Gülay Yalçınkaya, O. B. Ocak, I. Cakir, A. Inal, B. Gökyiğit
{"title":"Comparison of Two Different Photorefractors with Skiascopy Measurements in Healthy Children","authors":"Gülay Yalçınkaya, O. B. Ocak, I. Cakir, A. Inal, B. Gökyiğit","doi":"10.4274/jarem.galenos.2021.87597","DOIUrl":null,"url":null,"abstract":"Objective: Comparing refractive values determined by two photorefractory devices (Plusoptix® A09, GmbH, Nuremberg, Germany and SureSight®, Welch Allyn Co, New York, USA) with refractive values obtained by skiascope after cycloplegia with 1% cyclopentolate hydrochloride in healthy children. Methods: Cases aged between 48 and 132 months were evaluated for this cross-sectional study. Cases with no ophthalmic pathology in both eyes and uncorrected visual acuity of 1.0 were included in the study. In all cases, refractive measurements were performed first with Plusoptix® and then with SureSight®. After providing cycloplegia with 1% cyclopentolate hydrochloride, skiascopy was performed. The correlation between the measurements obtained was evaluated with the Pearson correlation coefficient, and the compatibility was evaluated with the Bland-Altman analysis. Results: The mean spherical values of 52 subjects measured by cycloplegic skiascopy, Plusoptix® and SureSight® were 0.88±2.07 dioptry (D), 0.69±1.78D and 1.64±1.06D, respectively; mean of cylindrical measurements are -0.73±0.68D, -0.92±0.67D and -0.83±0.70D, respectively. Comparing Plusoptix® with skiascopy, SureSight® with skiascopy, Plusoptix® with SureSight® measurements, a high positive correlation was found between spherical values (r=0.861, r=0.736, r=0.721, respectively); a positive correlation was also found between cylindrical values (r=0.602, r=0.675, r=0.901, respectively). Skiascopic spherical and cylindrical measurements with Plusoptix®, and skiascopic spherical and cylindrical measurements with SureSight® were found to be compatible with each other (within 95% confidence interval, lower limit: -2.65, upper limit: 1.89; lower limit: -1.35, upper limit: 0.95; lower limit: -3.56, upper limit: 2.06; lower limit: -1.20, upper limit: 0.98). Spherical and cylindrical measurements obtained with Plusoptix® and SureSight® were found to be consistent with each other (within 95% confidence interval, lower limit: -3.39, upper limit: 1.49; lower limit: -0.67, upper limit: 0.51). ÖZ Amaç: Sağlıklı çocuklarda, iki fotorefraktör cihaz (Plusoptix® A09, GmbH, Nuremberg, Almanya ve SureSight®, Welch Allyn Co, New York, ABD) ile saptanan refraktif değerleri, %1’lik siklopentolat hidrokloridle siklopleji sonrası skiaskopiyle elde edilen refraktif değerlerle karşılaştırmaktır. Yöntemler: Kesitsel çalışma için, yaşları 48 ile 132 ay arasındaki olgular değerlendirildi. Her iki gözde oftalmik patoloji saptanmayan ve düzeltilmemiş görme keskinliği 1,0 saptanan olgular çalışmaya alındı. Tüm olgularda önce Plusoptix®, ardından SureSight® ile refraktif ölçümler yapıldı. %1’lik siklopentolat hidrokloridle siklopleji sağlandıktan sonra skiaskopi gerçekleştirildi. Elde edilen ölçümler arasındaki ilişki Pearson korelasyon katsayısıyla, uyumluluksa Bland-Altman analiziyle değerlendirildi. Bulgular: Elli iki olgunun sikloplejik skiaskopi, Plusoptix® ve SureSight® ile ölçülen sferik değerlerinin ortalamaları sırasıyla 0.88±2.07 diyoptri (D), 0,69±1,78D ve 1,64±1,06D; silindirik ölçümlerinin