Nurse and Pharmacist Knowledge of Intravenous Smart Pump System Setup Requirements.

Karen K Giuliano, Jeannine W C Blake
{"title":"Nurse and Pharmacist Knowledge of Intravenous Smart Pump System Setup Requirements.","authors":"Karen K Giuliano, Jeannine W C Blake","doi":"10.2345/0899-8205-55.1.51","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>The primary purpose of this research was to describe nurse and pharmacist knowledge of setup requirements for intravenous (IV) smart pumps that require head height differentials for accurate fluid flow.</p><p><strong>Methods: </strong>A secondary analysis of anonymous electronic survey data using a database of prerecruited clinicians was conducted. A survey was sent by email to 173 pharmacists and 960 nurses. The response rate for pharmacists was 58% (100 of 173), and the response rate for nurses was 52% (500 of 960). After removing respondents who did not provide direct care and who did not use a head height differential IV infusion system, the final sample for analysis was 186 nurses and 25 pharmacists.</p><p><strong>Results: </strong>Overall, less than one-half of respondents (40%) were aware that manufacturer guidelines for positioning the primary infusion bag relative to the infusion pump were available. Slightly more (49.5%) were aware of the required head height differentials for secondary infusion. Only five respondents selected the correct primary head height, eight respondents selected the correct secondary head height, and one respondent selected both the correct primary and secondary head heights.</p><p><strong>Conclusion: </strong>The results of this study identify a substantial lack of knowledge among frontline clinicians regarding manufacturer recommendations for accurate IV administration of primary and secondary infusions for head height differential infusion systems. Both increased clinician education and innovative technology solutions are needed to improve IV smart pump safety and usability.</p>","PeriodicalId":35656,"journal":{"name":"Biomedical Instrumentation and Technology","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2021-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8641421/pdf/i0899-8205-55-1-51.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biomedical Instrumentation and Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2345/0899-8205-55.1.51","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: The primary purpose of this research was to describe nurse and pharmacist knowledge of setup requirements for intravenous (IV) smart pumps that require head height differentials for accurate fluid flow.

Methods: A secondary analysis of anonymous electronic survey data using a database of prerecruited clinicians was conducted. A survey was sent by email to 173 pharmacists and 960 nurses. The response rate for pharmacists was 58% (100 of 173), and the response rate for nurses was 52% (500 of 960). After removing respondents who did not provide direct care and who did not use a head height differential IV infusion system, the final sample for analysis was 186 nurses and 25 pharmacists.

Results: Overall, less than one-half of respondents (40%) were aware that manufacturer guidelines for positioning the primary infusion bag relative to the infusion pump were available. Slightly more (49.5%) were aware of the required head height differentials for secondary infusion. Only five respondents selected the correct primary head height, eight respondents selected the correct secondary head height, and one respondent selected both the correct primary and secondary head heights.

Conclusion: The results of this study identify a substantial lack of knowledge among frontline clinicians regarding manufacturer recommendations for accurate IV administration of primary and secondary infusions for head height differential infusion systems. Both increased clinician education and innovative technology solutions are needed to improve IV smart pump safety and usability.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
护士和药剂师了解静脉智能泵系统的设置要求。
研究目的本研究的主要目的是描述护士和药剂师对静脉注射(IV)智能泵的设置要求的了解情况,这些泵需要头高差来实现准确的液体流动:方法: 使用预先招募的临床医生数据库对匿名电子调查数据进行二次分析。通过电子邮件向 173 名药剂师和 960 名护士发送了调查问卷。药剂师的回复率为 58%(173 人中有 100 人回复),护士的回复率为 52%(960 人中有 500 人回复)。在剔除不提供直接护理和不使用头部高度差静脉输液系统的受访者后,最终的分析样本为 186 名护士和 25 名药剂师:总体而言,不到二分之一的受访者(40%)知道生产商提供了主输液袋相对于输液泵的定位指南。了解二次输液所需的头部高度差的受访者略多(49.5%)。只有 5 名受访者选择了正确的主输液头高度,8 名受访者选择了正确的副输液头高度,1 名受访者同时选择了正确的主输液头和副输液头高度:这项研究的结果表明,一线临床医生对制造商关于头部高度差输液系统的主要和次要输液的准确静脉输注建议缺乏了解。为提高静脉智能泵的安全性和可用性,需要加强对临床医生的教育并采用创新技术解决方案。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Biomedical Instrumentation and Technology
Biomedical Instrumentation and Technology Computer Science-Computer Networks and Communications
CiteScore
1.10
自引率
0.00%
发文量
16
期刊介绍: AAMI publishes Biomedical Instrumentation & Technology (BI&T) a bi-monthly peer-reviewed journal dedicated to the developers, managers, and users of medical instrumentation and technology.
期刊最新文献
Effect of Gamma and X-ray Irradiation on Polymers Commonly Used in Healthcare Products. Effect of Vaporized Hydrogen Peroxide and Nitrogen Dioxide Sterilization on Nitinol. Toward Enhanced Machine-Based Release in X-ray Sterilization. Beyond Endoscopes: Pilot Study of Surgical Instrument Lumen Inspection. Advantages and Limitations of Physical and Virtual Dose Mapping.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1