How is hip prosthesis and proximal femoral nail stability affected by lesser trochanter fractures: A comparative finite element analysis.

Mehmet Nuri Konya, Ahmet Aslan, Sibel Bakbak
{"title":"How is hip prosthesis and proximal femoral nail stability affected by lesser trochanter fractures: A comparative finite element analysis.","authors":"Mehmet Nuri Konya,&nbsp;Ahmet Aslan,&nbsp;Sibel Bakbak","doi":"10.5606/ehc.2018.58735","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>This study aims to evaluate the effects of lesser trochanter (LT) and iliopsoas tendon on implant stability by using finite element analysis (FEA).</p><p><strong>Materials and methods: </strong>Effects of iliacus and psoas major muscles on hip joint was evaluated with inverse dynamics methods to calculate joint reaction and muscle forces. Intertrochanteric femur fracture was simulated according to AO (Arbeitsgemeinschaft für Osteosynthesefragen) 31A1 and 31A2 classifications in three-dimensional modelling software. Uncemented three-dimensional model of modular nail prosthesis combination was used in FEA. All analyses were performed with Ti6Al4V's 114 GPa elastic modulus value. Effects of LT on implant stability were evaluated with two different implant designs using the same femoral stems and four different femoral models, two of which with intact LT and two of which with fractured LT.</p><p><strong>Results: </strong>Reaction forces of the hip joint decreased by 10% in the 0-40° hip flexion range. Maximum stress distribution for proximal femoral nail (PFN) model with fractured LT was 204.68 MPa at the distal locking screw on the interconnection point of PFN, while it was 335.35 MPa for the hip prosthesis with fractured LT. The direction of stress distribution for PFN model with fractured LT varied from medial to lateral and anterior to posterior. Maximum stress distribution for the hip prosthesis model with intact LT was 357.42 MPa, with direction of stress distribution from lateral to medial and posterior to anterior.</p><p><strong>Conclusion: </strong>Hip prosthesis models with intact or fractured LT were similar in terms of stress distribution and deformation values, while there were differences between PFN models with intact or fractured LT. Thus, intact LT was significant in PFN implant stability. Further clinical and experimental analyses are necessary on this topic.</p>","PeriodicalId":50551,"journal":{"name":"Eklem Hastaliklari Ve Cerrahisi-Joint Diseases and Related Surgery","volume":null,"pages":null},"PeriodicalIF":1.6000,"publicationDate":"2018-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"3","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Eklem Hastaliklari Ve Cerrahisi-Joint Diseases and Related Surgery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.5606/ehc.2018.58735","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 3

Abstract

Objectives: This study aims to evaluate the effects of lesser trochanter (LT) and iliopsoas tendon on implant stability by using finite element analysis (FEA).

Materials and methods: Effects of iliacus and psoas major muscles on hip joint was evaluated with inverse dynamics methods to calculate joint reaction and muscle forces. Intertrochanteric femur fracture was simulated according to AO (Arbeitsgemeinschaft für Osteosynthesefragen) 31A1 and 31A2 classifications in three-dimensional modelling software. Uncemented three-dimensional model of modular nail prosthesis combination was used in FEA. All analyses were performed with Ti6Al4V's 114 GPa elastic modulus value. Effects of LT on implant stability were evaluated with two different implant designs using the same femoral stems and four different femoral models, two of which with intact LT and two of which with fractured LT.

Results: Reaction forces of the hip joint decreased by 10% in the 0-40° hip flexion range. Maximum stress distribution for proximal femoral nail (PFN) model with fractured LT was 204.68 MPa at the distal locking screw on the interconnection point of PFN, while it was 335.35 MPa for the hip prosthesis with fractured LT. The direction of stress distribution for PFN model with fractured LT varied from medial to lateral and anterior to posterior. Maximum stress distribution for the hip prosthesis model with intact LT was 357.42 MPa, with direction of stress distribution from lateral to medial and posterior to anterior.

Conclusion: Hip prosthesis models with intact or fractured LT were similar in terms of stress distribution and deformation values, while there were differences between PFN models with intact or fractured LT. Thus, intact LT was significant in PFN implant stability. Further clinical and experimental analyses are necessary on this topic.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
小粗隆骨折对髋关节假体和股骨近端钉稳定性的影响:比较有限元分析。
目的:应用有限元分析(FEA)评价小粗隆(LT)和髂腰肌肌腱对假体稳定性的影响。材料与方法:采用逆动力学方法计算关节反作用力和肌力,评价髂大肌和腰肌对髋关节的影响。在三维建模软件中按照AO (Arbeitsgemeinschaft f r osteosynthesis efragen) 31A1和31A2分类模拟股骨粗隆间骨折。采用模块化甲假体组合的非胶结三维模型进行有限元分析。所有分析均采用Ti6Al4V的114 GPa弹性模量值进行。采用两种不同的假体设计,使用相同的股骨柄和四种不同的股骨模型,其中两种完整的假体和两种骨折的假体,评估了LT对假体稳定性的影响。结果:在0-40°髋关节屈曲范围内,髋关节的反作用力降低了10%。股骨近端股骨钉(PFN)模型在PFN连接点远端锁定螺钉处的最大应力分布为204.68 MPa,而LT骨折髋关节假体的最大应力分布为335.35 MPa。股骨近端股骨钉(PFN)模型的应力分布方向由内侧到外侧、前向后侧变化。完整LT髋关节假体模型最大应力分布为357.42 MPa,应力分布方向由外侧向内侧、后向前。结论:LT完整或骨折的假体模型在应力分布和变形值上相似,而LT完整或骨折的PFN模型之间存在差异。因此,LT完整对PFN假体的稳定性具有重要意义。进一步的临床和实验分析是必要的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
1.90
自引率
43.80%
发文量
0
审稿时长
>12 weeks
期刊介绍: Joint Diseases and Related Surgery (formerly published as Eklem Hastalıkları ve Cerrahisi) is the official publication of the Turkish Joint Diseases Foundation. Joint Diseases and Related Surgery is open access journal. The full text of the articles of the Journal is freely available without embargo since 1990. Joint Diseases and Related Surgery is international, double-blind peer-reviewed periodical journal bringing the latest developments in all aspects of joint diseases and related surgey.
期刊最新文献
The effects of focal brain damage on fracture healing: An experimental rat study. Endoscopic treatment of bilateral hallux saltans in an ordinary woman. Median nerve and carpal tunnel volume changes after two different surgical methods: A comparative magnetic resonance imaging study of mini-open and endoscopic carpal tunnel release. The effect of pentoxifylline on Achilles tendon healing in tenotomized rabbits. Icariin promotes early and late stages of fracture healing in rats.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1