Measures of knee internal and external rotation made with a digital inclinometer are consistent with the measures made with an electromagnetic tracking system

IF 2.7 Q2 ORTHOPEDICS Journal of Experimental Orthopaedics Pub Date : 2025-03-11 DOI:10.1002/jeo2.70209
Mark K. Timmons, Hunter G. Copley, Dustin T. Darnell, Seth T. Jude, Gary McIlvain
{"title":"Measures of knee internal and external rotation made with a digital inclinometer are consistent with the measures made with an electromagnetic tracking system","authors":"Mark K. Timmons,&nbsp;Hunter G. Copley,&nbsp;Dustin T. Darnell,&nbsp;Seth T. Jude,&nbsp;Gary McIlvain","doi":"10.1002/jeo2.70209","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Purpose</h3>\n \n <p>Individuals with excessive knee rotation have higher levels of dysfunction and disability. An inexpensive, and reliable method to assess knee internal (IR) and external (ER) rotation will improve the assessment of knee injuries. The study explored the reliability and measurement error of two methods of knee rotation measurement. The study tested the hypothesis that the digital inclinometer will provide reliable measures of knee internal and external rotation and that the digital inclinometer measurements will be consistent with the measurements made with the electromagnetic tracking system.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>Twenty (20) participants without knee injury participated. Knee IR and ER were measured using electromagnetic tracking and a digital inclinometer. The intraclass correlation coefficient (ICC<sub>(2,1)</sub>) was calculated for both techniques.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>The within device ICC values ranged from 0.826 to 0.939 for both devices. The within-device minimal detectable change (MDC) ranged from 1.2° to 1.9°. The ICC values for EI and IR measures collapsed between the devices, ranged from 0.717 to 0.859. The MDC calculated between devices ranged from 1.6° to 1.9°.</p>\n </section>\n \n <section>\n \n <h3> Conclusion</h3>\n \n <p>The results of the current study show that knee IR and ER can be measured reliably with both measurement techniques. The measurement of knee ER and IR did not differ between the two devices or between the right and left sides.</p>\n </section>\n \n <section>\n \n <h3> Level of Evidence</h3>\n \n <p>Level IV, diagnostic, case series study.</p>\n </section>\n </div>","PeriodicalId":36909,"journal":{"name":"Journal of Experimental Orthopaedics","volume":"12 1","pages":""},"PeriodicalIF":2.7000,"publicationDate":"2025-03-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/jeo2.70209","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Experimental Orthopaedics","FirstCategoryId":"1085","ListUrlMain":"https://esskajournals.onlinelibrary.wiley.com/doi/10.1002/jeo2.70209","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose

Individuals with excessive knee rotation have higher levels of dysfunction and disability. An inexpensive, and reliable method to assess knee internal (IR) and external (ER) rotation will improve the assessment of knee injuries. The study explored the reliability and measurement error of two methods of knee rotation measurement. The study tested the hypothesis that the digital inclinometer will provide reliable measures of knee internal and external rotation and that the digital inclinometer measurements will be consistent with the measurements made with the electromagnetic tracking system.

Methods

Twenty (20) participants without knee injury participated. Knee IR and ER were measured using electromagnetic tracking and a digital inclinometer. The intraclass correlation coefficient (ICC(2,1)) was calculated for both techniques.

Results

The within device ICC values ranged from 0.826 to 0.939 for both devices. The within-device minimal detectable change (MDC) ranged from 1.2° to 1.9°. The ICC values for EI and IR measures collapsed between the devices, ranged from 0.717 to 0.859. The MDC calculated between devices ranged from 1.6° to 1.9°.

Conclusion

The results of the current study show that knee IR and ER can be measured reliably with both measurement techniques. The measurement of knee ER and IR did not differ between the two devices or between the right and left sides.

Level of Evidence

Level IV, diagnostic, case series study.

Abstract Image

Abstract Image

Abstract Image

Abstract Image

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
用数字倾角仪测量的膝关节内外旋转与电磁跟踪系统测量的结果一致
目的:过度膝关节旋转的个体有更高水平的功能障碍和残疾。一种廉价、可靠的评估膝关节内(IR)和外(ER)旋转的方法将改善对膝关节损伤的评估。研究了两种膝关节旋转测量方法的可靠性和测量误差。该研究验证了以下假设:数字倾斜仪将提供可靠的膝关节内外旋转测量,并且数字倾斜仪的测量结果将与电磁跟踪系统的测量结果一致。方法20例无膝关节损伤的患者参与研究。使用电磁跟踪和数字倾角仪测量膝关节IR和ER。计算了两种技术的类内相关系数(ICC(2,1))。结果两种设备的设备内ICC值在0.826 ~ 0.939之间。设备内最小可检测变化(MDC)范围为1.2°至1.9°。EI和IR测量的ICC值在设备之间崩溃,范围从0.717到0.859。设备间的MDC取值范围为1.6°~ 1.9°。结论本研究结果表明,两种测量技术均可可靠地测量膝关节IR和ER。膝关节ER和IR的测量在两种装置之间或左右两侧之间没有差异。证据水平四级,诊断性,病例系列研究。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Journal of Experimental Orthopaedics
Journal of Experimental Orthopaedics Medicine-Orthopedics and Sports Medicine
CiteScore
3.20
自引率
5.60%
发文量
114
审稿时长
13 weeks
期刊最新文献
Posterior condylar offset and clinical outcomes in medial pivot total knee arthroplasty: A comparison of mechanical and kinematic alignment. From de-skilling to up-skilling: How artificial intelligence will augment the modern physician. Movement smoothness matters: The missing piece in the functional assessment of rotator cuff patients. Insights from a stereophotogrammetry experimental study. Implant selection and causes of aseptic failure in conversion from unicompartmental to total knee arthroplasty: A high-volume centre series. Knee kinesiography in 2026: Current and future applications for digital orthopaedics and personalised total knee arthroplasty: A narrative review.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1