Deviating glucose results in an international dual-center study. A root cause investigation.

IF 3.8 3区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Biochemia Medica Pub Date : 2022-02-15 Epub Date: 2021-12-15 DOI:10.11613/BM.2022.011001
Janne Cadamuro, Peter Bergsten, Katharina Mörwald, Anders Forslund, Marie Dahlbom, Jonas Bergquist, Iris Ciba, Susanne M Brunner, Jeanne Jabbour, Daniel Weghuber
{"title":"Deviating glucose results in an international dual-center study. A root cause investigation.","authors":"Janne Cadamuro,&nbsp;Peter Bergsten,&nbsp;Katharina Mörwald,&nbsp;Anders Forslund,&nbsp;Marie Dahlbom,&nbsp;Jonas Bergquist,&nbsp;Iris Ciba,&nbsp;Susanne M Brunner,&nbsp;Jeanne Jabbour,&nbsp;Daniel Weghuber","doi":"10.11613/BM.2022.011001","DOIUrl":null,"url":null,"abstract":"<p><p>During a dual-center study on obese and normal weight children and adolescents, focusing on glucose metabolism, we observed a marked difference in glucose results (N = 16,840) between the two sites, Salzburg, Austria and Uppsala, Sweden (P < 0.001). After excluding differences in patient characteristics between the two populations as cause of this finding, we investigated other preanalytic influences. Finally, only the tubes used for blood collection at the two sites were left to evaluate. While the Vacuette FC-Mix tube (Greiner Bio-One, Kremsmünster, Austria) was used in Uppsala, in Salzburg blood collections were performed with a lithium heparin tube (LH-Monovette, Sarstedt, Germany). To prove our hypothesis, we collected two blood samples in either of these tubes from 51 children (Salzburg N = 27, Uppsala N = 24) and compared the measured glucose results. Indeed, we found the suspected bias and calculated a correction formula, which significantly diminished the differences of glucose results between the two sites (<i>P</i> = 0.023). Our finding is in line with those of other studies and although this issue should be widely known, we feel that it is widely neglected, especially when comparing glucose concentrations across Europe, using large databases without any information on preanalytic sample handling.</p>","PeriodicalId":9021,"journal":{"name":"Biochemia Medica","volume":"32 1","pages":"011001"},"PeriodicalIF":3.8000,"publicationDate":"2022-02-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8672384/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biochemia Medica","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.11613/BM.2022.011001","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2021/12/15 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

During a dual-center study on obese and normal weight children and adolescents, focusing on glucose metabolism, we observed a marked difference in glucose results (N = 16,840) between the two sites, Salzburg, Austria and Uppsala, Sweden (P < 0.001). After excluding differences in patient characteristics between the two populations as cause of this finding, we investigated other preanalytic influences. Finally, only the tubes used for blood collection at the two sites were left to evaluate. While the Vacuette FC-Mix tube (Greiner Bio-One, Kremsmünster, Austria) was used in Uppsala, in Salzburg blood collections were performed with a lithium heparin tube (LH-Monovette, Sarstedt, Germany). To prove our hypothesis, we collected two blood samples in either of these tubes from 51 children (Salzburg N = 27, Uppsala N = 24) and compared the measured glucose results. Indeed, we found the suspected bias and calculated a correction formula, which significantly diminished the differences of glucose results between the two sites (P = 0.023). Our finding is in line with those of other studies and although this issue should be widely known, we feel that it is widely neglected, especially when comparing glucose concentrations across Europe, using large databases without any information on preanalytic sample handling.

Abstract Image

Abstract Image

Abstract Image

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
一项国际双中心研究结果显示葡萄糖偏离。根本原因调查。
在一项针对肥胖和正常体重儿童和青少年的双中心研究中,我们观察到两个地点(奥地利萨尔茨堡和瑞典乌普萨拉)的葡萄糖结果有显著差异(N = 16,840) (P < 0.001)。在排除了两个人群之间患者特征的差异作为这一发现的原因后,我们调查了其他分析前影响。最后,只留下两个地点用于采血的管子进行评估。乌普萨拉使用Vacuette FC-Mix管(Greiner Bio-One, kremsmnster,奥地利),在萨尔茨堡采集血液时使用锂肝素管(hl - monovette,德国萨斯泰特)。为了证明我们的假设,我们从51名儿童(Salzburg N = 27, Uppsala N = 24)中采集了两份血样,并比较了测量的葡萄糖结果。事实上,我们发现了可疑的偏倚,并计算了一个校正公式,显著减小了两个地点之间葡萄糖结果的差异(P = 0.023)。我们的发现与其他研究一致,尽管这个问题应该广为人知,但我们觉得它被广泛忽视了,特别是在比较欧洲各地的葡萄糖浓度时,使用大型数据库而没有任何分析前样品处理的信息。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Biochemia Medica
Biochemia Medica 医学-医学实验技术
CiteScore
5.50
自引率
3.00%
发文量
70
审稿时长
>12 weeks
期刊介绍: Biochemia Medica is the official peer-reviewed journal of the Croatian Society of Medical Biochemistry and Laboratory Medicine. Journal provides a wide coverage of research in all aspects of clinical chemistry and laboratory medicine. Following categories fit into the scope of the Journal: general clinical chemistry, haematology and haemostasis, molecular diagnostics and endocrinology. Development, validation and verification of analytical techniques and methods applicable to clinical chemistry and laboratory medicine are welcome as well as studies dealing with laboratory organization, automation and quality control. Journal publishes on a regular basis educative preanalytical case reports (Preanalytical mysteries), articles dealing with applied biostatistics (Lessons in biostatistics) and research integrity (Research integrity corner).
期刊最新文献
Quality assurance of add-on testing in plasma samples: stability limit for 29 biochemical analytes Two cases with discrepancy in the quantitative cytological assessment of cerebrospinal fluid in neonatal samples using light microscopy in comparison with Sysmex XN-1000 Common P-glycoprotein () polymorphisms do not seem to be associated with the risk of rivaroxaban-related bleeding events: Preliminary data CLSI-based verification and establishment of reference intervals for common biochemical assays in Croatian newborns Prediction interval: A powerful statistical tool for monitoring patients and analytical systems
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1