Assessment of coagulation assays on Roche Cobas t711 analyzer: performance and clinical implications.

IF 1.2 4区 医学 Q4 HEMATOLOGY Blood Coagulation & Fibrinolysis Pub Date : 2023-12-01 Epub Date: 2023-10-11 DOI:10.1097/MBC.0000000000001261
María Ordóñez-Robles, Oscar D Pons-Belda, María José Moína, Ángel Bernardo-Gutiérrez, Belén Prieto-García
{"title":"Assessment of coagulation assays on Roche Cobas t711 analyzer: performance and clinical implications.","authors":"María Ordóñez-Robles, Oscar D Pons-Belda, María José Moína, Ángel Bernardo-Gutiérrez, Belén Prieto-García","doi":"10.1097/MBC.0000000000001261","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>We performed an analytical assessment of five coagulation tests [i.e. prothrombin time (PT), activated partial thromboplastin time (aPTT), fibrinogen, thrombin time (TT) and D-dimer] on the Roche Cobas t711 analyzer and a comparison study with the methodology in use at our laboratory (i.e. Werfen ACL Top 750 analyzer), expanding the analysis to the clinical implications of Cobas t711 implementation.</p><p><strong>Methods: </strong>Imprecision studies were performed following the Clinical and Laboratory Standards Institute (CLSI) H57 A:2008 guideline. Linearity of D-dimer and fibrinogen tests was analysed according to the CLSI EP06-A: 2003 recommendations. For method comparison, the results were analyzed using the Bland-Altman plot and Passing-Bablok regression.</p><p><strong>Results: </strong>Imprecision met manufacturer claims for PT, aPTT and TT. D-dimer and fibrinogen tests showed a coefficient of variation (CV)% over manufacturer claims at certain concentration levels. Linearity ranges could not be verified. Comparison study revealed that results are not interchangeable for any test, a lower correlation for aPTT test and lower D-dimer results from Roche Cobas t711.</p><p><strong>Conclusion: </strong>The strength of this study relies on the analysis of the clinical implications of reporting Cobas t711 results compared to those obtained with the methodology in use at our laboratory. Different sensibility to factor deficiency, anticoagulant therapy and interferences might explain lower correlation rates obtained for the aPTT test. Different monoclonal antibodies used for D-dimer determination might explain the lower results obtained with the Cobas t711 analyzer. This aspect needs further studies given the relevance of D-dimer test to exclude thrombotic events and reinforces the need of harmonization in the haemostasis laboratory.</p>","PeriodicalId":8992,"journal":{"name":"Blood Coagulation & Fibrinolysis","volume":null,"pages":null},"PeriodicalIF":1.2000,"publicationDate":"2023-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood Coagulation & Fibrinolysis","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MBC.0000000000001261","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/10/11 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"HEMATOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Objectives: We performed an analytical assessment of five coagulation tests [i.e. prothrombin time (PT), activated partial thromboplastin time (aPTT), fibrinogen, thrombin time (TT) and D-dimer] on the Roche Cobas t711 analyzer and a comparison study with the methodology in use at our laboratory (i.e. Werfen ACL Top 750 analyzer), expanding the analysis to the clinical implications of Cobas t711 implementation.

Methods: Imprecision studies were performed following the Clinical and Laboratory Standards Institute (CLSI) H57 A:2008 guideline. Linearity of D-dimer and fibrinogen tests was analysed according to the CLSI EP06-A: 2003 recommendations. For method comparison, the results were analyzed using the Bland-Altman plot and Passing-Bablok regression.

Results: Imprecision met manufacturer claims for PT, aPTT and TT. D-dimer and fibrinogen tests showed a coefficient of variation (CV)% over manufacturer claims at certain concentration levels. Linearity ranges could not be verified. Comparison study revealed that results are not interchangeable for any test, a lower correlation for aPTT test and lower D-dimer results from Roche Cobas t711.

Conclusion: The strength of this study relies on the analysis of the clinical implications of reporting Cobas t711 results compared to those obtained with the methodology in use at our laboratory. Different sensibility to factor deficiency, anticoagulant therapy and interferences might explain lower correlation rates obtained for the aPTT test. Different monoclonal antibodies used for D-dimer determination might explain the lower results obtained with the Cobas t711 analyzer. This aspect needs further studies given the relevance of D-dimer test to exclude thrombotic events and reinforces the need of harmonization in the haemostasis laboratory.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Roche Cobas t711分析仪凝血测定的评估:性能和临床意义。
目的:我们在Roche Cobas t711分析仪上对五项凝血测试[即凝血酶原时间(PT)、活化部分凝血活酶时间(aPTT)、纤维蛋白原、凝血酶时间(TT)和D-二聚体]进行了分析评估,并与我们实验室使用的方法(即Werfen ACL Top 750分析仪)进行了比较研究,将分析扩展到Cobas t711实施的临床意义。方法:根据临床和实验室标准研究所(CLSI)H57A:2008指南进行不精确性研究。根据CLSI EP06-A:2003的建议分析D-二聚体和纤维蛋白原测试的线性。为了进行方法比较,使用Bland-Altman图和Passing-Balok回归分析结果。结果:不精确性符合制造商对PT、aPTT和TT的要求。D-二聚体和纤维蛋白原测试显示,在某些浓度水平下,与制造商的要求相比,变异系数(CV)为%。无法验证线性范围。比较研究表明,任何测试的结果都不能互换,aPTT测试的相关性较低,罗氏Cobas t711的D-二聚体结果较低。结论:与我们实验室使用的方法相比,本研究的优势在于分析报告Cobas t 711结果的临床意义。对因子缺乏、抗凝治疗和干扰的不同敏感性可能解释了aPTT测试获得的较低相关性。用于D-二聚体测定的不同单克隆抗体可能解释了Cobas t711分析仪获得的较低结果。鉴于D-二聚体检测与排除血栓事件的相关性,这方面需要进一步研究,并加强了止血实验室的协调需求。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
1.90
自引率
0.00%
发文量
111
审稿时长
4-8 weeks
期刊介绍: Blood Coagulation & Fibrinolysis is an international fully refereed journal that features review and original research articles on all clinical, laboratory and experimental aspects of haemostasis and thrombosis. The journal is devoted to publishing significant developments worldwide in the field of blood coagulation, fibrinolysis, thrombosis, platelets and the kininogen-kinin system, as well as dealing with those aspects of blood rheology relevant to haemostasis and the effects of drugs on haemostatic components
期刊最新文献
Variation among venous thromboembolism risk assessment tools for postcesarean patients: a retrospective cohort study. Is the combination of two automated rapid assays for diagnosis of heparin-induced thrombocytopenia necessary? A novel SERPINC1 c.119G>A (p.Cys40Tyr) mutation with variable clinical expression in an Indian family. A novel γ-chain mutation p.Asp318His in a Chinese family with dysfibrinogenemia. Molecular mechanism analysis of a family with hereditary coagulation FXI deficiency caused by compound heterozygous mutations.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1