Fracture strength of all-ceramic (Y-TZP) three- and four-unit fixed dental prostheses with different connector design and production history.

Q Dentistry Swedish dental journal Pub Date : 2013-01-01
Deyar Jallal Hadi Mahmood, Ewa H Linderoth, Per Vult Von Steyern, Ann Wennerberg
{"title":"Fracture strength of all-ceramic (Y-TZP) three- and four-unit fixed dental prostheses with different connector design and production history.","authors":"Deyar Jallal Hadi Mahmood,&nbsp;Ewa H Linderoth,&nbsp;Per Vult Von Steyern,&nbsp;Ann Wennerberg","doi":"","DOIUrl":null,"url":null,"abstract":"<p><strong>Aims: </strong>To investigate how different default settings for connector design of two different CAD/CAM systems, i.e. to compare how different radii of curvature in the embrasure area of the connector affect the fracture strength and the fracture mode of 3-unit and 4-unit all-ceramic FDPs made from Y-TZP and to investigate how the number of pontics affect the fracture strength of Y-TZP.</p><p><strong>Material and methods: </strong>32 all-ceramic three (3Z) and four (4Z)-unit, Y-TZP, FDP cores were made and divided in 4 groups, with 8 FDP in each group. The groups 3Z:1 and 4Z:1 were generated with a mechanical scanner, Procera Forte and the FDPs in group 3Z:2 and 4Z:2 were generated with an optical scanner, NobelProcera Scanner. The connector dimensions were set to 3 mm x 3 mm and core was set to 0.7 mm. The design of radius of the gingival and occlusal embrasures in the connector areas was set according to default settings and the manufacturer's recommendations. All the FDP cores were subjected to heat treatment, thermocycled for 5,000 cycles, preloaded for 10,000 cycles to simulate ageing and finally loaded to fracture.</p><p><strong>Results: </strong>Regarding connector design a significant difference was found between group 3Z:1 and 3Z:2 (P<0.05), and group 4Z:1 (50% of the FDPs fractured during preloading 30-300N) and 4Z:2 (P<0.05). An extra pontic decreased the fracture strength up to 45%.</p><p><strong>Conclusions: </strong>The default settings of the two different CAD/CAM systems had a great impact on the fracture strength. It is important that a CAD/CAM system is equipped with possibilities to design a connector that fulfils the clinical demands of mechanical function and longevity. The most crucial factor for the load-bearing capacity isthe design of the radius of the gingival embrasures. Increasing the number of pontics from three to four decreases the load-bearing capacity nearly twice.</p>","PeriodicalId":22114,"journal":{"name":"Swedish dental journal","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2013-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Swedish dental journal","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q","JCRName":"Dentistry","Score":null,"Total":0}
引用次数: 0

Abstract

Aims: To investigate how different default settings for connector design of two different CAD/CAM systems, i.e. to compare how different radii of curvature in the embrasure area of the connector affect the fracture strength and the fracture mode of 3-unit and 4-unit all-ceramic FDPs made from Y-TZP and to investigate how the number of pontics affect the fracture strength of Y-TZP.

Material and methods: 32 all-ceramic three (3Z) and four (4Z)-unit, Y-TZP, FDP cores were made and divided in 4 groups, with 8 FDP in each group. The groups 3Z:1 and 4Z:1 were generated with a mechanical scanner, Procera Forte and the FDPs in group 3Z:2 and 4Z:2 were generated with an optical scanner, NobelProcera Scanner. The connector dimensions were set to 3 mm x 3 mm and core was set to 0.7 mm. The design of radius of the gingival and occlusal embrasures in the connector areas was set according to default settings and the manufacturer's recommendations. All the FDP cores were subjected to heat treatment, thermocycled for 5,000 cycles, preloaded for 10,000 cycles to simulate ageing and finally loaded to fracture.

Results: Regarding connector design a significant difference was found between group 3Z:1 and 3Z:2 (P<0.05), and group 4Z:1 (50% of the FDPs fractured during preloading 30-300N) and 4Z:2 (P<0.05). An extra pontic decreased the fracture strength up to 45%.

Conclusions: The default settings of the two different CAD/CAM systems had a great impact on the fracture strength. It is important that a CAD/CAM system is equipped with possibilities to design a connector that fulfils the clinical demands of mechanical function and longevity. The most crucial factor for the load-bearing capacity isthe design of the radius of the gingival embrasures. Increasing the number of pontics from three to four decreases the load-bearing capacity nearly twice.

分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
不同接头设计和生产历史的全陶瓷(Y-TZP)三单元和四单元固定义齿的断裂强度。
目的:研究两种不同CAD/CAM系统对连接器设计的不同默认设置,即比较由Y-TZP制成的3-unit和4-unit全陶瓷fdp的不同曲率半径对连接器断裂强度和断裂模式的影响,以及研究柱数对Y-TZP断裂强度的影响。材料与方法:制作全陶瓷三(3Z)和四(4Z)-单元,Y-TZP, FDP芯32个,分为4组,每组8个FDP。3Z:1和4Z:1组用机械扫描仪生成Procera Forte, 3Z:2和4Z:2组用光学扫描仪NobelProcera scanner生成fdp。连接器尺寸设置为3mm x 3mm,芯设置为0.7 mm。根据默认设置和制造商的建议设置连接区域的牙龈半径和咬合膜的设计。所有FDP芯都经过热处理,热循环5000次,预加载10000次以模拟老化,最后加载至断裂。结果:3Z:1组与3Z:2组在连接器设计上存在显著差异(p)。结论:两种不同CAD/CAM系统的默认设置对断裂强度有很大影响。重要的是,CAD/CAM系统配备了设计连接器的可能性,以满足临床对机械功能和寿命的要求。龈膜半径的设计是影响龈膜承载能力的关键因素。将桥架的数量从3个增加到4个,其承载能力降低了近两倍。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Swedish dental journal
Swedish dental journal 医学-牙科与口腔外科
CiteScore
0.85
自引率
0.00%
发文量
0
审稿时长
>12 weeks
期刊介绍: Swedish Dental Journal is the scientific journal of the Swedish Dental Association and the Swedish Dental Society. It is published 4 times a year to promote practice, education and research within odontology. Manuscripts containing original research are accepted for consideraion if neither the article nor any part of its essential substance has been or will be published elsewhere. Reviews, Case Reports and Short Communications will also be considered for publication.
期刊最新文献
Adverbs Verbs Punctuation Conjunctions Prepositions
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1