Li Guanqiang, Hu Bo, Sun Yuan, Chen Ming, Zhang Xicheng, Huang Xianchen
{"title":"Invention of a new percutaneous closure technique for vascular haemostasis in percutaneous veno-arterial extracorporeal membrane oxygenation.","authors":"Li Guanqiang, Hu Bo, Sun Yuan, Chen Ming, Zhang Xicheng, Huang Xianchen","doi":"10.1186/s40001-025-02307-x","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To assess the feasibility and safety of a novel closure technique, namely the parallel closure technique (Par-CT), in which the sheath is placed parallel to the ProGlide system during veno-arterial extracorporeal membrane oxygenation (VA-ECMO) to enhance the efficiency of percutaneous vascular haemostasis.</p><p><strong>Methods: </strong>The abdominal aortas of goats were punctured with a 17F sheath to simulate arterial cannula vascular access for VA-ECMO, and the Par-CT was used to achieve vascular haemostasis. In five goats, the abdominal aorta was punctured with a 17F sheath and then a 6 to 10F vascular sheath was introduced. This resulted in a smaller vascular rupture area and successful closure of the aortic rupture using the Par-CT.</p><p><strong>Results: </strong>If the sheath tube that was introduced was < 8 F, the puncture site could not be closed and would therefore have to be resutured and the Par-CT would have to be repeated. However, the puncture site could be sutured appropriately if the sheath tube that was introduced was ≥ 8F.</p><p><strong>Conclusions: </strong>This study suggested that the Par-CT is both safe and effective for haemostasis after weaning from VA-ECMO.</p>","PeriodicalId":11949,"journal":{"name":"European Journal of Medical Research","volume":"30 1","pages":"119"},"PeriodicalIF":2.8000,"publicationDate":"2025-02-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11837478/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"European Journal of Medical Research","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1186/s40001-025-02307-x","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MEDICINE, RESEARCH & EXPERIMENTAL","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: To assess the feasibility and safety of a novel closure technique, namely the parallel closure technique (Par-CT), in which the sheath is placed parallel to the ProGlide system during veno-arterial extracorporeal membrane oxygenation (VA-ECMO) to enhance the efficiency of percutaneous vascular haemostasis.
Methods: The abdominal aortas of goats were punctured with a 17F sheath to simulate arterial cannula vascular access for VA-ECMO, and the Par-CT was used to achieve vascular haemostasis. In five goats, the abdominal aorta was punctured with a 17F sheath and then a 6 to 10F vascular sheath was introduced. This resulted in a smaller vascular rupture area and successful closure of the aortic rupture using the Par-CT.
Results: If the sheath tube that was introduced was < 8 F, the puncture site could not be closed and would therefore have to be resutured and the Par-CT would have to be repeated. However, the puncture site could be sutured appropriately if the sheath tube that was introduced was ≥ 8F.
Conclusions: This study suggested that the Par-CT is both safe and effective for haemostasis after weaning from VA-ECMO.
期刊介绍:
European Journal of Medical Research publishes translational and clinical research of international interest across all medical disciplines, enabling clinicians and other researchers to learn about developments and innovations within these disciplines and across the boundaries between disciplines. The journal publishes high quality research and reviews and aims to ensure that the results of all well-conducted research are published, regardless of their outcome.