Mini drug pump for ophthalmic use.

Saloomeh Saati, Ronalee Lo, Po-Ying Li, Ellis Meng, Rohit Varma, Mark S Humayun
{"title":"Mini drug pump for ophthalmic use.","authors":"Saloomeh Saati,&nbsp;Ronalee Lo,&nbsp;Po-Ying Li,&nbsp;Ellis Meng,&nbsp;Rohit Varma,&nbsp;Mark S Humayun","doi":"","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To evaluate the feasibility of developing a novel mini drug pump for ophthalmic use.</p><p><strong>Methods: </strong>Using principles of microelectromechanical systems engineering, a mini drug pump was fabricated. The pumping mechanism is based on electrolysis, and the pump includes a drug refill port as well as a check valve to control drug delivery. Drug pumps were tested first on the benchtop and then after implantation in rabbits. For the latter, we implanted 4 elliptical (9.9 x 7.7 x 1.8 mm) non-electrically active pumps into 4 rabbits. The procedure is similar to implantation of a glaucoma seton. To determine the ability to refill and also the patency of the cannula, at intervals of 4 to 6 weeks after implantation, we accessed the drug reservoir with a transconjunctival needle and delivered approximately as low as 1 microL of trypan blue solution (0.06%) into the anterior chamber. Animals were followed up by slit-lamp examination, photography, and fluorescein angiography.</p><p><strong>Results: </strong>Benchtop testing showed 2.0 microL/min delivery when using 0.4 mW of power for electrolysis. One-way valves showed reliable opening pressures of 470 mm Hg. All implanted devices refilled at 4- to 6-week intervals for 4 to 6 months. No infection was seen. No devices extruded. No filtering bleb formed over the implant.</p><p><strong>Conclusions: </strong>A prototype ocular mini drug pump was built, implanted, and refilled. Such a platform needs more testing to determine the long-term biocompatibility of an electrically controlled implanted pump. Testing with various pharmacologic agents is needed to determine its ultimate potential for ophthalmic use.</p>","PeriodicalId":23166,"journal":{"name":"Transactions of the American Ophthalmological Society","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2009-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2814560/pdf/1545-6110_v107_p060.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Transactions of the American Ophthalmological Society","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: To evaluate the feasibility of developing a novel mini drug pump for ophthalmic use.

Methods: Using principles of microelectromechanical systems engineering, a mini drug pump was fabricated. The pumping mechanism is based on electrolysis, and the pump includes a drug refill port as well as a check valve to control drug delivery. Drug pumps were tested first on the benchtop and then after implantation in rabbits. For the latter, we implanted 4 elliptical (9.9 x 7.7 x 1.8 mm) non-electrically active pumps into 4 rabbits. The procedure is similar to implantation of a glaucoma seton. To determine the ability to refill and also the patency of the cannula, at intervals of 4 to 6 weeks after implantation, we accessed the drug reservoir with a transconjunctival needle and delivered approximately as low as 1 microL of trypan blue solution (0.06%) into the anterior chamber. Animals were followed up by slit-lamp examination, photography, and fluorescein angiography.

Results: Benchtop testing showed 2.0 microL/min delivery when using 0.4 mW of power for electrolysis. One-way valves showed reliable opening pressures of 470 mm Hg. All implanted devices refilled at 4- to 6-week intervals for 4 to 6 months. No infection was seen. No devices extruded. No filtering bleb formed over the implant.

Conclusions: A prototype ocular mini drug pump was built, implanted, and refilled. Such a platform needs more testing to determine the long-term biocompatibility of an electrically controlled implanted pump. Testing with various pharmacologic agents is needed to determine its ultimate potential for ophthalmic use.

Abstract Image

Abstract Image

Abstract Image

分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
眼科用迷你药物泵。
目的:评价研制一种新型眼科微型药物泵的可行性。方法:利用微机电系统工程原理,制作微型药物泵。所述泵送机构基于电解,所述泵包括药物补给口以及控制药物输送的止回阀。药物泵首先在实验台上进行测试,然后在植入兔子体内后进行测试。对于后者,我们将4个椭圆形(9.9 x 7.7 x 1.8 mm)非电活性泵植入4只兔子体内。手术过程类似于青光眼植入。为了确定重新填充的能力和插管的通畅程度,在植入后4至6周的间隔时间内,我们使用经结膜针进入药物储存库,并向前房注入大约低至1微升的台泮蓝溶液(0.06%)。动物随访采用裂隙灯检查、摄影和荧光素血管造影。结果:台式测试显示,电解功率为0.4 mW时,输出量为2.0 microL/min。单向阀显示可靠的开启压力为470毫米汞柱。所有植入的装置每隔4至6周重新填充一次,持续4至6个月。未见感染。没有挤压设备。植入物上方未形成滤泡。结论:建立了一个眼部微型药物泵的原型,植入并重新填充。这样的平台需要更多的测试来确定电控植入泵的长期生物相容性。需要用各种药理学试剂进行试验,以确定其在眼科应用的最终潜力。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
The Utility of Color Duplex Ultrasonography in the Diagnosis of Giant Cell Arteritis: A Prospective, Masked Study. (An American Ophthalmological Society Thesis). Autoimmune Retinopathy: Current Concepts and Practices (An American Ophthalmological Society Thesis). The Relationship Between Ocular Itch, Ocular Pain, and Dry Eye Symptoms (An American Ophthalmological Society Thesis). Erratum: Predictors of Intraocular Pressure After Phacoemulsification in Primary Open-Angle Glaucoma Eyes with Wide Versus Narrower Angles (An American Ophthalmological Society Thesis). The Effects of Phacoemulsification and Intraocular Lens Implantation on Anatomical and Functional Parameters in Patients with Primary Angle Closure: A Prospective Study. (An American Ophthalmological Society Thesis).
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1