Sebastian Bäurle, Anja Viestenz, Berthold Seitz, Arne Viestenz
{"title":"[Intraocular pressure elevation after vitrectomy-Goldmann applanation tonometry measures lower intraocular pressure than dynamic contour tonometry].","authors":"Sebastian Bäurle, Anja Viestenz, Berthold Seitz, Arne Viestenz","doi":"10.1007/s00347-021-01443-z","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The dynamic contour tonometer PASCAL (DCT) is a direct noninvasive digital tonometer matching the corneal contour. We compared the DCT with the Goldmann applanation tonometer (GAT).</p><p><strong>Methods: </strong>A total of 100 eyes were included in this prospective comparative clinical investigation and intraocular pressure (IOP) was measured with GAT and DCT before and after pars plana vitrectomy (ppV). Different intraocular endotamponades, such as gas and silicone oil were used. Preoperative IOP, changes in IOP postoperatively and intertonometer differences were measured.</p><p><strong>Results: </strong>The preoperative mean IOP was 15.8 ± 5.2 mm Hg measured with the GAT and 17.5 ± 5.9 mm Hg with DCT. On the first postoperative day, in eyes with gas endotamponade the mean IOP rise was 2.5 mm Hg (p = 0.035) and 18 eyes were measured with an IOP of ≥ 25 mm Hg (19.1%; DCT). Postoperative IOP measured by GAT was 2.5 mm Hg lower and in eyes with gas endotamponade the GAT measured the IOP 3.0 mm Hg lower than the DCT. In eyes with a postoperative IOP above 20 mm Hg, the GAT measured the IOP 4.7 mm Hg lower compared to the DCT. Out of 18 eyes 10 with IOP ≥ 25 mm Hg were not recognized as hypertensive (≥ 25 mm Hg) by GAT. In 13% of the eyes the DCT measured at least 6 mm Hg higher IOP levels than GAT on the first postoperative day. In one extreme case with gas endotamponade, GAT underestimated the IOP by 12 mm Hg compared to the DCT.</p><p><strong>Conclusion: </strong>The IOP spikes after ppV are a feared complication and can lead to irreversible visual loss. Depending on the endotamponade, GAT measures lower IOP than DCT, especially in IOP spikes caused by expansive gas endotamponade. Postoperative IOP measurement is important and readings between GAT and DCT can differ.</p>","PeriodicalId":54676,"journal":{"name":"Ophthalmologe","volume":" ","pages":"71-76"},"PeriodicalIF":0.0000,"publicationDate":"2022-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1007/s00347-021-01443-z","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Ophthalmologe","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00347-021-01443-z","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2021/7/6 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 1
Abstract
Background: The dynamic contour tonometer PASCAL (DCT) is a direct noninvasive digital tonometer matching the corneal contour. We compared the DCT with the Goldmann applanation tonometer (GAT).
Methods: A total of 100 eyes were included in this prospective comparative clinical investigation and intraocular pressure (IOP) was measured with GAT and DCT before and after pars plana vitrectomy (ppV). Different intraocular endotamponades, such as gas and silicone oil were used. Preoperative IOP, changes in IOP postoperatively and intertonometer differences were measured.
Results: The preoperative mean IOP was 15.8 ± 5.2 mm Hg measured with the GAT and 17.5 ± 5.9 mm Hg with DCT. On the first postoperative day, in eyes with gas endotamponade the mean IOP rise was 2.5 mm Hg (p = 0.035) and 18 eyes were measured with an IOP of ≥ 25 mm Hg (19.1%; DCT). Postoperative IOP measured by GAT was 2.5 mm Hg lower and in eyes with gas endotamponade the GAT measured the IOP 3.0 mm Hg lower than the DCT. In eyes with a postoperative IOP above 20 mm Hg, the GAT measured the IOP 4.7 mm Hg lower compared to the DCT. Out of 18 eyes 10 with IOP ≥ 25 mm Hg were not recognized as hypertensive (≥ 25 mm Hg) by GAT. In 13% of the eyes the DCT measured at least 6 mm Hg higher IOP levels than GAT on the first postoperative day. In one extreme case with gas endotamponade, GAT underestimated the IOP by 12 mm Hg compared to the DCT.
Conclusion: The IOP spikes after ppV are a feared complication and can lead to irreversible visual loss. Depending on the endotamponade, GAT measures lower IOP than DCT, especially in IOP spikes caused by expansive gas endotamponade. Postoperative IOP measurement is important and readings between GAT and DCT can differ.
背景:动态轮廓眼压计PASCAL (DCT)是一种直接、无创的与角膜轮廓匹配的数字眼压计。我们比较了DCT和Goldmann眼压计(GAT)。方法:对100只眼进行前瞻性比较临床研究,采用GAT和DCT测量玻璃体切割(ppV)前后的眼压(IOP)。采用不同的眼内填塞液,如气体和硅油。测量术前眼压、术后眼压变化及眼压计间差异。结果:术前GAT平均眼压为15.8 ±5.2 mm Hg, DCT平均眼压为17.5 ±5.9 mm Hg。术后第1天,气体内压填塞眼的平均眼压升高2.5 mm Hg (p = 0.035),18只眼的眼压≥ 25 mm Hg (19.1%;DCT)。术后GAT测得的IOP比DCT低2.5 mm Hg,气体内压填塞眼GAT测得的IOP比DCT低3.0 mm Hg。对于术后IOP高于20 mm Hg的眼睛,GAT测量的IOP比DCT低4.7 mm Hg。在18只眼中,10只眼IOP≥ 25 mm Hg, GAT未识别为高血压(≥ 25 mm Hg)。在13%的眼睛中,术后第一天DCT测量的IOP水平比GAT至少高6 mm Hg。在一个气体内压填塞的极端病例中,与DCT相比,GAT低估了12 mm Hg的IOP。结论:ppV术后眼压尖峰是一种可怕的并发症,可导致不可逆的视力丧失。根据内压填塞情况,GAT测量的IOP比DCT低,特别是在膨胀气体内压填塞引起的IOP峰值时。术后IOP测量很重要,GAT和DCT的读数可能不同。
期刊介绍:
Der Ophthalmologe is an internationally recognized journal dealing with all aspects of ophthalmology. The journal serves both the scientific exchange and the continuing education of ophthalmologists.
Freely submitted original papers allow the presentation of important clinical studies and serve scientific exchange.
Case reports feature interesting cases and aim at optimizing diagnostic and therapeutic strategies.
Comprehensive reviews on a specific topical issue focus on providing evidenced based information on diagnostics and therapy.
Review articles under the rubric ''Continuing Medical Education'' present verified results of scientific research and their integration into daily practice.