Comparative Evaluation of Canal Transportation and Centering Ratio in Curved Canals: A Study of Cone-beam Computed Tomography and Micro-computed Tomography.
Yazdan Shantiaee, Babak Zandi, Ali Rahbar Taramsari, Alireza Akbarzadeh Baghban, Nazanin Zargar, Shiva Shojaeian, Niloofar Mortezapour, Kourosh Shantiaee, Fatemeh Soltaninejad
{"title":"Comparative Evaluation of Canal Transportation and Centering Ratio in Curved Canals: A Study of Cone-beam Computed Tomography and Micro-computed Tomography.","authors":"Yazdan Shantiaee, Babak Zandi, Ali Rahbar Taramsari, Alireza Akbarzadeh Baghban, Nazanin Zargar, Shiva Shojaeian, Niloofar Mortezapour, Kourosh Shantiaee, Fatemeh Soltaninejad","doi":"10.22037/iej.v18i4.41541","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>This study aimed to compare the accuracy and agreement between cone-beam computed tomography (CBCT) and micro-computed tomography (micro-CT) in the assessment of canal transportation and centering ratio following root canal instrumentation with rotary files.</p><p><strong>Material and methods: </strong>Twenty mesiobuccal canals of mandibular molars were prepared using the 2Shape sequential rotary system. CBCT and micro-CT scans were performed before and after instrumentation, and the magnitude of transportation and centering ratio were measured. The acceptable transportation was set at ≤0.15 mm. The accuracy and agreement between CBCT and micro-CT were calculated, and the intra-class correlation coefficient (ICC) and kappa coefficient were determined to assess the agreement between the two modalities. Statistical analyses were performed using repeated measures ANOVA.</p><p><strong>Results: </strong>Transportation was detected by both modalities at all distances from the apex after instrumentation. The agreement between CBCT and micro-CT in assessing canal transportation was observed in 80%, 85%, 75%, and 75% of specimens at 1-, 3-, 5-, and 7-mm from the apex, respectively. The ICC for transportation and centering ratio was much lower than 0.75, indicating poor agreement between the modalities. The kappa coefficient did not show acceptable agreement between the methods.</p><p><strong>Conclusions: </strong>CBCT and micro-CT demonstrated poor agreement in assessing canal transportation and centering ratio. Micro-CT remains the preferred modality for in vitro investigations, while CBCT should be limited to clinical settings.</p>","PeriodicalId":14534,"journal":{"name":"Iranian Endodontic Journal","volume":"18 4","pages":"241-247"},"PeriodicalIF":0.0000,"publicationDate":"2023-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/32/78/IEJ-18-241.PMC10566004.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Iranian Endodontic Journal","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.22037/iej.v18i4.41541","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Dentistry","Score":null,"Total":0}
引用次数: 0
Abstract
Introduction: This study aimed to compare the accuracy and agreement between cone-beam computed tomography (CBCT) and micro-computed tomography (micro-CT) in the assessment of canal transportation and centering ratio following root canal instrumentation with rotary files.
Material and methods: Twenty mesiobuccal canals of mandibular molars were prepared using the 2Shape sequential rotary system. CBCT and micro-CT scans were performed before and after instrumentation, and the magnitude of transportation and centering ratio were measured. The acceptable transportation was set at ≤0.15 mm. The accuracy and agreement between CBCT and micro-CT were calculated, and the intra-class correlation coefficient (ICC) and kappa coefficient were determined to assess the agreement between the two modalities. Statistical analyses were performed using repeated measures ANOVA.
Results: Transportation was detected by both modalities at all distances from the apex after instrumentation. The agreement between CBCT and micro-CT in assessing canal transportation was observed in 80%, 85%, 75%, and 75% of specimens at 1-, 3-, 5-, and 7-mm from the apex, respectively. The ICC for transportation and centering ratio was much lower than 0.75, indicating poor agreement between the modalities. The kappa coefficient did not show acceptable agreement between the methods.
Conclusions: CBCT and micro-CT demonstrated poor agreement in assessing canal transportation and centering ratio. Micro-CT remains the preferred modality for in vitro investigations, while CBCT should be limited to clinical settings.
期刊介绍:
The Iranian Endodontic Journal (IEJ) is an international peer-reviewed biomedical publication, the aim of which is to provide a scientific medium of communication for researchers throughout the globe. IEJ aims to publish the highest quality articles, both clinical and scientific, on all aspects of Endodontics. The journal is an official Journal of the Iranian Center for Endodontic Research (ICER) and the Iranian Association of Endodontists (IAE). The Journal welcomes articles related to the scientific or applied aspects of endodontics e.g. original researches, systematic reviews, meta-analyses, review articles, clinical trials, case series/reports, hypotheses, letters to the editor, etc. From the beginning (i.e. since 2006), the IEJ was the first open access endodontic journal in the world, which gave readers free and instant access to published articles and enabling them faster discovery of the latest endodontic research.