Contactless Measurement of Integrity of Silicone Coating on Self-Expandable Esophageal Nitinol Stents.

Martin Kopeček, Jiří Záhora, Aleš Bezrouk
{"title":"Contactless Measurement of Integrity of Silicone Coating on Self-Expandable Esophageal Nitinol Stents.","authors":"Martin Kopeček,&nbsp;Jiří Záhora,&nbsp;Aleš Bezrouk","doi":"10.14712/18059694.2022.11","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>A stent is a mesh tube inserted into a natural passage in the body to prevent disease induction. Self-expandable esophageal nitinol stents such as SX-ELLA Stent Esophageal HV (HV Stent Plus) can be indicated for palliation of malignant esophageal strictures, for the treatment of benign esophageal strictures that are refractory to standard therapy and for the treatment of esophago-respiratory fistulas. A silicone-stent coating is used for tumor in-growth prevention and esophago-respiratory fistula occlusion. The thickness of the stent and the overall integrity of the silicone coating of all wires indicate the overall mechanical properties of the esophageal stent and the resistance to external adverse events such as corrosion and mechanical and chemical resistance.</p><p><strong>Methods: </strong>The polymer multicomponent epoxy resin - a mixture of Epon and Durcupan - was used as a method for robust sample stabilization. A cutting system using a thin water beam with a powder (Blue Line) was chosen as the best variant to obtain 6 samples for both-sided measurement (10 measuring sides). The optical microscopic reflective light method was used to examine wire crossing points in the sections. Fifty values were measured on either sample side for the internal, external and mesh thickness of the silicone stent layer. The wire crossing points were selected so that the silicone layer structure could be clearly seen, and the wires approached each other most closely. Only approximately 4 to 8 crossing points in each section could be measured when applying this approach. The resolution of the microscope and calibration (based on the camera used) was 0.677 μm/pixel.</p><p><strong>Results: </strong>Additional data could be obtained on 8 planes. Two boundary samples were destroyed by the cutting process. Whole coating of the stent was around all mesh wires, especially in areas with higher mechanical stress (wire crossing). The minimum detectable and admissible value determined for all 3 measuring areas (internal, external, mesh) on the wire crossings was 6.77 μm, i.e., 10 pixels, based on the microscope resolution and manufacturer's methodology. The results were characterized by p < 0.001 for all 3 parameters. We tested opposite samples in each section to verify the section quality and data consistency. For the 4 areas, the data were significantly different, but the thickness differences were only on the order of units percent, so the measurements were not appreciably affected. We assume that the material cutting loss, making up 1-2 mm, contributed to the differences in the sections.</p><p><strong>Conclusion: </strong>We examined the overall integrity of the silicone coating of the esophageal stent. The method of HV stent anchoring in a polymeric bath followed by cutting with a waterjet and sample measurement under an optical microscope proved to be very simple and reliable. Sufficient thicknesses of the silicone layer on the wire cross sections were verified. The coated silicone layer thickness appeared to be significantly different along the stent from the proximal part to the distant part, presumably due to the manufacturing technology.</p>","PeriodicalId":35758,"journal":{"name":"Acta medica (Hradec Kralove)","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2022-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Acta medica (Hradec Kralove)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.14712/18059694.2022.11","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

Abstract

Objectives: A stent is a mesh tube inserted into a natural passage in the body to prevent disease induction. Self-expandable esophageal nitinol stents such as SX-ELLA Stent Esophageal HV (HV Stent Plus) can be indicated for palliation of malignant esophageal strictures, for the treatment of benign esophageal strictures that are refractory to standard therapy and for the treatment of esophago-respiratory fistulas. A silicone-stent coating is used for tumor in-growth prevention and esophago-respiratory fistula occlusion. The thickness of the stent and the overall integrity of the silicone coating of all wires indicate the overall mechanical properties of the esophageal stent and the resistance to external adverse events such as corrosion and mechanical and chemical resistance.

Methods: The polymer multicomponent epoxy resin - a mixture of Epon and Durcupan - was used as a method for robust sample stabilization. A cutting system using a thin water beam with a powder (Blue Line) was chosen as the best variant to obtain 6 samples for both-sided measurement (10 measuring sides). The optical microscopic reflective light method was used to examine wire crossing points in the sections. Fifty values were measured on either sample side for the internal, external and mesh thickness of the silicone stent layer. The wire crossing points were selected so that the silicone layer structure could be clearly seen, and the wires approached each other most closely. Only approximately 4 to 8 crossing points in each section could be measured when applying this approach. The resolution of the microscope and calibration (based on the camera used) was 0.677 μm/pixel.

