of 26 mT/m and 45 mT/m/ms, respectively. A secondary aim was to evaluate three different sizes of MR-visible markers.</p><p><strong>Method: </strong>Sheaths were placed in the femoral vein and artery of juvenile Yorkshire pigs under general anesthesia. Air-filled balloon wedge catheters and Judkins Right catheters were used for R&LHC, respectively, aided by an MR-compatible guidewire. Ferumoxytol was administered as a T1-shortening contrast agent and real-time visualization was carried out with a research interactive sequence using different spatiotemporal resolution settings. Inferior vena cava (IVC) angioplasty was performed using balloons filled with 1% gadolinium, and IVC stenting was performed with stainless steel stents and one platinum iridium covered stent.</p><p><strong>Result: </strong>RHC was successful in all 8 attempts with balloon tip visibility in all real-time protocols. One pig expired with ferumoxytol infusion, but the catheterization was completed post-mortem. LHC and IVC angioplasty were attempted and successful in 2 and 4 pigs, respectively. For stenting, higher resolution, lower frame rate imaging was used. All 6 attempted stent implantations were successful. The MR markers on the angioplasty balloon with widths of 0.5 and 1mm were more visible than the 0.25mm markers. Marker placement affected distinguishability from the crimped stent. Stainless steel stents created only minimal signal voids pre- and post-deployment; however, the platinum-iridium stent caused significant artifact obscuring wall apposition assessment.</p><p><strong>Conclusion: </strong>This study is the first to demonstrate the technical feasibility of R&LHC, IVC angioplasty, and IVC stenting using real-time MRI on a commercially available low field scanner.</p>","PeriodicalId":15221,"journal":{"name":"Journal of Cardiovascular Magnetic Resonance","volume":" ","pages":"101858"},"PeriodicalIF":4.2000,"publicationDate":"2025-02-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Cardiovascular Magnetic Resonance","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.jocmr.2025.101858","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Low-field (0.55T) magnetic resonance imaging (MRI) may allow MRI-guided interventions using available catheterization equipment, as radiofrequency-induced heating of interventional devices is reduced at low field.

Objective: The purpose of this study was to test the feasibility of real-time MRI-guided right and left heart catheterization (R&LHC) as well as angioplasty and stenting of the inferior vena cava using a commercially available 0.55T MRI system (MAGNETOM Free.Max, Siemens Healthineers AG, Erlangen, Germany) with 80cm patient bore and maximum gradient amplitude and slew rate of 26 mT/m and 45 mT/m/ms, respectively. A secondary aim was to evaluate three different sizes of MR-visible markers.

Method: Sheaths were placed in the femoral vein and artery of juvenile Yorkshire pigs under general anesthesia. Air-filled balloon wedge catheters and Judkins Right catheters were used for R&LHC, respectively, aided by an MR-compatible guidewire. Ferumoxytol was administered as a T1-shortening contrast agent and real-time visualization was carried out with a research interactive sequence using different spatiotemporal resolution settings. Inferior vena cava (IVC) angioplasty was performed using balloons filled with 1% gadolinium, and IVC stenting was performed with stainless steel stents and one platinum iridium covered stent.

Result: RHC was successful in all 8 attempts with balloon tip visibility in all real-time protocols. One pig expired with ferumoxytol infusion, but the catheterization was completed post-mortem. LHC and IVC angioplasty were attempted and successful in 2 and 4 pigs, respectively. For stenting, higher resolution, lower frame rate imaging was used. All 6 attempted stent implantations were successful. The MR markers on the angioplasty balloon with widths of 0.5 and 1mm were more visible than the 0.25mm markers. Marker placement affected distinguishability from the crimped stent. Stainless steel stents created only minimal signal voids pre- and post-deployment; however, the platinum-iridium stent caused significant artifact obscuring wall apposition assessment.

Conclusion: This study is the first to demonstrate the technical feasibility of R&LHC, IVC angioplasty, and IVC stenting using real-time MRI on a commercially available low field scanner.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
10.90
自引率
12.50%
发文量
61
审稿时长
6-12 weeks
期刊介绍: Journal of Cardiovascular Magnetic Resonance (JCMR) publishes high-quality articles on all aspects of basic, translational and clinical research on the design, development, manufacture, and evaluation of cardiovascular magnetic resonance (CMR) methods applied to the cardiovascular system. Topical areas include, but are not limited to: New applications of magnetic resonance to improve the diagnostic strategies, risk stratification, characterization and management of diseases affecting the cardiovascular system. New methods to enhance or accelerate image acquisition and data analysis. Results of multicenter, or larger single-center studies that provide insight into the utility of CMR. Basic biological perceptions derived by CMR methods.
期刊最新文献
How low can we go? The effect of acquisition duration on cardiac volume and function measurements in free-running cardiac and respiratory motion-resolved 5D whole-heart cine MRI at 1.5T. Hypothermia as an adjunctive therapy to percutaneous intervention after ST-elevation myocardial infarction - Effects on regional myocardial contractility. Late Gadolinium-Enhanced Cardiac Magnetic Resonance for Predicting Left Ventricular Reverse Remodeling in Dilated Cardiomyopathy A Comprehensive Review and Meta-Analysis. Doppler Ultrasound Gating for Adult Cardiovascular Magnetic Resonance: Initial Experience. Diagnostic value of bone scintigraphy versus CMR in cardiac amyloidosis.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1