Development of a One-Micrometer-Diameter Particle Size Standard Reference Material.

G W Mulholland, A W Hartman, G G Hembree, Egon Marx, T R Lettieri
{"title":"Development of a One-Micrometer-Diameter Particle Size Standard Reference Material.","authors":"G W Mulholland,&nbsp;A W Hartman,&nbsp;G G Hembree,&nbsp;Egon Marx,&nbsp;T R Lettieri","doi":"10.6028/jres.090.001","DOIUrl":null,"url":null,"abstract":"<p><p>The average diameter of the first micrometer particle size standard (Standard Reference Material 1690), an aqueous suspension of monosized polystyrene spheres with a nominal 1 μm diameter, was accurately determined by three independent techniques. In one technique the intensity of light scattered by a diluted suspension of polystyrene spheres was measured as a function of scattering angle, using a He-Ne laser polarized in the vertical direction. The second technique consisted of measuring as a function of angle the intensity of light scattered from individual polystyrene spheres suspended in air, using a He-Cd laser with light polarized parallel and perpendicular to the scattering plane. The measurement of row length by optical microscopy for polystyrene spheres arranged in close-packed, two-dimensional hexagonal arrays was the basis of the third technique. The measurement errors for each technique were quantitatively assessed. For the light scattering experiments, this required simulation with numerical experiments. The average diameter determined by each technique agreed within 0.5% with the most accurate value being 0.895±0.007 μm based on light scattering by an aqueous suspension. Transmission electron microscopy, flow through electrical sensing zone counter measurements, and optical microscopy were also used to obtain more detailed information on the size distribution including the standard deviation (0.0095 μm), fraction of off-size particles, and the fraction of agglomerated doublets (1.5%).</p>","PeriodicalId":93321,"journal":{"name":"Journal of research of the National Bureau of Standards (1977)","volume":"90 1","pages":"3-26"},"PeriodicalIF":0.0000,"publicationDate":"1985-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6768151/pdf/jres-90-003.pdf","citationCount":"25","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of research of the National Bureau of Standards (1977)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.6028/jres.090.001","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 25

Abstract

The average diameter of the first micrometer particle size standard (Standard Reference Material 1690), an aqueous suspension of monosized polystyrene spheres with a nominal 1 μm diameter, was accurately determined by three independent techniques. In one technique the intensity of light scattered by a diluted suspension of polystyrene spheres was measured as a function of scattering angle, using a He-Ne laser polarized in the vertical direction. The second technique consisted of measuring as a function of angle the intensity of light scattered from individual polystyrene spheres suspended in air, using a He-Cd laser with light polarized parallel and perpendicular to the scattering plane. The measurement of row length by optical microscopy for polystyrene spheres arranged in close-packed, two-dimensional hexagonal arrays was the basis of the third technique. The measurement errors for each technique were quantitatively assessed. For the light scattering experiments, this required simulation with numerical experiments. The average diameter determined by each technique agreed within 0.5% with the most accurate value being 0.895±0.007 μm based on light scattering by an aqueous suspension. Transmission electron microscopy, flow through electrical sensing zone counter measurements, and optical microscopy were also used to obtain more detailed information on the size distribution including the standard deviation (0.0095 μm), fraction of off-size particles, and the fraction of agglomerated doublets (1.5%).

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
一种直径为一微米的粒度标准参考物质的研制。
通过三种独立技术准确测定了第一微米粒度标准品(标准参考物质1690)的平均直径,该标准品是标称直径为1μm的单尺寸聚苯乙烯球的水悬浮液。在一种技术中,使用在垂直方向上偏振的He-Ne激光器,测量由聚苯乙烯球的稀释悬浮液散射的光的强度作为散射角的函数。第二种技术是使用平行和垂直于散射平面偏振的He-Cd激光器,测量悬浮在空气中的单个聚苯乙烯球散射的光的强度,作为角度的函数。通过光学显微镜测量排列在紧密堆积的二维六边形阵列中的聚苯乙烯球的行长是第三种技术的基础。对每种技术的测量误差进行了定量评估。对于光散射实验,这需要用数值实验进行模拟。根据水悬浮液的光散射,每种技术确定的平均直径在0.5%以内,最准确的值为0.895±0.007μm。还使用透射电子显微镜、流通电感应区计数器测量和光学显微镜来获得关于尺寸分布的更详细信息,包括标准偏差(0.0095μm)、粒径偏小颗粒的分数和团聚双峰的分数(1.5%)。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
The Temperature Dependence of Spectral Broadening in the Hg (61S0-63P1) Multiplet At High Optical Densities. Absolute Isotopic Abundance Ratio And Atomic Weight Of a Reference Sample of Gallium. Thermal Expansion of Platinum And Platinum-Rhodium Alloys. The Triple Point of Oxygen In Sealed Transportable Cells. A Multi-kilogram Capacity Calorimeter For Heterogeneous Materials.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1