Results: Additional data could be obtained on 8 planes. Two boundary samples were destroyed by the cutting process. Whole coating of the stent was around all mesh wires, especially in areas with higher mechanical stress (wire crossing). The minimum detectable and admissible value determined for all 3 measuring areas (internal, external, mesh) on the wire crossings was 6.77 μm, i.e., 10 pixels, based on the microscope resolution and manufacturer's methodology. The results were characterized by p < 0.001 for all 3 parameters. We tested opposite samples in each section to verify the section quality and data consistency. For the 4 areas, the data were significantly different, but the thickness differences were only on the order of units percent, so the measurements were not appreciably affected. We assume that the material cutting loss, making up 1-2 mm, contributed to the differences in the sections.

Conclusion: We examined the overall integrity of the silicone coating of the esophageal stent. The method of HV stent anchoring in a polymeric bath followed by cutting with a waterjet and sample measurement under an optical microscope proved to be very simple and reliable. Sufficient thicknesses of the silicone layer on the wire cross sections were verified. The coated silicone layer thickness appeared to be significantly different along the stent from the proximal part to the distant part, presumably due to the manufacturing technology.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
自膨胀食管镍钛诺支架硅胶涂层完整性的非接触测量。
目的:支架是一种插入体内自然通道的网状管,以防止疾病的诱发。SX-ELLA Stent esophageal HV (HV Stent Plus)等自膨胀式食管镍钛诺支架可用于缓解恶性食管狭窄、治疗标准治疗难治性良性食管狭窄和治疗食管呼吸瘘。硅胶支架涂层用于肿瘤生长预防和食管-呼吸瘘闭塞。支架的厚度和所有金属丝硅胶涂层的整体完整性表明了食管支架的整体力学性能以及对外部不良事件的抵抗力,如腐蚀和机械耐化学性。方法:采用多组分环氧树脂(Epon和Durcupan的混合物)作为稳压稳定样品的方法。选择使用细水束和粉末(蓝线)的切割系统作为最佳变体,获得6个样品用于双面测量(10个测量边)。采用光学显微反射光法对截面中的导线交叉点进行检测。在每个样品侧测量硅胶支架层的内部、外部和网格厚度的50个值。导线交叉点的选择使硅酮层结构清晰可见,导线之间最接近。当采用这种方法时,每个剖面只能测量大约4到8个交叉点。显微镜和定标分辨率(基于所用相机)为0.677 μm/pixel。结果:在8个平面上获得了额外的数据。切割过程破坏了两个边界样品。支架的整个涂层包裹在所有网丝周围,特别是在机械应力较高的区域(线交叉)。根据显微镜分辨率和制造商的方法,在导线交叉处的所有3个测量区域(内部,外部,网格)确定的最小可检测值和允许值为6.77 μm,即10像素。3个参数的结果均以p < 0.001为特征。为了验证剖面质量和数据的一致性,我们在每个剖面中测试了相反的样本。对于4个区域,数据差异显著,但厚度差异仅在单位百分比量级,因此测量结果没有明显影响。我们假设材料切割损失占1-2毫米,导致了截面的差异。结论:我们检查了食管支架硅胶涂层的整体完整性。HV支架在聚合物浴中锚定,然后用水射流切割,在光学显微镜下进行样品测量,这种方法非常简单可靠。验证了导线截面上硅酮层的足够厚度。从支架近端到远端,涂层硅胶层厚度沿支架方向明显不同,可能是由于制造技术的原因。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Acta medica (Hradec Kralove)
Acta medica (Hradec Kralove) Medicine-Medicine (all)
CiteScore
1.10
自引率
0.00%
发文量
8
审稿时长
20 weeks
期刊介绍: Acta Medica (Hradec Králové) is a multidisciplinary medical journal published by the Faculty of Medicine in Hradec Králové - Karolinum Press, the publishing house of Charles University. The journal is peer-reviewed and published quarterly in both paper and electronic form. The language of Acta Medica is English. Offerings include review articles, original articles, brief communications, case reports, announcements and notices. The journal was founded in 1958 under the title "A Collection of Scientific Works of the Charles University Faculty of Medicine in Hradec Kralove."
期刊最新文献
Antithrombin Deficiency: Frequency in Patients with Thrombosis and Thrombophilic Families. Drug Rash with Eosinophilia and Systemic Symptoms (DRESS Syndrome). Wireless Monitoring of Gastrointestinal Transit Time, Intra-luminal pH, Pressure and Temperature in Experimental Pigs: A Pilot Study. The Sex Ratio at Birth Is Higher in Māori than in Non-Māori Populations in Aotearoa New Zealand. First Croatian Case of Double Aneuploidy: A Child With Klinefelter and Edwards Syndrome (48,XXY,+18) - Possible Causes and Contributing Factors.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1