American Clinical Neurophysiology Society Technical Standards for Electrical Stimulation With Intracranial Electrodes for Functional Brain Mapping and Seizure Induction.

IF 2.3 4区 医学 Q3 CLINICAL NEUROLOGY Journal of Clinical Neurophysiology Pub Date : 2025-02-12 DOI:10.1097/WNP.0000000000001149
Ravindra Arya, Fiona M Baumer, Patrick Chauvel, Birgit Frauscher, Prasanna Jayakar, Ammar Kheder, Bradley Lega, Ronald P Lesser, Kai J Miller, Marc R Nuwer, Nigel P Pedersen, Anthony L Ritaccio, David S Sabsevitz, Saurabh R Sinha, Elson L So, William O Tatum, Jessica W Templer, Stephan U Schuele
{"title":"American Clinical Neurophysiology Society Technical Standards for Electrical Stimulation With Intracranial Electrodes for Functional Brain Mapping and Seizure Induction.","authors":"Ravindra Arya, Fiona M Baumer, Patrick Chauvel, Birgit Frauscher, Prasanna Jayakar, Ammar Kheder, Bradley Lega, Ronald P Lesser, Kai J Miller, Marc R Nuwer, Nigel P Pedersen, Anthony L Ritaccio, David S Sabsevitz, Saurabh R Sinha, Elson L So, William O Tatum, Jessica W Templer, Stephan U Schuele","doi":"10.1097/WNP.0000000000001149","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>These American Clinical Neurophysiology Society technical standards suggest best practices for electrical stimulation mapping (ESM) with subdural and stereotactic depth electrodes for seizure induction and mapping of brain function.</p><p><strong>Methods: </strong>A working group was convened from American Clinical Neurophysiology Society membership with expertise in ESM. PubMed searches were performed to identify pertinent peer-reviewed literature. Recurrent meetings reviewed progress, built consensus by discussion, and developed evidence-based recommendations to the extent possible.</p><p><strong>Recommendations: </strong>Stimulators used for ESM should have sufficient dynamic range, ability to interrupt a stimulus train, and ictal disrupt mechanism(s). Charge density should be calculated for the specific electrodes and ESM settings, the maximum safe limits being 52 to 57 μC/cm2/phase for subdural electrodes and not established for stereotactic depth electrodes. Subdural ESM for functional mapping is typically performed at 50-Hz pulse frequency, 200- to 300-µs pulse width, 2- to 8-s train duration, and 1- to 20-mA current strength. Stereo ESM is commonly performed using 2 different pulse frequencies: 1 Hz (300-500 µs pulse width, train duration up to 30 s, and often a constant current of 3-5 mA), and 50 Hz (100-500 µs pulse width, train duration 2-8 s, and 0.5-10 mA current intensity).</p><p><strong>Conclusions: </strong>This guideline provides technical standards for the performance of ESM, which will likely evolve over time with advances in technology and additional evidence (also see Graphical Abstract).</p>","PeriodicalId":15516,"journal":{"name":"Journal of Clinical Neurophysiology","volume":" ","pages":""},"PeriodicalIF":2.3000,"publicationDate":"2025-02-12","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Clinical Neurophysiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/WNP.0000000000001149","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CLINICAL NEUROLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: These American Clinical Neurophysiology Society technical standards suggest best practices for electrical stimulation mapping (ESM) with subdural and stereotactic depth electrodes for seizure induction and mapping of brain function.

Methods: A working group was convened from American Clinical Neurophysiology Society membership with expertise in ESM. PubMed searches were performed to identify pertinent peer-reviewed literature. Recurrent meetings reviewed progress, built consensus by discussion, and developed evidence-based recommendations to the extent possible.

Recommendations: Stimulators used for ESM should have sufficient dynamic range, ability to interrupt a stimulus train, and ictal disrupt mechanism(s). Charge density should be calculated for the specific electrodes and ESM settings, the maximum safe limits being 52 to 57 μC/cm2/phase for subdural electrodes and not established for stereotactic depth electrodes. Subdural ESM for functional mapping is typically performed at 50-Hz pulse frequency, 200- to 300-µs pulse width, 2- to 8-s train duration, and 1- to 20-mA current strength. Stereo ESM is commonly performed using 2 different pulse frequencies: 1 Hz (300-500 µs pulse width, train duration up to 30 s, and often a constant current of 3-5 mA), and 50 Hz (100-500 µs pulse width, train duration 2-8 s, and 0.5-10 mA current intensity).

Conclusions: This guideline provides technical standards for the performance of ESM, which will likely evolve over time with advances in technology and additional evidence (also see Graphical Abstract).

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
求助全文
约1分钟内获得全文 去求助
来源期刊
Journal of Clinical Neurophysiology
Journal of Clinical Neurophysiology 医学-临床神经学
CiteScore
4.60
自引率
4.20%
发文量
198
审稿时长
6-12 weeks
期刊介绍: ​The Journal of Clinical Neurophysiology features both topical reviews and original research in both central and peripheral neurophysiology, as related to patient evaluation and treatment. Official Journal of the American Clinical Neurophysiology Society.
期刊最新文献
Epilepsy Surgery Education: A Survey of US Epilepsy Fellowship Program Directors. American Clinical Neurophysiology Society Technical Standards for Electrical Stimulation With Intracranial Electrodes for Functional Brain Mapping and Seizure Induction. Reference Values of the Tube-Based Laryngeal Adductor Reflex of the Healthy Recurrent Laryngeal Nerve. Accuracy of SEEG Source Localization: A Pilot Study Using Corticocortical Evoked Potentials. Real-Time Seizure Detection Using Behind-the-Ear Wearable System.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1