Failure analysis of a fractured stainless steel femoral bone plate

IF 5.7 2区 工程技术 Q1 ENGINEERING, MECHANICAL Engineering Failure Analysis Pub Date : 2025-02-26 DOI:10.1016/j.engfailanal.2025.109466
Luiza de Brito Fantin, Marcelo Ferreira Moreira, José Marcos Paz de Souza
{"title":"Failure analysis of a fractured stainless steel femoral bone plate","authors":"Luiza de Brito Fantin,&nbsp;Marcelo Ferreira Moreira,&nbsp;José Marcos Paz de Souza","doi":"10.1016/j.engfailanal.2025.109466","DOIUrl":null,"url":null,"abstract":"<div><div>This study investigated a femoral bone plate that fractured after 6 months of use in a patient who had already experienced a similar implant failure. Chemical analyses, Vickers hardness measurements, and metallographic and fractographic examinations showed that the compression plate was following the chemical and microstructural requirements set by the ISO 5832 standard, and no material or manufacturing defects could be associated with the premature failure of the plate. The fracture occurred on the seventh of the fourteen holes of the plate and the failure mode was identified as bending fatigue. The main fatigue cracks nucleated on the edge of the compression hole and propagated through the plate towards the bone. Finite elements analysis results confirmed that fatigue crack nucleation sites were regions of maximum stress concentration. Lack of bone healing was the most probable cause for the excessive loading of the plate, and the failure might have been prevented with better stabilization of the bone fracture region, although analyses of surgical and post-operative medical procedures were outside the scope of this investigation. Additionally, a brief discussion regarding orthopedic implant failures in Brazil was conducted. By analyzing 25 other failure cases, it was found that most of them are related to the device itself and might have been prevented through a better technical certification process and implant monitoring by the national health regulatory agency.</div></div>","PeriodicalId":11677,"journal":{"name":"Engineering Failure Analysis","volume":"173 ","pages":"Article 109466"},"PeriodicalIF":5.7000,"publicationDate":"2025-02-26","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Engineering Failure Analysis","FirstCategoryId":"5","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1350630725002079","RegionNum":2,"RegionCategory":"工程技术","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"ENGINEERING, MECHANICAL","Score":null,"Total":0}
引用次数: 0

Abstract

This study investigated a femoral bone plate that fractured after 6 months of use in a patient who had already experienced a similar implant failure. Chemical analyses, Vickers hardness measurements, and metallographic and fractographic examinations showed that the compression plate was following the chemical and microstructural requirements set by the ISO 5832 standard, and no material or manufacturing defects could be associated with the premature failure of the plate. The fracture occurred on the seventh of the fourteen holes of the plate and the failure mode was identified as bending fatigue. The main fatigue cracks nucleated on the edge of the compression hole and propagated through the plate towards the bone. Finite elements analysis results confirmed that fatigue crack nucleation sites were regions of maximum stress concentration. Lack of bone healing was the most probable cause for the excessive loading of the plate, and the failure might have been prevented with better stabilization of the bone fracture region, although analyses of surgical and post-operative medical procedures were outside the scope of this investigation. Additionally, a brief discussion regarding orthopedic implant failures in Brazil was conducted. By analyzing 25 other failure cases, it was found that most of them are related to the device itself and might have been prevented through a better technical certification process and implant monitoring by the national health regulatory agency.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
不锈钢股骨接骨板骨折的失效分析
本研究调查了一例股骨接骨板在使用6个月后骨折的患者,该患者已经经历了类似的植入失败。化学分析、维氏硬度测量、金相和断口学检查表明,压缩板符合ISO 5832标准规定的化学和微观结构要求,没有材料或制造缺陷与板的过早失效有关。板的14个孔中的第7个孔发生断裂,破坏模式为弯曲疲劳。主要疲劳裂纹在压缩孔边缘成核,并通过板向骨方向扩展。有限元分析结果证实,疲劳裂纹成核部位是应力集中最大的区域。骨愈合不足是造成钢板过度负荷的最可能原因,骨折区域的更好稳定可能会避免失败,尽管手术和术后医疗程序的分析超出了本研究的范围。此外,对巴西骨科植入失败进行了简要讨论。通过分析其他25个失败案例,发现其中大多数与设备本身有关,并且可以通过国家卫生监管机构更好的技术认证程序和植入物监控来预防。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Engineering Failure Analysis
Engineering Failure Analysis 工程技术-材料科学:表征与测试
CiteScore
7.70
自引率
20.00%
发文量
956
审稿时长
47 days
期刊介绍: Engineering Failure Analysis publishes research papers describing the analysis of engineering failures and related studies. Papers relating to the structure, properties and behaviour of engineering materials are encouraged, particularly those which also involve the detailed application of materials parameters to problems in engineering structures, components and design. In addition to the area of materials engineering, the interacting fields of mechanical, manufacturing, aeronautical, civil, chemical, corrosion and design engineering are considered relevant. Activity should be directed at analysing engineering failures and carrying out research to help reduce the incidences of failures and to extend the operating horizons of engineering materials. Emphasis is placed on the mechanical properties of materials and their behaviour when influenced by structure, process and environment. Metallic, polymeric, ceramic and natural materials are all included and the application of these materials to real engineering situations should be emphasised. The use of a case-study based approach is also encouraged. Engineering Failure Analysis provides essential reference material and critical feedback into the design process thereby contributing to the prevention of engineering failures in the future. All submissions will be subject to peer review from leading experts in the field.
期刊最新文献
Editorial Board Retraction notice to “Finite element evaluation of seismic response to RC interventions in historical masonry: Erzurum stone Granaries” [Eng. Fail. Anal. 178 (2025) 109727] Rotating bending fatigue analysis of 3D-printed PLA polymer composites: Effect of short fiber reinforcement and epoxy coating Wheel profile optimization for high-speed locomotives to suppress carbody hunting stability failure under complex wheel-rail conditions Guided wave characteristics and nondestructive testing of corrosion defects in threaded steel resin bolts
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1