{"title":"Development and validation of a measurement system for continuously monitoring postoperative reservoir levels.","authors":"S Eichhorn, T Reisinger, J Böhm, S Voss, S Doppler, R Lange, M Krane","doi":"10.1007/s13246-019-00746-5","DOIUrl":null,"url":null,"abstract":"<p><p>Following cardiac surgical procedures, multiple drainage systems remain in place inside the patient's chest to prevent the development of pericardial effusion or pneumothorax. Therefore, postoperative bleeding must be diligently observed. Currently, observation of the exudate rate is performed through periodical visual inspection of the reservoir. To improve postoperative monitoring, a measurement system based on load cells was developed to automatically detect bleeding rates. A reservoir retaining bracket was instrumented with a load cell. The signal was digitized by a microcontroller and then processed and displayed on customized software written in LabView. In cases where bleeding rates reach critical levels, the device will automatically sound an alarm. Additionally, the bleeding rate is displayed on the screen with the status of the alarm, as well as the fluid level of the reservoir. These data are all logged to a file. The measurement system has been validated for gain stability and drift, as well as for sensor accuracy, with different in vitro examinations. Additionally, performance of the measurement device was tested in a clinical pilot study on patients recovering from cardiac surgical procedures. The in vitro investigation showed that the monitoring device had excellent gain and drift stability, as well as sensor accuracy, with a resolution of 2.6 mL/h for the bleeding rate. During the clinical examination, bleeding rates of all patients were correctly measured. Continuously recording drainage volume using the developed system was comparable to manual measurements performed every 30 min by a nurse. Implementation of continuous digital measurements could improve patient safety and reduce the workload of medical professionals working in intensive care units.</p>","PeriodicalId":55430,"journal":{"name":"Australasian Physical & Engineering Sciences in Medicine","volume":"42 2","pages":"611-617"},"PeriodicalIF":0.0000,"publicationDate":"2019-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1007/s13246-019-00746-5","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Australasian Physical & Engineering Sciences in Medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/s13246-019-00746-5","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2019/3/13 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"Biochemistry, Genetics and Molecular Biology","Score":null,"Total":0}
引用次数: 1
Abstract
Following cardiac surgical procedures, multiple drainage systems remain in place inside the patient's chest to prevent the development of pericardial effusion or pneumothorax. Therefore, postoperative bleeding must be diligently observed. Currently, observation of the exudate rate is performed through periodical visual inspection of the reservoir. To improve postoperative monitoring, a measurement system based on load cells was developed to automatically detect bleeding rates. A reservoir retaining bracket was instrumented with a load cell. The signal was digitized by a microcontroller and then processed and displayed on customized software written in LabView. In cases where bleeding rates reach critical levels, the device will automatically sound an alarm. Additionally, the bleeding rate is displayed on the screen with the status of the alarm, as well as the fluid level of the reservoir. These data are all logged to a file. The measurement system has been validated for gain stability and drift, as well as for sensor accuracy, with different in vitro examinations. Additionally, performance of the measurement device was tested in a clinical pilot study on patients recovering from cardiac surgical procedures. The in vitro investigation showed that the monitoring device had excellent gain and drift stability, as well as sensor accuracy, with a resolution of 2.6 mL/h for the bleeding rate. During the clinical examination, bleeding rates of all patients were correctly measured. Continuously recording drainage volume using the developed system was comparable to manual measurements performed every 30 min by a nurse. Implementation of continuous digital measurements could improve patient safety and reduce the workload of medical professionals working in intensive care units.
期刊介绍:
Australasian Physical & Engineering Sciences in Medicine (APESM) is a multidisciplinary forum for information and research on the application of physics and engineering to medicine and human physiology. APESM covers a broad range of topics that include but is not limited to:
- Medical physics in radiotherapy
- Medical physics in diagnostic radiology
- Medical physics in nuclear medicine
- Mathematical modelling applied to medicine and human biology
- Clinical biomedical engineering
- Feature extraction, classification of EEG, ECG, EMG, EOG, and other biomedical signals;
- Medical imaging - contributions to new and improved methods;
- Modelling of physiological systems
- Image processing to extract information from images, e.g. fMRI, CT, etc.;
- Biomechanics, especially with applications to orthopaedics.
- Nanotechnology in medicine
APESM offers original reviews, scientific papers, scientific notes, technical papers, educational notes, book reviews and letters to the editor.
APESM is the journal of the Australasian College of Physical Scientists and Engineers in Medicine, and also the official journal of the College of Biomedical Engineers, Engineers Australia and the Asia-Oceania Federation of Organizations for Medical Physics.