{"title":"Visual performance in pseudophakia with standard and ultraviolet-absorbing intraocular lenses: a preliminary report.","authors":"H M Hammer, M Yap, J R Weatherill","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>There is currently considerably interest in the vulnerability of the pseudophakic eye ultraviolet light. Contrast sensitivity was measured in patients receiving UV-absorbing and standard posterior chamber intraocular lenses (IOLs) and in normal age-matched observers. The results suggest that in the relative short-term there is no difference in visual performance with either type of IOL. Modulation Transfer Function studies on 3 batches of IOLs (2 standard and 1 UV-absorbing) showed slight variations in optical performance within and between batches. Spectral sensitivity testing demonstrated increased macular protection from short wave-length blue light by the crystalline lens when compared with both types of IOL.</p>","PeriodicalId":76757,"journal":{"name":"Transactions of the ophthalmological societies of the United Kingdom","volume":"105 ( Pt 4) ","pages":"441-6"},"PeriodicalIF":0.0000,"publicationDate":"1986-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Transactions of the ophthalmological societies of the United Kingdom","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
There is currently considerably interest in the vulnerability of the pseudophakic eye ultraviolet light. Contrast sensitivity was measured in patients receiving UV-absorbing and standard posterior chamber intraocular lenses (IOLs) and in normal age-matched observers. The results suggest that in the relative short-term there is no difference in visual performance with either type of IOL. Modulation Transfer Function studies on 3 batches of IOLs (2 standard and 1 UV-absorbing) showed slight variations in optical performance within and between batches. Spectral sensitivity testing demonstrated increased macular protection from short wave-length blue light by the crystalline lens when compared with both types of IOL.