{"title":"FIx controller: an instrument to automatically adjust inspired oxygen fraction using feedback control from a pulse oximeter.","authors":"D B Raemer, X B Ji, G P Topulos","doi":"10.1023/a:1007384122804","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To develop an instrument to help prevent pulmonary O2 toxicity, a syndrome that manifests itself in adult intensive care patients.</p><p><strong>Methods: </strong>We designed, built, and tested a device that controls FIO2 exposure using oxygen saturation measured with a pulse oximeter (SpO2) in a negative feedback control system. A target SpO2 is designated by the clinician and the system adjusts the FIO2 from a mechanical ventilator so as to minimize the difference between the measured SpO2 and the target. Important elements of the system include a conservative artifact rejection algorithm, a gainscheduled sampled-data proportional-integral-derivative (PID) controller, and a safety system to prevent inspired mixtures with undesirably low FIO2 due to device failure.</p><p><strong>Results: </strong>The control system was tuned in a series of animal experiments. Acceptable clinical response of the system was obtained using a gain-scheduled controller algorithm whereby the gain of the proportional term of a PID controller was adjusted based on the error signal and measured minute ventilation. Also, the artifact rejection algorithm and safety systems were successfully tested using simulation.</p><p><strong>Conclusions: </strong>Testing the effectiveness of this instrument will require comparison with manual control of FIO2 in an appropriately designed trial.</p>","PeriodicalId":77199,"journal":{"name":"Journal of clinical monitoring","volume":"13 2","pages":"91-101"},"PeriodicalIF":0.0000,"publicationDate":"1997-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1023/a:1007384122804","citationCount":"19","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of clinical monitoring","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1023/a:1007384122804","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 19
Abstract
Objective: To develop an instrument to help prevent pulmonary O2 toxicity, a syndrome that manifests itself in adult intensive care patients.
Methods: We designed, built, and tested a device that controls FIO2 exposure using oxygen saturation measured with a pulse oximeter (SpO2) in a negative feedback control system. A target SpO2 is designated by the clinician and the system adjusts the FIO2 from a mechanical ventilator so as to minimize the difference between the measured SpO2 and the target. Important elements of the system include a conservative artifact rejection algorithm, a gainscheduled sampled-data proportional-integral-derivative (PID) controller, and a safety system to prevent inspired mixtures with undesirably low FIO2 due to device failure.
Results: The control system was tuned in a series of animal experiments. Acceptable clinical response of the system was obtained using a gain-scheduled controller algorithm whereby the gain of the proportional term of a PID controller was adjusted based on the error signal and measured minute ventilation. Also, the artifact rejection algorithm and safety systems were successfully tested using simulation.
Conclusions: Testing the effectiveness of this instrument will require comparison with manual control of FIO2 in an appropriately designed trial.