Berkan Guleyupoglu, Nicole Febles, Preet Minhas, Christoph Hahn, Marom Bikson
{"title":"Reduced discomfort during high-definition transcutaneous stimulation using 6% benzocaine.","authors":"Berkan Guleyupoglu, Nicole Febles, Preet Minhas, Christoph Hahn, Marom Bikson","doi":"10.3389/fneng.2014.00028","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>High-Definition transcranial Direct Current Stimulation (HD-tDCS) allows for non-invasive neuromodulation using an array of compact (approximately 1 cm(2) contact area) \"High-Definition\" (HD) electrodes, as compared to conventional tDCS (which uses two large pads that are approximately 35 cm(2)). In a previous transcutaneous study, we developed and validated designs for HD electrodes that reduce discomfort over >20 min session with 2 mA electrode current.</p><p><strong>Objective: </strong>The purpose of this study was to investigate the use of a chemical pretreatment with 6% benzocaine (topical numbing agent) to further reduce subjective discomfort during transcutaneous stimulation and to allow for better sham controlled studies.</p><p><strong>Methods: </strong>Pre-treatment with 6% benzocaine was compared with control (no pretreatment) for 22 min 2 mA of stimulation, with either CCNY-4 or Lectron II electroconductive gel, for both cathodal and anodal transcutaneous (forearm) stimulation (eight different combinations).</p><p><strong>Results: </strong>RESULTS show that for all conditions and polarities tested, stimulation with HD electrodes is safe and well tolerated and that pretreatment further reduced subjective discomfort.</p><p><strong>Conclusion: </strong>Pretreatment with a mild analgesic reduces discomfort during HD-tDCS.</p>","PeriodicalId":73093,"journal":{"name":"Frontiers in neuroengineering","volume":" ","pages":"28"},"PeriodicalIF":0.0000,"publicationDate":"2014-07-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.3389/fneng.2014.00028","citationCount":"28","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Frontiers in neuroengineering","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3389/fneng.2014.00028","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2014/1/1 0:00:00","PubModel":"eCollection","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 28
Abstract
Background: High-Definition transcranial Direct Current Stimulation (HD-tDCS) allows for non-invasive neuromodulation using an array of compact (approximately 1 cm(2) contact area) "High-Definition" (HD) electrodes, as compared to conventional tDCS (which uses two large pads that are approximately 35 cm(2)). In a previous transcutaneous study, we developed and validated designs for HD electrodes that reduce discomfort over >20 min session with 2 mA electrode current.
Objective: The purpose of this study was to investigate the use of a chemical pretreatment with 6% benzocaine (topical numbing agent) to further reduce subjective discomfort during transcutaneous stimulation and to allow for better sham controlled studies.
Methods: Pre-treatment with 6% benzocaine was compared with control (no pretreatment) for 22 min 2 mA of stimulation, with either CCNY-4 or Lectron II electroconductive gel, for both cathodal and anodal transcutaneous (forearm) stimulation (eight different combinations).
Results: RESULTS show that for all conditions and polarities tested, stimulation with HD electrodes is safe and well tolerated and that pretreatment further reduced subjective discomfort.
Conclusion: Pretreatment with a mild analgesic reduces discomfort during HD-tDCS.