Yu Zhang, W. Dai, can-can Lin, Qiao-yuan Li, Chengjun Guo
{"title":"Implantation of an S-ICD in a Patient with a DDD Pacemaker and Congenitally Corrected Transposition of the Great Arteries","authors":"Yu Zhang, W. Dai, can-can Lin, Qiao-yuan Li, Chengjun Guo","doi":"10.15212/cvia.2019.0597","DOIUrl":null,"url":null,"abstract":"Subcutaneous implantable cardioverter-defibrillator (S-ICD) therapy has become a viable alternative to conventional transvenous ICD implantation. Patients with congenitally corrected transposition of the great arteries (ccTGA) have a high risk of sudden cardiac death due to malignant\n arrhythmia. The interaction between the S-ICD system and the transvenous pacemaker system is not fully understood. We report a case of S-ICD implantation in a patient with ccTGA and a DDD pacemaker. The patient was a 30-year-old man with a previously placed pacemaker with diagnoses of congenital\n heart disease, ccTGA (SLL), left atrioventricular valve insufficiency, and third-degree atrioventricular block. He presented with an out-of-hospital cardiac arrest, and an S-ICD was implanted to prevent sudden cardiac death. Defibrillation checks were performed successfully. We tested the\n compatibility of the DDD pacemaker with the S-ICD and found that there was no interference between them. In conclusion, an S-ICD system is a reasonable and safe option in patients with ccTGA.","PeriodicalId":41559,"journal":{"name":"Cardiovascular Innovations and Applications","volume":"1 1","pages":""},"PeriodicalIF":0.9000,"publicationDate":"2021-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Cardiovascular Innovations and Applications","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.15212/cvia.2019.0597","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Subcutaneous implantable cardioverter-defibrillator (S-ICD) therapy has become a viable alternative to conventional transvenous ICD implantation. Patients with congenitally corrected transposition of the great arteries (ccTGA) have a high risk of sudden cardiac death due to malignant
arrhythmia. The interaction between the S-ICD system and the transvenous pacemaker system is not fully understood. We report a case of S-ICD implantation in a patient with ccTGA and a DDD pacemaker. The patient was a 30-year-old man with a previously placed pacemaker with diagnoses of congenital
heart disease, ccTGA (SLL), left atrioventricular valve insufficiency, and third-degree atrioventricular block. He presented with an out-of-hospital cardiac arrest, and an S-ICD was implanted to prevent sudden cardiac death. Defibrillation checks were performed successfully. We tested the
compatibility of the DDD pacemaker with the S-ICD and found that there was no interference between them. In conclusion, an S-ICD system is a reasonable and safe option in patients with ccTGA.