T Kanluan, S Tangvarasittichai, O Tangvarasittichai
{"title":"Evaluation of a random access analyser: BM/Hitachi 911.","authors":"T Kanluan, S Tangvarasittichai, O Tangvarasittichai","doi":"10.1155/S1463924695000332","DOIUrl":null,"url":null,"abstract":"<p><p>The performance of Boehringer Mannheim's BM/Hitachi 911 was evaluated for three months. The mean coeffcient of variation (CV) of the within-run and between-run imprecision of the 16 analytes were less than 1.16% (range 0.47-2.38%) and 1.35% (range 0.62-2.93,%), respectively. A linearity study for the various assays covered clinically important levels. No relevant drift was observed during an eight-hour assay nor was any sample-related carry-over detected. In all cases, the regression analyses (slopes) of the results obtainedfrom BM/Hitachi 911 and 717 were between the extreme values of 0.94 and 1.05. During the three months of operation, no major problem was encountered. The BM/Hitachi 911 was found to be easily operated, to require minimal attention and simple daily maintenance during operation.</p>","PeriodicalId":22600,"journal":{"name":"The Journal of Automatic Chemistry","volume":"17 6","pages":"213-8"},"PeriodicalIF":0.0000,"publicationDate":"1995-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1155/S1463924695000332","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"The Journal of Automatic Chemistry","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1155/S1463924695000332","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
The performance of Boehringer Mannheim's BM/Hitachi 911 was evaluated for three months. The mean coeffcient of variation (CV) of the within-run and between-run imprecision of the 16 analytes were less than 1.16% (range 0.47-2.38%) and 1.35% (range 0.62-2.93,%), respectively. A linearity study for the various assays covered clinically important levels. No relevant drift was observed during an eight-hour assay nor was any sample-related carry-over detected. In all cases, the regression analyses (slopes) of the results obtainedfrom BM/Hitachi 911 and 717 were between the extreme values of 0.94 and 1.05. During the three months of operation, no major problem was encountered. The BM/Hitachi 911 was found to be easily operated, to require minimal attention and simple daily maintenance during operation.