Initial experience with a steerable intravascular ultrasound catheter in the aorta and pulmonary artery.

American journal of cardiac imaging Pub Date : 1995-07-01
G Görge, J Ge, M Haude, D Baumgart, T Buck, R Erbel
{"title":"Initial experience with a steerable intravascular ultrasound catheter in the aorta and pulmonary artery.","authors":"G Görge,&nbsp;J Ge,&nbsp;M Haude,&nbsp;D Baumgart,&nbsp;T Buck,&nbsp;R Erbel","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>The aim of this protocol was to test the feasibility and safety of a prototype steerable intravascular ultrasound (IVUS) catheter (Boston Scientific, Waterton, MA) in comparison with standard IVUS catheters. A 3.5F, 20-MHz mechanical echo transducer was incorporated into a bendable sheath with a blunt tip. The flexible IVUS catheter was compared with a standard IVUS catheter in 13 patients. Seven patients underwent catheterization of the left side of the heart, and six patients had catheterization of the right side of the heart for suspected recurrent pulmonary embolism. In the aorta, three lumen area measurements were made: (1) midway between the aortic arch and the aortic root, (2) at the most cranial part of the aorta, and (3) in the descending aorta at the level of the diaphragm. Evaluation of the accuracy of luminal dimension measurements by both types of catheters in perpendicular positions to the vessel wall was evaluated in a hollow rubber cast of an human aorta and its side branches, representing luminal diameters from 3 to 26 mm. We performed 20 measurements with each type of catheter. The results were compared with ruler measurements, after the cast had been cut in slices. The equation for the standard 3.5F IVUS catheter was: y = 0.89x + 0.15; SE = 0.17; r = .97; for the 4.8F 20-MHz standard IVUS catheter: y = 0.97x + 0.05; SE = 0.18; r = .98; and for the steerable catheter, y = 0.94x + 0.09; SE = 0.12; r = 0.97.(ABSTRACT TRUNCATED AT 250 WORDS)</p>","PeriodicalId":79315,"journal":{"name":"American journal of cardiac imaging","volume":"9 3","pages":"180-4"},"PeriodicalIF":0.0000,"publicationDate":"1995-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"American journal of cardiac imaging","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

The aim of this protocol was to test the feasibility and safety of a prototype steerable intravascular ultrasound (IVUS) catheter (Boston Scientific, Waterton, MA) in comparison with standard IVUS catheters. A 3.5F, 20-MHz mechanical echo transducer was incorporated into a bendable sheath with a blunt tip. The flexible IVUS catheter was compared with a standard IVUS catheter in 13 patients. Seven patients underwent catheterization of the left side of the heart, and six patients had catheterization of the right side of the heart for suspected recurrent pulmonary embolism. In the aorta, three lumen area measurements were made: (1) midway between the aortic arch and the aortic root, (2) at the most cranial part of the aorta, and (3) in the descending aorta at the level of the diaphragm. Evaluation of the accuracy of luminal dimension measurements by both types of catheters in perpendicular positions to the vessel wall was evaluated in a hollow rubber cast of an human aorta and its side branches, representing luminal diameters from 3 to 26 mm. We performed 20 measurements with each type of catheter. The results were compared with ruler measurements, after the cast had been cut in slices. The equation for the standard 3.5F IVUS catheter was: y = 0.89x + 0.15; SE = 0.17; r = .97; for the 4.8F 20-MHz standard IVUS catheter: y = 0.97x + 0.05; SE = 0.18; r = .98; and for the steerable catheter, y = 0.94x + 0.09; SE = 0.12; r = 0.97.(ABSTRACT TRUNCATED AT 250 WORDS)

分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
在主动脉和肺动脉植入可操控的血管内超声导管的初步经验。
该方案的目的是测试原型可操纵血管内超声(IVUS)导管(Boston Scientific, Waterton, MA)与标准IVUS导管的可行性和安全性。一个3.5F, 20 mhz的机械回波传感器被整合到一个钝尖端的可弯曲护套中。13例患者将柔性IVUS导管与标准IVUS导管进行比较。7例患者因疑似复发性肺栓塞而行左心置管,6例患者行右心置管。在主动脉中,测量三个管腔面积:(1)在主动脉弓和主动脉根之间,(2)在主动脉最颅部,(3)在降主动脉横膈膜水平。评估两种导管垂直于血管壁位置的管腔尺寸测量的准确性,在人主动脉及其侧分支的中空橡胶铸型中进行评估,代表管腔直径从3到26毫米。我们对每种类型的导管进行了20次测量。将铸件切成薄片后,将结果与直尺测量结果进行比较。标准3.5F IVUS导管的方程为:y = 0.89x + 0.15;Se = 0.17;R = 0.97;4.8F 20 mhz标准IVUS导管:y = 0.97x + 0.05;Se = 0.18;R = 0.98;对于导向导管,y = 0.94x + 0.09;Se = 0.12;R = 0.97。(摘要删节250字)
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Variability of a three-layered appearance in intravascular ultrasound coronary images: a comparison of morphometric measurements with four intravascular ultrasound systems. Contrast-enhanced radiographic computed tomographic findings in patients with straight back syndrome. Comparison of 10, 20, and 40 level electron beam computed tomography studies for coronary calcium. Variation of heart rate and electrocardiograph trigger interval during ultrafast computed tomography. Evaluation of reference systems for quantitative wall motion analysis from three-dimensional endocardial surface reconstruction: an echocardiographic study in subjects with and without myocardial infarction.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1