ortalamalarıysa sırasıyla -0,73±0,68D, -0,92±0,67D ve -0,83±0,70D’dir. Plusoptix® ile skiaskopi, SureSight® ile skiaskopi, Plusoptix® ile SureSight® ölçümleri karşılaştırıldığında sferik değerler arasında yüksek pozitif korelasyon saptanmıştır (sırasıyla r=0,861, r=0,736, r=0,721); silindirik değerler arasında da pozitif korelasyon bulunmuştur (sırasıyla r=0,602, r=0,675, r=0,901). Plusoptix® ile skiaskopik sferik ve silindirik ölçümler, SureSight® ile skiaskopik sferik ve silindirik ölçümler birbirleriyle uyumlu saptanmıştır (%95 güven aralığı içinde sırasıyla alt sınır: -2,65, üst sınır: 1,89; alt sınır: -1,35, üst sınır: 0,95; alt sınır: -3,56, üst sınır: 2,06; alt sınır: -1,20, üst sınır: 0,98). Plusoptix® ve SureSight® ile elde edilen sferik ve silindirik ölçümler birbiriyle uyumlu bulunmuştur (%95 güven aralığı içinde sırasıyla alt sınır: -3,39, üst sınır: 1,49; alt sınır: -0,67, üst sınır: 0,51). Sonuç: Çocukluk çağında bu üç ölçümün de sonuçlarının birbiriyle uyumlu olduğu sonucuna ulaşıldı. Anahtar kelimeler: Fotorefraktometre, Plusoptix®, retinoskopi, sikloplejik skiaskopi, SureSight® Geliş Tarihi/Received Date: 13.01.2021 Kabul Tarihi/Accepted Date: 25.06.2021 Sorumlu Yazar/Corresponding Author: Gülay Yalçınkaya, E-posta: ykgulay@gmail.com DOI: 10.4274/jarem.galenos.2021.87597 Yalçınkaya ve ark. İki Ayrı Fotorefraktör ve Skiaskopi Orijinal Araştırma / Original Investigation Yalçınkaya ve ark. İki Ayrı Fotorefraktör ve Skiaskopi J Acad Res Med 2021;11(2):206-10 207 Conclusion: It was concluded that the results of these three measurements in childhood were compatible with each other.","PeriodicalId":56162,"journal":{"name":"Journal of Academic Research in Medicine-JAREM","volume":" ","pages":""},"PeriodicalIF":0.1000,"publicationDate":"2021-08-27","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Academic Research in Medicine-JAREM","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4274/jarem.galenos.2021.87597","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"MEDICINE, GENERAL & INTERNAL","Score":null,"Total":0}
引用次数: 1
Abstract
Objective: Comparing refractive values determined by two photorefractory devices (Plusoptix® A09, GmbH, Nuremberg, Germany and SureSight®, Welch Allyn Co, New York, USA) with refractive values obtained by skiascope after cycloplegia with 1% cyclopentolate hydrochloride in healthy children. Methods: Cases aged between 48 and 132 months were evaluated for this cross-sectional study. Cases with no ophthalmic pathology in both eyes and uncorrected visual acuity of 1.0 were included in the study. In all cases, refractive measurements were performed first with Plusoptix® and then with SureSight®. After providing cycloplegia with 1% cyclopentolate hydrochloride, skiascopy was performed. The correlation between the measurements obtained was evaluated with the Pearson correlation coefficient, and the compatibility was evaluated with the Bland-Altman analysis. Results: The mean spherical values of 52 subjects measured by cycloplegic skiascopy, Plusoptix® and SureSight® were 0.88±2.07 dioptry (D), 0.69±1.78D and 1.64±1.06D, respectively; mean of cylindrical measurements are -0.73±0.68D, -0.92±0.67D and -0.83±0.70D, respectively. Comparing Plusoptix® with skiascopy, SureSight® with skiascopy, Plusoptix® with SureSight® measurements, a high positive correlation was found between spherical values (r=0.861, r=0.736, r=0.721, respectively); a positive correlation was also found between cylindrical values (r=0.602, r=0.675, r=0.901, respectively). Skiascopic spherical and cylindrical measurements with Plusoptix®, and skiascopic spherical and cylindrical measurements with SureSight® were found to be compatible with each other (within 95% confidence interval, lower limit: -2.65, upper limit: 1.89; lower limit: -1.35, upper limit: 0.95; lower limit: -3.56, upper limit: 2.06; lower limit: -1.20, upper limit: 0.98). Spherical and cylindrical measurements obtained with Plusoptix® and SureSight® were found to be consistent with each other (within 95% confidence interval, lower limit: -3.39, upper limit: 1.49; lower limit: -0.67, upper limit: 0.51). ÖZ Amaç: Sağlıklı çocuklarda, iki fotorefraktör cihaz (Plusoptix® A09, GmbH, Nuremberg, Almanya ve SureSight®, Welch Allyn Co, New York, ABD) ile saptanan refraktif değerleri, %1’lik siklopentolat hidrokloridle siklopleji sonrası skiaskopiyle elde edilen refraktif değerlerle karşılaştırmaktır. Yöntemler: Kesitsel çalışma için, yaşları 48 ile 132 ay arasındaki olgular değerlendirildi. Her iki gözde oftalmik patoloji saptanmayan ve düzeltilmemiş görme keskinliği 1,0 saptanan olgular çalışmaya alındı. Tüm olgularda önce Plusoptix®, ardından SureSight® ile refraktif ölçümler yapıldı. %1’lik siklopentolat hidrokloridle siklopleji sağlandıktan sonra skiaskopi gerçekleştirildi. Elde edilen ölçümler arasındaki ilişki Pearson korelasyon katsayısıyla, uyumluluksa Bland-Altman analiziyle değerlendirildi. Bulgular: Elli iki olgunun sikloplejik skiaskopi, Plusoptix® ve SureSight® ile ölçülen sferik değerlerinin ortalamaları sırasıyla 0.88±2.07 diyoptri (D), 0,69±1,78D ve 1,64±1,06D; silindirik ölçümlerinin ortalamalarıysa sırasıyla -0,73±0,68D, -0,92±0,67D ve -0,83±0,70D’dir. Plusoptix® ile skiaskopi, SureSight® ile skiaskopi, Plusoptix® ile SureSight® ölçümleri karşılaştırıldığında sferik değerler arasında yüksek pozitif korelasyon saptanmıştır (sırasıyla r=0,861, r=0,736, r=0,721); silindirik değerler arasında da pozitif korelasyon bulunmuştur (sırasıyla r=0,602, r=0,675, r=0,901). Plusoptix® ile skiaskopik sferik ve silindirik ölçümler, SureSight® ile skiaskopik sferik ve silindirik ölçümler birbirleriyle uyumlu saptanmıştır (%95 güven aralığı içinde sırasıyla alt sınır: -2,65, üst sınır: 1,89; alt sınır: -1,35, üst sınır: 0,95; alt sınır: -3,56, üst sınır: 2,06; alt sınır: -1,20, üst sınır: 0,98). Plusoptix® ve SureSight® ile elde edilen sferik ve silindirik ölçümler birbiriyle uyumlu bulunmuştur (%95 güven aralığı içinde sırasıyla alt sınır: -3,39, üst sınır: 1,49; alt sınır: -0,67, üst sınır: 0,51). Sonuç: Çocukluk çağında bu üç ölçümün de sonuçlarının birbiriyle uyumlu olduğu sonucuna ulaşıldı. Anahtar kelimeler: Fotorefraktometre, Plusoptix®, retinoskopi, sikloplejik skiaskopi, SureSight® Geliş Tarihi/Received Date: 13.01.2021 Kabul Tarihi/Accepted Date: 25.06.2021 Sorumlu Yazar/Corresponding Author: Gülay Yalçınkaya, E-posta: ykgulay@gmail.com DOI: 10.4274/jarem.galenos.2021.87597 Yalçınkaya ve ark. İki Ayrı Fotorefraktör ve Skiaskopi Orijinal Araştırma / Original Investigation Yalçınkaya ve ark. İki Ayrı Fotorefraktör ve Skiaskopi J Acad Res Med 2021;11(2):206-10 207 Conclusion: It was concluded that the results of these three measurements in childhood were compatible with each other.