Analytical performance of a real-time PCR-based assay for V600 mutations in the BRAF gene, used as the companion diagnostic test for the novel BRAF inhibitor vemurafenib in metastatic melanoma.

Harkanwal Halait, Kelli Demartin, Sweta Shah, Stephen Soviero, Rachel Langland, Suzanne Cheng, Grantland Hillman, Lin Wu, H Jeffrey Lawrence
{"title":"Analytical performance of a real-time PCR-based assay for V600 mutations in the BRAF gene, used as the companion diagnostic test for the novel BRAF inhibitor vemurafenib in metastatic melanoma.","authors":"Harkanwal Halait,&nbsp;Kelli Demartin,&nbsp;Sweta Shah,&nbsp;Stephen Soviero,&nbsp;Rachel Langland,&nbsp;Suzanne Cheng,&nbsp;Grantland Hillman,&nbsp;Lin Wu,&nbsp;H Jeffrey Lawrence","doi":"10.1097/PDM.0b013e31823b216f","DOIUrl":null,"url":null,"abstract":"<p><p>Melanomas frequently harbor BRAFV600 mutations. Vemurafenib (RG7204/PLX4032), a small-molecule inhibitor of mutant BRAF, has shown striking clinical efficacy in BRAFV600 mutant melanoma, creating the need for a well-validated companion diagnostic to select patients for treatment. We describe analytic performance characteristics of the cobas 4800 BRAF V600 Mutation Test, the test used to select patients for the pivotal vemurafenib trials. This real-time polymerase chain reaction assay was designed to detect the V600E (1799T>A) mutation DNA from formalin-fixed paraffin-embedded tissue samples. Sensitivity was assessed using blends of cell lines or tumor DNA, and tumor specimens with low levels of mutant alleles, as determined by 454 sequencing (a quantitative next-generation pyrosequencing method). A >96% hit rate was obtained across all specimen types with 5% mutant alleles at a DNA input of 125 ng, an amount readily obtained from one 5-μm section. The cobas test showed a higher sensitivity and specificity than direct bidirectional sequencing in a panel of 219 melanoma specimens. Cross reactivity with V600K and V600D was observed. Repeated testing of 5 specimens by 2 operators, using different instruments and reagent lots, yielded correct calls in 158/160 tests (98.8%). A set of 26 highly pigmented samples were identified that gave invalid test results. A simple 1:2 dilution resulted in a valid test result of 76% in such cases. The cobas test is a reproducible assay that detects some non-V600E mutations and is more accurate than direct sequencing in detecting BRAFV600E.</p>","PeriodicalId":11235,"journal":{"name":"Diagnostic Molecular Pathology","volume":"21 1","pages":"1-8"},"PeriodicalIF":0.0000,"publicationDate":"2012-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1097/PDM.0b013e31823b216f","citationCount":"152","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Diagnostic Molecular Pathology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1097/PDM.0b013e31823b216f","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 152

Abstract

Melanomas frequently harbor BRAFV600 mutations. Vemurafenib (RG7204/PLX4032), a small-molecule inhibitor of mutant BRAF, has shown striking clinical efficacy in BRAFV600 mutant melanoma, creating the need for a well-validated companion diagnostic to select patients for treatment. We describe analytic performance characteristics of the cobas 4800 BRAF V600 Mutation Test, the test used to select patients for the pivotal vemurafenib trials. This real-time polymerase chain reaction assay was designed to detect the V600E (1799T>A) mutation DNA from formalin-fixed paraffin-embedded tissue samples. Sensitivity was assessed using blends of cell lines or tumor DNA, and tumor specimens with low levels of mutant alleles, as determined by 454 sequencing (a quantitative next-generation pyrosequencing method). A >96% hit rate was obtained across all specimen types with 5% mutant alleles at a DNA input of 125 ng, an amount readily obtained from one 5-μm section. The cobas test showed a higher sensitivity and specificity than direct bidirectional sequencing in a panel of 219 melanoma specimens. Cross reactivity with V600K and V600D was observed. Repeated testing of 5 specimens by 2 operators, using different instruments and reagent lots, yielded correct calls in 158/160 tests (98.8%). A set of 26 highly pigmented samples were identified that gave invalid test results. A simple 1:2 dilution resulted in a valid test result of 76% in such cases. The cobas test is a reproducible assay that detects some non-V600E mutations and is more accurate than direct sequencing in detecting BRAFV600E.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
基于实时pcr的BRAF基因V600突变检测的分析性能,用于新型BRAF抑制剂vemurafenib在转移性黑色素瘤中的伴随诊断测试。
黑色素瘤经常携带BRAFV600突变。Vemurafenib (RG7204/PLX4032)是一种突变BRAF的小分子抑制剂,在BRAFV600突变黑色素瘤中显示出惊人的临床疗效,这就需要一种经过充分验证的伴随诊断来选择治疗患者。我们描述了cobas 4800 BRAF V600突变测试的分析性能特征,该测试用于选择关键vemurafenib试验的患者。采用实时聚合酶链反应法检测福尔马林固定石蜡包埋组织样品中V600E (1799T>A)突变DNA。通过454测序(一种定量下一代焦磷酸测序方法)确定的细胞系或肿瘤DNA混合物和具有低水平突变等位基因的肿瘤标本的敏感性。当DNA输入为125 ng时,在所有类型的样品中,5%的突变等位基因的命中率都大于96%,这一数字很容易从一个5-μm切片中获得。cobas测试在219个黑色素瘤样本中显示出比直接双向测序更高的敏感性和特异性。观察到与V600K和V600D的交叉反应。2名操作人员使用不同仪器和试剂批号对5个标本进行重复检测,158/160次检测结果正确(98.8%)。一组26个高度着色的样品被鉴定出给出无效的测试结果。在这种情况下,简单的1:2稀释可获得76%的有效测试结果。cobas试验是一种可重复的试验,可检测一些非v600e突变,在检测BRAFV600E方面比直接测序更准确。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
审稿时长
>12 weeks
期刊介绍: Diagnostic Molecular Pathology focuses on providing clinical and academic pathologists with coverage of the latest molecular technologies, timely reviews of established techniques, and papers on the applications of these methods to all aspects of surgical pathology and laboratory medicine. It publishes original, peer-reviewed contributions on molecular probes for diagnosis, such as tumor suppressor genes, oncogenes, the polymerase chain reaction (PCR), and in situ hybridization. Articles demonstrate how these highly sensitive techniques can be applied for more accurate diagnosis.
期刊最新文献
Index Molecular Testing for Respiratory Viruses Molecular Testing in Emerging Infectious Diseases Frequent PIK3CA mutations in radial scars. Pyrosequencing for EGFR mutation detection: diagnostic accuracy and clinical implications.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1