{"title":"A new M23-based ELISA assay for anti-aquaporin 4 autoantibodies: diagnostic accuracy and clinical correlation.","authors":"Marilina Tampoia, Letizia Abbracciavento, Giuseppina Barberio, Martina Fabris, Nicola Bizzaro","doi":"10.1186/s13317-019-0115-7","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Although many assays have been developed to detect anti-aquaporin-4 (AQP4) antibodies, most of these assays require sophisticated techniques and are thus only available at specialized laboratories. The aim of this study was to evaluate the analytical and clinical performance of a new commercial enzyme-linked immunosorbent assay (ELISA RSR, AQP4 Ab Version 2) to detect anti-AQP4 antibodies performed on a fully automated system (SkyLAB 752).</p><p><strong>Methods: </strong>Serum samples from 64 patients with neuromyelitis optica spectrum disorders (NMOSD) (including NMO, longitudinally extensive myelitis-LETM, optical neuritis and myelitis) and 27 controls were tested for anti-AQP4 antibodies. All sera were previously tested using an indirect immunofluorescence (IIF) method on primate tissue, as the reference method. Commercial control sera were used to determine within-run, between-day and within-laboratory precision (CLSI guidelines).</p><p><strong>Results: </strong>At a cut-off value of 2.1 U/mL as determined by ROC curves, sensitivity and specificity for NMO were 83.3% and 100%, respectively. The ELISA assay provided 100% concordant results with the reference IIF method. The median concentration of anti-AQP4 antibodies was statistically higher in patients with NMO than in patients with LETM (<i>p </i>= 0.0006) or with other NMOSD and in controls (<i>p </i>< 0.0001). At the concentration of 12.4 and 28.1 U/mL, the within-run, between-day and within-laboratory coefficients of variation (CV) were 3.2% and 3%, 7.6% and 7.4%, and 8.2% and 8.0%, respectively.</p><p><strong>Conclusions: </strong>This new ELISA method performed on a fully automated system, showed high sensitivity and absolute specificity, good CV in precision tests, and provided observer-independent quantitative results.</p>","PeriodicalId":8655,"journal":{"name":"Auto-Immunity Highlights","volume":"10 1","pages":"5"},"PeriodicalIF":0.0000,"publicationDate":"2019-06-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1186/s13317-019-0115-7","citationCount":"8","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Auto-Immunity Highlights","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1186/s13317-019-0115-7","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2019/12/1 0:00:00","PubModel":"eCollection","JCR":"Q1","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 8
Abstract
Purpose: Although many assays have been developed to detect anti-aquaporin-4 (AQP4) antibodies, most of these assays require sophisticated techniques and are thus only available at specialized laboratories. The aim of this study was to evaluate the analytical and clinical performance of a new commercial enzyme-linked immunosorbent assay (ELISA RSR, AQP4 Ab Version 2) to detect anti-AQP4 antibodies performed on a fully automated system (SkyLAB 752).
Methods: Serum samples from 64 patients with neuromyelitis optica spectrum disorders (NMOSD) (including NMO, longitudinally extensive myelitis-LETM, optical neuritis and myelitis) and 27 controls were tested for anti-AQP4 antibodies. All sera were previously tested using an indirect immunofluorescence (IIF) method on primate tissue, as the reference method. Commercial control sera were used to determine within-run, between-day and within-laboratory precision (CLSI guidelines).
Results: At a cut-off value of 2.1 U/mL as determined by ROC curves, sensitivity and specificity for NMO were 83.3% and 100%, respectively. The ELISA assay provided 100% concordant results with the reference IIF method. The median concentration of anti-AQP4 antibodies was statistically higher in patients with NMO than in patients with LETM (p = 0.0006) or with other NMOSD and in controls (p < 0.0001). At the concentration of 12.4 and 28.1 U/mL, the within-run, between-day and within-laboratory coefficients of variation (CV) were 3.2% and 3%, 7.6% and 7.4%, and 8.2% and 8.0%, respectively.
Conclusions: This new ELISA method performed on a fully automated system, showed high sensitivity and absolute specificity, good CV in precision tests, and provided observer-independent quantitative results.