Dual-Layer Detector Head CT to Maintain Image Quality While Reducing the Radiation Dose in Pediatric Patients.

IF 3.1 3区 医学 Q2 CLINICAL NEUROLOGY American Journal of Neuroradiology Pub Date : 2023-10-01 Epub Date: 2023-09-21 DOI:10.3174/ajnr.A7999
Zhengwu Tan, Lan Zhang, Xiaojie Sun, Ming Yang, Joyman Makamure, Hongying Wu, Jing Wang
{"title":"Dual-Layer Detector Head CT to Maintain Image Quality While Reducing the Radiation Dose in Pediatric Patients.","authors":"Zhengwu Tan, Lan Zhang, Xiaojie Sun, Ming Yang, Joyman Makamure, Hongying Wu, Jing Wang","doi":"10.3174/ajnr.A7999","DOIUrl":null,"url":null,"abstract":"<p><strong>Background and purpose: </strong>Radiation exposure in the CT diagnostic imaging process is a conspicuous concern in pediatric patients. This study aimed to evaluate whether 60-keV virtual monoenergetic images of the pediatric cranium in dual-layer CT can reduce the radiation dose while maintaining image quality compared with conventional images.</p><p><strong>Materials and methods: </strong>One hundred six unenhanced pediatric head scans acquired by dual-layer CT were retrospectively assessed. The patients were assigned to 2 groups of 53 and scanned with 250 and 180 mAs, respectively. Dose-length product values were retrieved, and noise, SNR, and contrast-to-noise ratio were calculated for each case. Two radiologists blinded to the reconstruction technique used evaluated image quality on a 5-point Likert scale. Statistical assessment was performed with ANOVA and the Wilcoxon test, adjusted for multiple comparisons.</p><p><strong>Results: </strong>Mean dose-length product values were 717.47 (SD, 41.52) mGy×cm and 520.74 (SD, 42) mGy×cm for the 250- and 180-mAs groups, respectively. Irrespective of the radiation dose, noise was significantly lower, SNR and contrast-to-noise ratio were significantly higher, and subjective analysis revealed significant superiority of 60-keV virtual monoenergetic images compared with conventional images (all <i>P</i> < .001). SNR, contrast-to-noise ratio, and subjective evaluation in 60-keV virtual monoenergetic images were not significantly different between the 2 scan groups (<i>P</i> > .05). Radiation dose parameters were significantly lower in the 180-mAs group compared with the 250-mAs group (<i>P</i> < .001).</p><p><strong>Conclusions: </strong>Dual-layer CT 60-keV virtual monoenergetic images allowed a radiation dose reduction of 28% without image-quality loss in pediatric cranial CT.</p>","PeriodicalId":7875,"journal":{"name":"American Journal of Neuroradiology","volume":null,"pages":null},"PeriodicalIF":3.1000,"publicationDate":"2023-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10549953/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"American Journal of Neuroradiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3174/ajnr.A7999","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/9/21 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"CLINICAL NEUROLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Background and purpose: Radiation exposure in the CT diagnostic imaging process is a conspicuous concern in pediatric patients. This study aimed to evaluate whether 60-keV virtual monoenergetic images of the pediatric cranium in dual-layer CT can reduce the radiation dose while maintaining image quality compared with conventional images.

Materials and methods: One hundred six unenhanced pediatric head scans acquired by dual-layer CT were retrospectively assessed. The patients were assigned to 2 groups of 53 and scanned with 250 and 180 mAs, respectively. Dose-length product values were retrieved, and noise, SNR, and contrast-to-noise ratio were calculated for each case. Two radiologists blinded to the reconstruction technique used evaluated image quality on a 5-point Likert scale. Statistical assessment was performed with ANOVA and the Wilcoxon test, adjusted for multiple comparisons.

Results: Mean dose-length product values were 717.47 (SD, 41.52) mGy×cm and 520.74 (SD, 42) mGy×cm for the 250- and 180-mAs groups, respectively. Irrespective of the radiation dose, noise was significantly lower, SNR and contrast-to-noise ratio were significantly higher, and subjective analysis revealed significant superiority of 60-keV virtual monoenergetic images compared with conventional images (all P < .001). SNR, contrast-to-noise ratio, and subjective evaluation in 60-keV virtual monoenergetic images were not significantly different between the 2 scan groups (P > .05). Radiation dose parameters were significantly lower in the 180-mAs group compared with the 250-mAs group (P < .001).

Conclusions: Dual-layer CT 60-keV virtual monoenergetic images allowed a radiation dose reduction of 28% without image-quality loss in pediatric cranial CT.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
双层探头CT在降低儿科患者辐射剂量的同时保持图像质量。
背景和目的:CT诊断成像过程中的辐射暴露是儿科患者关注的一个突出问题。本研究旨在评估与传统图像相比,双层CT中60 keV的儿童颅骨虚拟单能图像是否可以在保持图像质量的同时减少辐射剂量。材料和方法:对106例儿童头部双层CT平扫进行回顾性评价。将患者分为2组,每组53人,分别用250和180毫安进行扫描。检索剂量-长度乘积值,并计算每个病例的噪声、信噪比和对比噪声比。两名对重建技术不知情的放射科医生使用5点Likert量表评估了图像质量。采用方差分析和Wilcoxon检验进行统计评估,并对多重比较进行调整。结果:250和180 mAs组的平均剂量-长度乘积值分别为717.47(SD,41.52)mGy×cm和520.74(SD,42)mGy。无论辐射剂量如何,噪声都显著降低,信噪比和对比噪声比显著提高,主观分析显示60keV虚拟单能图像与传统图像相比具有显著优势(均P P > .05)。与250mAs组相比,180mAs组的辐射剂量参数显著降低(P<.001)。结论:双层CT 60keV虚拟单能图像使儿童颅骨CT的辐射剂量减少了28%,而图像质量没有损失。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
7.10
自引率
5.70%
发文量
506
审稿时长
2 months
期刊介绍: The mission of AJNR is to further knowledge in all aspects of neuroimaging, head and neck imaging, and spine imaging for neuroradiologists, radiologists, trainees, scientists, and associated professionals through print and/or electronic publication of quality peer-reviewed articles that lead to the highest standards in patient care, research, and education and to promote discussion of these and other issues through its electronic activities.
期刊最新文献
Artificial Intelligence-Generated Editorials in Radiology: Can Expert Editors Detect Them? The many faces of myxopapillary ependynomas. Vestibular schwannoma-related increased labyrinthine post-gadolinium 3D-FLAIR signal intensity and association with hearing impairment. CNS Embryonal Tumor with PLAGL Amplification, a New Tumor Type in Children and Adolescents: Insights from a Comprehensive MRI Analysis. DSA Quantitative Analysis and Predictive Modeling of Obliteration in Cerebral AVM following Stereotactic Radiosurgery.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1