A handheld open-field infant keratometer (an american ophthalmological society thesis).

Joseph M Miller
{"title":"A handheld open-field infant keratometer (an american ophthalmological society thesis).","authors":"Joseph M Miller","doi":"","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To design and evaluate a new infant keratometer that incorporates an unobstructed view of the infant with both eyes (open-field design).</p><p><strong>Methods: </strong>The design of the open-field infant keratometer is presented, and details of its construction are given. The design incorporates a single-ring keratoscope for measurement of corneal astigmatism over a 4-mm region of the cornea and includes a rectangular grid target concentric within the ring to allow for the study of higher-order aberrations of the eye. In order to calibrate the lens and imaging system, a novel telecentric test object was constructed and used. The system was bench calibrated against steel ball bearings of known dimensions and evaluated for accuracy while being used in handheld mode in a group of 16 adult cooperative subjects. It was then evaluated for testability in a group of 10 infants and toddlers.</p><p><strong>Results: </strong>Results indicate that while the device achieved the goal of creating an open-field instrument containing a single-ring keratoscope with a concentric grid array for the study of higher-order aberrations, additional work is required to establish better control of the vertex distance.</p><p><strong>Conclusion: </strong>The handheld open-field infant keratometer demonstrates testability suitable for the study of infant corneal astigmatism. Use of collimated light sources in future iterations of the design must be incorporated in order to achieve the accuracy required for clinical investigation.</p>","PeriodicalId":23166,"journal":{"name":"Transactions of the American Ophthalmological Society","volume":"108 ","pages":"77-95"},"PeriodicalIF":0.0000,"publicationDate":"2010-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3016080/pdf/1545-6110_v108_p077.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Transactions of the American Ophthalmological Society","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: To design and evaluate a new infant keratometer that incorporates an unobstructed view of the infant with both eyes (open-field design).

Methods: The design of the open-field infant keratometer is presented, and details of its construction are given. The design incorporates a single-ring keratoscope for measurement of corneal astigmatism over a 4-mm region of the cornea and includes a rectangular grid target concentric within the ring to allow for the study of higher-order aberrations of the eye. In order to calibrate the lens and imaging system, a novel telecentric test object was constructed and used. The system was bench calibrated against steel ball bearings of known dimensions and evaluated for accuracy while being used in handheld mode in a group of 16 adult cooperative subjects. It was then evaluated for testability in a group of 10 infants and toddlers.

Results: Results indicate that while the device achieved the goal of creating an open-field instrument containing a single-ring keratoscope with a concentric grid array for the study of higher-order aberrations, additional work is required to establish better control of the vertex distance.

Conclusion: The handheld open-field infant keratometer demonstrates testability suitable for the study of infant corneal astigmatism. Use of collimated light sources in future iterations of the design must be incorporated in order to achieve the accuracy required for clinical investigation.

Abstract Image

分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
手持式开视野婴儿角膜计(美国眼科学会论文)。
目的:设计并评估一种新的婴儿角膜计,该角膜计能让婴儿双眼不受阻碍地观察(开视场设计):方法:介绍开视野婴儿角膜计的设计及其构造细节。该设计包含一个单环角膜塑形镜,用于测量角膜上 4 毫米区域的角膜散光,还包括一个同心环内的矩形网格目标,以便研究眼球的高阶像差。为了校准镜头和成像系统,我们制作并使用了一个新颖的远心测试物体。该系统根据已知尺寸的钢珠轴承进行了台架校准,并在一组 16 名成年合作受试者中以手持模式使用时对其准确性进行了评估。然后在一组 10 名婴幼儿中对其可测试性进行了评估:结果表明,虽然该设备实现了创建一个包含单环角膜塑形镜和同心网格阵列的开视野仪器以研究高阶像差的目标,但还需要做更多的工作来更好地控制顶点距离:结论:手持式开视野婴儿角膜塑形镜证明了其适用于婴儿角膜散光研究的可测试性。在未来的迭代设计中,必须使用准直光源,以达到临床研究所需的精确度。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
The Utility of Color Duplex Ultrasonography in the Diagnosis of Giant Cell Arteritis: A Prospective, Masked Study. (An American Ophthalmological Society Thesis). Autoimmune Retinopathy: Current Concepts and Practices (An American Ophthalmological Society Thesis). The Relationship Between Ocular Itch, Ocular Pain, and Dry Eye Symptoms (An American Ophthalmological Society Thesis). Erratum: Predictors of Intraocular Pressure After Phacoemulsification in Primary Open-Angle Glaucoma Eyes with Wide Versus Narrower Angles (An American Ophthalmological Society Thesis). The Effects of Phacoemulsification and Intraocular Lens Implantation on Anatomical and Functional Parameters in Patients with Primary Angle Closure: A Prospective Study. (An American Ophthalmological Society Thesis).
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1