Usefulness of Hammering Sound Frequency Analysis as an Evaluation Method for the Prevention of Trouble during Hip Replacement

R. Sakai, K. Uchiyama, N. Takahira, M. Kakeshita, Y. Otsu, Kazuhiro Yoshida, M. Ujihira
{"title":"Usefulness of Hammering Sound Frequency Analysis as an Evaluation Method for the Prevention of Trouble during Hip Replacement","authors":"R. Sakai, K. Uchiyama, N. Takahira, M. Kakeshita, Y. Otsu, Kazuhiro Yoshida, M. Ujihira","doi":"10.4236/jbise.2020.135007","DOIUrl":null,"url":null,"abstract":"In total hip arthroplasty, judgment of the appropriateness of stem hammering is dependent on the experience and feelings of the surgeon and no objective evaluation method has been established. In this study, a frequency analysis of the hammering sounds in total hip arthroplasty was performed to investigate objective judgment criteria capable of preventing problems during surgery. Stem hammering was applied following the surgeon’s feelings as usual in an operating room. A directional microphone was placed at a distance about 2 m from the surgical field and the peak frequency reaching the maximum amplitude was determined by Fourier analysis. It was clarified that the same peak frequency repeats when appropriate fixation is acquired during surgery, suggesting that intraoperative fracture and postoperative loosening can be prevented by stopping hammering at the time the peak frequency converged. Investigation of changes in the hammering sound frequency may serve as objective judgment criteria capable of preventing problems during surgery.","PeriodicalId":64231,"journal":{"name":"生物医学工程(英文)","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2020-05-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"4","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"生物医学工程(英文)","FirstCategoryId":"1087","ListUrlMain":"https://doi.org/10.4236/jbise.2020.135007","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 4

Abstract

In total hip arthroplasty, judgment of the appropriateness of stem hammering is dependent on the experience and feelings of the surgeon and no objective evaluation method has been established. In this study, a frequency analysis of the hammering sounds in total hip arthroplasty was performed to investigate objective judgment criteria capable of preventing problems during surgery. Stem hammering was applied following the surgeon’s feelings as usual in an operating room. A directional microphone was placed at a distance about 2 m from the surgical field and the peak frequency reaching the maximum amplitude was determined by Fourier analysis. It was clarified that the same peak frequency repeats when appropriate fixation is acquired during surgery, suggesting that intraoperative fracture and postoperative loosening can be prevented by stopping hammering at the time the peak frequency converged. Investigation of changes in the hammering sound frequency may serve as objective judgment criteria capable of preventing problems during surgery.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
锤击声频率分析作为髋关节置换术中预防问题的评估方法的有用性
在全髋关节置换术中,对锤击柄是否合适的判断依赖于术者的经验和感受,目前尚无客观的评价方法。在本研究中,对全髋关节置换术中锤击声的频率进行了分析,以探讨能够在手术中预防问题的客观判断标准。在手术室里,按照外科医生的感觉,像往常一样使用茎锤。在距离手术场约2 m处放置定向麦克风,通过傅里叶分析确定达到最大振幅的峰值频率。研究表明,在手术中获得适当的固定时,相同的峰值频率会重复出现,这表明可以通过在峰值频率汇聚时停止锤击来防止术中骨折和术后松动。研究锤击声频率的变化可以作为预防手术中出现问题的客观判断标准。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
900
期刊最新文献
Tectona grandis (Teak Tree) Young Leaf Extract as a Histological Stain An Alternative Method for Incorporating Fiber Meshes in Complete Upper Dentures Research on the Chemical Hazard Risk of Toys Exported to EU from China Based on the Analysis of the EU “Safety Gate” Alerts Cases Impact of Association between Functional Training and Respiratory Muscle Training in Elderly: A Randomized Controlled Trial Morphogenesis of Floating Bone Segments: A Legacy of Serial Tensile Cross-Strut Microdamage in Trabecular Disconnection “Crumple Zones”?
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1