{"title":"On the claibration of laser-cone calorimeters.","authors":"F Hillenkamp","doi":"10.1088/0022-3735/1/10/413","DOIUrl":null,"url":null,"abstract":"A method is described which allows the experimental determination of the heat capacity of laser-cone calorimeters. The method includes the usual calibration with a current pulse applied to the cone's heating coil, and a second calibration with d.c. current. Uncertainties in the calibration of such calorimeters due to possible incomplete heat transfer between the coil, the cone and the temperature sensors are thereby eliminated.","PeriodicalId":16983,"journal":{"name":"Journal of Scientific Instruments","volume":"1 10","pages":"1022-3"},"PeriodicalIF":0.0000,"publicationDate":"1968-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1088/0022-3735/1/10/413","citationCount":"2","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Scientific Instruments","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1088/0022-3735/1/10/413","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 2
Abstract
A method is described which allows the experimental determination of the heat capacity of laser-cone calorimeters. The method includes the usual calibration with a current pulse applied to the cone's heating coil, and a second calibration with d.c. current. Uncertainties in the calibration of such calorimeters due to possible incomplete heat transfer between the coil, the cone and the temperature sensors are thereby eliminated.