{"title":"Application of ArcCHECK system in the verification of tomotherapy plan for esophageal cancer","authors":"Zhenli Wang, A. Zheng, Jian Zhu, H. Du, Xiaodong Sun, Dong Wang, Huitao Wang, Yaowen Zhang, Qingshan Zhu, Z. Cui","doi":"10.3760/CMA.J.CN113030-20181221-00005","DOIUrl":null,"url":null,"abstract":"Objective \nTo explore the application of ArcCheck system in the validation of Helical and Direct tomotherapy plans for esophageal cancer and summarize relevant experience. \n \n \nMethods \nThe Helical and Direct tomotherapy verification plans were established for 32 patients with esophageal cancer at different positions according to the doctor′s instructions, which were verified by the ArcCHECK system to compare the passing rate of the results.The correlation between the volume of the target area and the passing rate of the planned verification was analyzed. The therapeutic verification plan with a small target volume was made. The target area was placed at the center of ArcCHECK phantom and the area of detectors to statistically compare the verification passing rates. \n \n \nResults \nHelical plan showed a significantly higher passing rate than the Direct plan (P 0.05). \n \n \nConclusions \nThe passing rate of the Helical plan is generally higher than that of the Direct plan, which may be related to the angular response of the ArcCHECK detector and the fact that more reference points are not included for calculation due to low-dose radiation. In addition, it may also be related to the higher requirements of Direct plan for tomotherapy dose control system. In the Helical verification plan, when the 3%/3mm criterion is adopted, the larger the target volume, the higher the possibility of lower passing rate, whereas the correlation coefficient between them is relatively low. The high-dose area can be verified by the plans at the center of the phantom or the detection point. With the comprehensive consideration, we suggest putting it at the center of the phantom. \n \n \nKey words: \nEsophageal neoplasm/helical tomotherapy; ArcCHECK system; Plan verification; Quality assurance","PeriodicalId":10288,"journal":{"name":"中华放射肿瘤学杂志","volume":"29 1","pages":"273-277"},"PeriodicalIF":0.0000,"publicationDate":"2020-04-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"中华放射肿瘤学杂志","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3760/CMA.J.CN113030-20181221-00005","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Objective
To explore the application of ArcCheck system in the validation of Helical and Direct tomotherapy plans for esophageal cancer and summarize relevant experience.
Methods
The Helical and Direct tomotherapy verification plans were established for 32 patients with esophageal cancer at different positions according to the doctor′s instructions, which were verified by the ArcCHECK system to compare the passing rate of the results.The correlation between the volume of the target area and the passing rate of the planned verification was analyzed. The therapeutic verification plan with a small target volume was made. The target area was placed at the center of ArcCHECK phantom and the area of detectors to statistically compare the verification passing rates.
Results
Helical plan showed a significantly higher passing rate than the Direct plan (P 0.05).
Conclusions
The passing rate of the Helical plan is generally higher than that of the Direct plan, which may be related to the angular response of the ArcCHECK detector and the fact that more reference points are not included for calculation due to low-dose radiation. In addition, it may also be related to the higher requirements of Direct plan for tomotherapy dose control system. In the Helical verification plan, when the 3%/3mm criterion is adopted, the larger the target volume, the higher the possibility of lower passing rate, whereas the correlation coefficient between them is relatively low. The high-dose area can be verified by the plans at the center of the phantom or the detection point. With the comprehensive consideration, we suggest putting it at the center of the phantom.
Key words:
Esophageal neoplasm/helical tomotherapy; ArcCHECK system; Plan verification; Quality assurance
期刊介绍:
The Chinese Journal of Radiation Oncology is a national academic journal sponsored by the Chinese Medical Association. It was founded in 1992 and the title was written by Chen Minzhang, the former Minister of Health. Its predecessor was the Chinese Journal of Radiation Oncology, which was founded in 1987. The journal is an authoritative journal in the field of radiation oncology in my country. It focuses on clinical tumor radiotherapy, tumor radiation physics, tumor radiation biology, and thermal therapy. Its main readers are middle and senior clinical doctors and scientific researchers. It is now a monthly journal with a large 16-page format and 80 pages of text. For many years, it has adhered to the principle of combining theory with practice and combining improvement with popularization. It now has columns such as monographs, head and neck tumors (monographs), chest tumors (monographs), abdominal tumors (monographs), physics, technology, biology (monographs), reviews, and investigations and research.