Assessing the effectiveness of Compare Assistant for improving intra-rater reliability of ultrasound-measured muscle thickness.

IF 1.9 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Journal of Medical Ultrasonics Pub Date : 2024-01-01 Epub Date: 2023-10-07 DOI:10.1007/s10396-023-01367-y
Tyler D Baer, Kevin A Rice, Emely Urbina, Dominic V Whitener, Scott J Dankel
{"title":"Assessing the effectiveness of Compare Assistant for improving intra-rater reliability of ultrasound-measured muscle thickness.","authors":"Tyler D Baer, Kevin A Rice, Emely Urbina, Dominic V Whitener, Scott J Dankel","doi":"10.1007/s10396-023-01367-y","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Muscle thickness measured via ultrasound is commonly used to assess muscle size. The purpose of this study was to determine if the reliability of this measurement will improve if using the Compare Assistant tool, and whether this depends on technician experience and the muscle being assessed.</p><p><strong>Methods: </strong>Individuals came to the laboratory for two visits each separated by 24 h. On day 1, two ultrasound images were taken on the individual's anterior upper arm (elbow flexors) and anterior lower leg (tibialis anterior) by two inexperienced and one experienced ultrasound technician. On day 2, three images were taken: (1) without looking at the previous images taken on day 1; (2) after re-examining the images taken on day 1, and (3) side-by-side with the images taken on day 1 via Compare Assistant. Bayes Factors (BF<sub>10</sub>) were used to provide evidence for the null (< 0.33) or alternative (> 3) hypotheses.</p><p><strong>Results: </strong>There was no rater by measurement technique interaction (upper body: BF<sub>10</sub> = 0.04, lower body: BF<sub>10</sub> = 0.138), nor was there a main effect of measurement technique (upper body: BF<sub>10</sub> = 0.052, lower body: BF<sub>10</sub> = 0.331), indicating that reliability measures were not improved for either the upper body (CV%, no look: 2.92 vs. Compare Assistant: 2.87) or lower body (CV%, no look: 1.81 vs. Compare Assistant: 1.34) as a result of using Compare Assistant.</p><p><strong>Conclusion: </strong>The results of this study suggest that day-to-day reliability of muscle thickness measurement may be limited by random biological variability as opposed to technician error.</p>","PeriodicalId":50130,"journal":{"name":"Journal of Medical Ultrasonics","volume":null,"pages":null},"PeriodicalIF":1.9000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Medical Ultrasonics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s10396-023-01367-y","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/10/7 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: Muscle thickness measured via ultrasound is commonly used to assess muscle size. The purpose of this study was to determine if the reliability of this measurement will improve if using the Compare Assistant tool, and whether this depends on technician experience and the muscle being assessed.

Methods: Individuals came to the laboratory for two visits each separated by 24 h. On day 1, two ultrasound images were taken on the individual's anterior upper arm (elbow flexors) and anterior lower leg (tibialis anterior) by two inexperienced and one experienced ultrasound technician. On day 2, three images were taken: (1) without looking at the previous images taken on day 1; (2) after re-examining the images taken on day 1, and (3) side-by-side with the images taken on day 1 via Compare Assistant. Bayes Factors (BF10) were used to provide evidence for the null (< 0.33) or alternative (> 3) hypotheses.

Results: There was no rater by measurement technique interaction (upper body: BF10 = 0.04, lower body: BF10 = 0.138), nor was there a main effect of measurement technique (upper body: BF10 = 0.052, lower body: BF10 = 0.331), indicating that reliability measures were not improved for either the upper body (CV%, no look: 2.92 vs. Compare Assistant: 2.87) or lower body (CV%, no look: 1.81 vs. Compare Assistant: 1.34) as a result of using Compare Assistant.

Conclusion: The results of this study suggest that day-to-day reliability of muscle thickness measurement may be limited by random biological variability as opposed to technician error.

Abstract Image

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
评估Compare Assistant提高超声测量肌肉厚度评分器内可靠性的有效性。
目的:通过超声波测量肌肉厚度通常用于评估肌肉大小。本研究的目的是确定如果使用比较助手工具,该测量的可靠性是否会提高,以及这是否取决于技术人员的经验和正在评估的肌肉。方法:个体来到实验室进行两次访问,每次访问间隔24小时。在第1天,由两名经验不足和一名经验丰富的超声技师在个体的前上臂(肘部屈肌)和前小腿(胫骨前肌)上拍摄了两张超声图像。在第2天,拍摄了三张图像:(1)不看之前在第1天拍摄的图像;(2) 以及(3)通过Compare Assistant与第1天拍摄的图像并排。贝叶斯因子(BF10)用于为零( 3) 假设。结果:没有通过测量技术交互进行评分(上身:BF10 = 0.04,下体:BF10 = 0.138),也没有测量技术的主要影响(上身:BF10 = 0.052,下体:BF10 = 0.331),表明上半身(CV%,无表情:2.92 vs.Compare Assistant:2.87)或下半身(CV%,无眼神:1.81 vs。Compare Assistant:1.34)。结论:本研究的结果表明,肌肉厚度测量的日常可靠性可能受到随机生物变异性的限制,而不是技术人员的错误。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
3.30
自引率
11.10%
发文量
102
审稿时长
>12 weeks
期刊介绍: The Journal of Medical Ultrasonics is the official journal of the Japan Society of Ultrasonics in Medicine. The main purpose of the journal is to provide forum for the publication of papers documenting recent advances and new developments in the entire field of ultrasound in medicine and biology, encompassing both the medical and the engineering aspects of the science.The journal welcomes original articles, review articles, images, and letters to the editor.The journal also provides state-of-the-art information such as announcements from the boards and the committees of the society.
期刊最新文献
Correction: Principle of contrast-enhanced ultrasonography. Evaluation under loading detects medial meniscus extrusion in patients with reconstructed anterior cruciate ligament and restricted knee extension. Resolution of oval thrombus in a case of external jugular venous aneurysm. Quantification of the size of subchorionic hematoma causing pregnancy-related complications: a retrospective cohort study. Correction: Third-look contrast-enhanced ultrasonography plus needle biopsy for differential diagnosis of magnetic resonance imaging-only detected breast lesions.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1