Filippo Scalise, Davide Margonato, Andrea Sole, Antonio Sorropago, Giovanni Sorropago, Giuseppe Mancia
{"title":"Ambulatory blood pressure monitoring by a novel cuffless device: a pilot study.","authors":"Filippo Scalise, Davide Margonato, Andrea Sole, Antonio Sorropago, Giovanni Sorropago, Giuseppe Mancia","doi":"10.1080/08037051.2020.1785273","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Cuffless blood pressure (CL-BP) measurements are believed to be a potentially alternative to cuff-occlusion-based (C-BP) measurement. A new cuffless device was developed for ambulatory BP monitoring. We assessed the accuracy of a new CL-BP device compared to a standard oscillatory C-BP device over the 24 h.</p><p><strong>Materials and methods: </strong>Eighty-four consecutive patients were included in the study. BP was measured simultaneously by the CL-BP device and by a C-BP device over the 24 h. Calculations included 24 h mean systolic (S) BP, the mean diastolic (D) BP and the heart rate (HR). Correlations between the CL-BP and C-BP measurements were sought using Pearson's correlation coefficients and Bland-Altman plots.</p><p><strong>Results: </strong>Using the C-BP device, the 24 h SBP value for the cohort was 125.4 ± 10.9 mmHg (mean ± SD); the corresponding DBP value being 75 ± 8.3 mmHg. Mean SBP/DBP were higher with the CL-BP device, i.e. 131.1 ± 15.9/80.2 ± 9.7 mmHg . The correlation coefficients between the two sets of values were significant (SBP: <i>r</i> = 0.58, DBP: <i>r</i> = 0.65). Better correlations for SBP and DBP were found 1) in patients with BMI > 25 (SBP: <i>r</i> = 0.65, DBP: <i>r</i> = 0.70) compared to those with BMI <25 and 2) in males compared to females (SBP: <i>r</i> = 0.71, DBP: <i>r</i> = 0.77).</p><p><strong>Conclusions: </strong>In our patients a CL-BP device estimated 24 h mean SBP and DBP differently from the classical oscillometric device, with a moderate correlation. CL-BP measurements were most accurate on male and overweight subjects.</p>","PeriodicalId":55591,"journal":{"name":"Blood Pressure","volume":"29 6","pages":"375-381"},"PeriodicalIF":2.3000,"publicationDate":"2020-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1080/08037051.2020.1785273","citationCount":"5","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood Pressure","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1080/08037051.2020.1785273","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2020/7/1 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 5
Abstract
Purpose: Cuffless blood pressure (CL-BP) measurements are believed to be a potentially alternative to cuff-occlusion-based (C-BP) measurement. A new cuffless device was developed for ambulatory BP monitoring. We assessed the accuracy of a new CL-BP device compared to a standard oscillatory C-BP device over the 24 h.
Materials and methods: Eighty-four consecutive patients were included in the study. BP was measured simultaneously by the CL-BP device and by a C-BP device over the 24 h. Calculations included 24 h mean systolic (S) BP, the mean diastolic (D) BP and the heart rate (HR). Correlations between the CL-BP and C-BP measurements were sought using Pearson's correlation coefficients and Bland-Altman plots.
Results: Using the C-BP device, the 24 h SBP value for the cohort was 125.4 ± 10.9 mmHg (mean ± SD); the corresponding DBP value being 75 ± 8.3 mmHg. Mean SBP/DBP were higher with the CL-BP device, i.e. 131.1 ± 15.9/80.2 ± 9.7 mmHg . The correlation coefficients between the two sets of values were significant (SBP: r = 0.58, DBP: r = 0.65). Better correlations for SBP and DBP were found 1) in patients with BMI > 25 (SBP: r = 0.65, DBP: r = 0.70) compared to those with BMI <25 and 2) in males compared to females (SBP: r = 0.71, DBP: r = 0.77).
Conclusions: In our patients a CL-BP device estimated 24 h mean SBP and DBP differently from the classical oscillometric device, with a moderate correlation. CL-BP measurements were most accurate on male and overweight subjects.
期刊介绍:
For outstanding coverage of the latest advances in hypertension research, turn to Blood Pressure, a primary source for authoritative and timely information on all aspects of hypertension research and management.
Features include:
• Physiology and pathophysiology of blood pressure regulation
• Primary and secondary hypertension
• Cerebrovascular and cardiovascular complications of hypertension
• Detection, treatment and follow-up of hypertension
• Non pharmacological and pharmacological management
• Large outcome trials in hypertension.