Dosimetric analysis of treatment planning optimization techniques for CO-60 and IR-192 HDR brachytherapy sources in carcinoma cervix patients

Q4 Medicine Libri Oncologici Pub Date : 2022-05-12 DOI:10.20471/lo.2022.50.01.02
Rajni Verma, Gourav Kumar Jain, A. Chougule
{"title":"Dosimetric analysis of treatment planning optimization techniques for CO-60 and IR-192 HDR brachytherapy sources in carcinoma cervix patients","authors":"Rajni Verma, Gourav Kumar Jain, A. Chougule","doi":"10.20471/lo.2022.50.01.02","DOIUrl":null,"url":null,"abstract":"Summary Introduction: the study was designed to dosimetrically compare the treatment planning optimization techniques to analyze the equivalence of co-60 and ir-192 HDr Brachytherapy sources. further, the present study planned to analyze the equivalence on the basis of technical, logistical and clinical aspects for both the sources. Methods: Thirty-two patients with confirmed diagnosis of carcinoma cervix were included in the study. Comparative treatment plans were made for three different treatment planning optimization techniques for both the sources Co-60 and ir-192 i.e. (a) plans with prescription on point a (Manchester system) with identical dwell position, (b) plans with inverse planning and (c) plans with prescription on Hr-ctV with manual optimization. treatment plans were evaluated on the basis of clinical parameters Hr-ctV V200%, V150%, V100%, D95%, D50%, D2cc Bladder, D2cc rectum and point dose on icrU rectum and Bladder reference point. Results: For the plans with prescription on Point A with identical dwell position, the average percentage difference were found Hr-ctV V200% (6.3%), V150% (5.1%), V100% (1.8%), D2cc rectum (2.3%) and icrU rectum reference point (2.1%) and all these parameters were statistically significant for both the sources. For the Plans with inverse planning opti -mization, these parameters were found significant with average percentage difference HR-CTV V200% (6.1%), V150% (4.9%), D50% (2.1%) and icrU rectum reference point (2.3%). for the plans with dose prescription on Hr-ctV and manual optimization, most of the average percentage differences were found non-significant and clinical parameters were ob served clinically comparable to each other for both the sources. further, it was also observed that on an average ir-192 source required only 42% of the treatment time required by co-60 for the same treatment plan delivery. Conclusion: among all the three planning techniques, the planning technique with prescription on Hr-ctV & manual optimization was found to have comparable clinical quality for both the sources. this analysis revealed that geometry of source placement can overcome the differences in individual source characteristics.","PeriodicalId":53700,"journal":{"name":"Libri Oncologici","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2022-05-12","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Libri Oncologici","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.20471/lo.2022.50.01.02","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 1

Abstract

Summary Introduction: the study was designed to dosimetrically compare the treatment planning optimization techniques to analyze the equivalence of co-60 and ir-192 HDr Brachytherapy sources. further, the present study planned to analyze the equivalence on the basis of technical, logistical and clinical aspects for both the sources. Methods: Thirty-two patients with confirmed diagnosis of carcinoma cervix were included in the study. Comparative treatment plans were made for three different treatment planning optimization techniques for both the sources Co-60 and ir-192 i.e. (a) plans with prescription on point a (Manchester system) with identical dwell position, (b) plans with inverse planning and (c) plans with prescription on Hr-ctV with manual optimization. treatment plans were evaluated on the basis of clinical parameters Hr-ctV V200%, V150%, V100%, D95%, D50%, D2cc Bladder, D2cc rectum and point dose on icrU rectum and Bladder reference point. Results: For the plans with prescription on Point A with identical dwell position, the average percentage difference were found Hr-ctV V200% (6.3%), V150% (5.1%), V100% (1.8%), D2cc rectum (2.3%) and icrU rectum reference point (2.1%) and all these parameters were statistically significant for both the sources. For the Plans with inverse planning opti -mization, these parameters were found significant with average percentage difference HR-CTV V200% (6.1%), V150% (4.9%), D50% (2.1%) and icrU rectum reference point (2.3%). for the plans with dose prescription on Hr-ctV and manual optimization, most of the average percentage differences were found non-significant and clinical parameters were ob served clinically comparable to each other for both the sources. further, it was also observed that on an average ir-192 source required only 42% of the treatment time required by co-60 for the same treatment plan delivery. Conclusion: among all the three planning techniques, the planning technique with prescription on Hr-ctV & manual optimization was found to have comparable clinical quality for both the sources. this analysis revealed that geometry of source placement can overcome the differences in individual source characteristics.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
宫颈癌患者CO-60和IR-192 HDR近距离放疗源治疗方案优化技术的剂量学分析
摘要简介:本研究旨在对治疗计划优化技术进行剂量比较,以分析co-60和ir-192 HDr近距离治疗来源的等效性。此外,本研究计划从技术、后勤和临床方面分析这两种来源的等效性。方法:将32例确诊为宫颈癌的患者纳入研究。对Co-60和ir-192来源的三种不同治疗计划优化技术进行了比较治疗计划,即(a)在a点(曼彻斯特系统)上具有相同停留位置的处方的计划,(b)具有反向计划的计划,以及(c)在Hr ctV上具有手动优化处方的计划。根据临床参数Hr-ctV V200%、V150%、V100%、D95%、D50%、D2cc膀胱、D2cc直肠以及icrU直肠上的点剂量和膀胱参考点来评估治疗方案。结果:对于停药位置相同的A点处方方案,Hr ctV V200%(6.3%)、V150%(5.1%)、V100%(1.8%)、D2cc直肠(2.3%)和icrU直肠参考点(2.1%)的平均百分比差异,所有这些参数对两个来源都具有统计学意义。对于具有反向计划优化的计划,这些参数具有显著性,平均百分比差异为HR-CTV V200%(6.1%)、V150%(4.9%)、D50%(2.1%)和icrU直肠参考点(2.3%),大多数平均百分比差异被发现没有显著性,并且两种来源的临床参数在临床上具有可比性。此外,还观察到,对于相同的治疗计划,平均而言,ir-192来源只需要co-60所需治疗时间的42%。结论:在所有三种计划技术中,Hr-ctV处方和手动优化的计划技术对两种来源具有可比的临床质量。该分析表明,源的几何布局可以克服单个源特性的差异。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Libri Oncologici
Libri Oncologici Medicine-Oncology
CiteScore
0.30
自引率
0.00%
发文量
9
审稿时长
8 weeks
期刊介绍: - Genitourinary cancer: the potential role of imaging - Hemoglobin level and neoadjuvant chemoradiation in patients with locally advanced cervical carcinoma
期刊最新文献
Results of multicenter testing of PIK3CA somatic mutations in hormone-receptor positive HER2-negative advanced breast cancer Use of nasolabial flap for reconstruction of the floor of the mouth defects Hereditary breast and ovarian cancer – University Hospital of Split experiences Breast cancer radiotherapy - changes in fractionation schemes through decades Latest perspectives on the benefits of neoadjuvant therapy for patients with advanced gastric cancer
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1