Rupture of Subcutaneous Peritoneal Dialysis Catheter by Stretching Exercise: A Case Report.

Nagayuki Kaneshiro, Tsutomu Sakurada, Rina Hachisuka, Michiko Hisamichi, Masahiko Yazawa, Yugo Shibagaki
{"title":"Rupture of Subcutaneous Peritoneal Dialysis Catheter by Stretching Exercise: A Case Report.","authors":"Nagayuki Kaneshiro,&nbsp;Tsutomu Sakurada,&nbsp;Rina Hachisuka,&nbsp;Michiko Hisamichi,&nbsp;Masahiko Yazawa,&nbsp;Yugo Shibagaki","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>Rupture of the peritoneal dialysis (PD) catheter is rare complication. Here, we report a case of catheter rupture that occurred because of exercise after partial catheter reimplantation.A 66-year-old man with a history of end-stage kidney disease secondary to diabetic nephropathy experienced refractory exit-site and tunnel infection. After the infected parts of the catheter were excised, a partial catheter reimplantation was performed. At the time of that surgery, a presternal location was selected for the new exit site, and a titanium extender was used to connect the two catheters. The patient was discharged on postoperative day 3, but was readmitted for a pericatheter leak 5 days later. Fluoroscopy performed to investigate the cause demonstrated a pericatheter leak from the connecting portion between the titanium extender and the catheter.Surgery performed to repair the leak revealed that the catheter had ruptured. We believe that the cause of the rupture was mechanical stress induced by the patient's stretching exercise program. The PD catheter was made of silicone rubber with high elasticity. Even when such resilient materials are used, we recommend that, to avoid PD catheter rupture after partial reimplantation, clinicians should discourage the patient from stretching excessively.</p>","PeriodicalId":7361,"journal":{"name":"Advances in peritoneal dialysis. Conference on Peritoneal Dialysis","volume":"32 ","pages":"3-6"},"PeriodicalIF":0.0000,"publicationDate":"2016-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Advances in peritoneal dialysis. Conference on Peritoneal Dialysis","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Rupture of the peritoneal dialysis (PD) catheter is rare complication. Here, we report a case of catheter rupture that occurred because of exercise after partial catheter reimplantation.A 66-year-old man with a history of end-stage kidney disease secondary to diabetic nephropathy experienced refractory exit-site and tunnel infection. After the infected parts of the catheter were excised, a partial catheter reimplantation was performed. At the time of that surgery, a presternal location was selected for the new exit site, and a titanium extender was used to connect the two catheters. The patient was discharged on postoperative day 3, but was readmitted for a pericatheter leak 5 days later. Fluoroscopy performed to investigate the cause demonstrated a pericatheter leak from the connecting portion between the titanium extender and the catheter.Surgery performed to repair the leak revealed that the catheter had ruptured. We believe that the cause of the rupture was mechanical stress induced by the patient's stretching exercise program. The PD catheter was made of silicone rubber with high elasticity. Even when such resilient materials are used, we recommend that, to avoid PD catheter rupture after partial reimplantation, clinicians should discourage the patient from stretching excessively.

分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
伸展运动致皮下腹膜透析导管破裂1例。
腹膜透析(PD)导管破裂是罕见的并发症。在此,我们报告一例在部分导管再植后因运动导致导管破裂的病例。66岁男性,有继发于糖尿病肾病的终末期肾病病史,出现难治性出口和隧道感染。导管感染部分切除后,进行部分导管再植。在手术时,选择胸骨前位置作为新的出口位置,并使用钛扩展器连接两根导管。患者术后第3天出院,但5天后因导管外漏再次入院。进行透视检查以调查原因显示从钛扩展器和导管之间的连接部分导管周泄漏。手术修复泄漏后发现导管已经破裂。我们认为破裂的原因是由病人的伸展运动项目引起的机械应力。PD导管采用高弹性硅橡胶制成。即使使用这种有弹性的材料,我们建议,为了避免局部再植术后PD导管破裂,临床医生应劝阻患者过度拉伸。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Changes in Peritoneal Transport and Peritoneal Damage in Japanese Patients Undergoing Peritoneal Dialysis Using Neutral-pH Dialysate: A Retrospective Cohort Study at Two Centers. Maintaining Peritoneal Dialysis Adequacy: The Process of Incremental Prescription. Comparison of Various Scaling Parameters and Energy Expenditure in Peritoneal Dialysis Patients. Prevention of Peritoneal Dialysis Drop-Out. Accuracy of a Newly-Introduced Oscillometric Device for the Estimation of Arterial Stiffness Indices in Patients on Peritoneal Dialysis: A Preliminary Validation Study.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1