Effect of pupil dilation on biometry measurements and intraocular lens power calculation formula with intraocular lens master 500

Suchitra Kumari, Antabha Bandyopadhyay, Taruni Kumari
{"title":"Effect of pupil dilation on biometry measurements and intraocular lens power calculation formula with intraocular lens master 500","authors":"Suchitra Kumari, Antabha Bandyopadhyay, Taruni Kumari","doi":"10.4103/jcor.jcor_43_23","DOIUrl":null,"url":null,"abstract":"Purpose: The purpose was to study the effect of pupil dilation on biometric measurements and intraocular lens (IOL) power calculation using IOLMaster 500. Materials and Methods: A prospective observational study on 52 eyes of 52 cataract patients. The patients were examined with IOL Master 500 before and after pupil dilation. Axial length (AL), corneal power, anterior chamber depth (ACD), and the corresponding IOL power with Haigis and Sanders, Retzlaff, and Kraff (theoretical formula) (SRK/T) formula were compared. Results: There was no statistically significant difference in AL (+0.005 mm; P = .467) and corneal power (+0.001 D; P = .889) before and after dilation. The ACD increased significantly after dilation (+0.12 ±0.35 mm; P < 0.01). IOL power calculation based on SRK/T formula (−0.02 ±0.16, P = 0.20) was not affected, whereas a significant change in the calculation was found using the Haigis formula (+0.07 ±0.54, P = 0.03). Conclusion: There is no clinically significant effect of pupil dilation on the IOL Master 500 measurements of AL, corneal power, and IOL power calculated using the SRK/T formula. However, a significant increase in ACD and IOL power calculation is found using the Haigis Formula.","PeriodicalId":33073,"journal":{"name":"Journal of Clinical Ophthalmology and Research","volume":"11 1","pages":"190 - 193"},"PeriodicalIF":0.0000,"publicationDate":"2023-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Clinical Ophthalmology and Research","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/jcor.jcor_43_23","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: The purpose was to study the effect of pupil dilation on biometric measurements and intraocular lens (IOL) power calculation using IOLMaster 500. Materials and Methods: A prospective observational study on 52 eyes of 52 cataract patients. The patients were examined with IOL Master 500 before and after pupil dilation. Axial length (AL), corneal power, anterior chamber depth (ACD), and the corresponding IOL power with Haigis and Sanders, Retzlaff, and Kraff (theoretical formula) (SRK/T) formula were compared. Results: There was no statistically significant difference in AL (+0.005 mm; P = .467) and corneal power (+0.001 D; P = .889) before and after dilation. The ACD increased significantly after dilation (+0.12 ±0.35 mm; P < 0.01). IOL power calculation based on SRK/T formula (−0.02 ±0.16, P = 0.20) was not affected, whereas a significant change in the calculation was found using the Haigis formula (+0.07 ±0.54, P = 0.03). Conclusion: There is no clinically significant effect of pupil dilation on the IOL Master 500 measurements of AL, corneal power, and IOL power calculated using the SRK/T formula. However, a significant increase in ACD and IOL power calculation is found using the Haigis Formula.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
瞳孔扩张对生物测量和人工晶状体度数计算公式的影响
目的:利用IOLMaster 500研究瞳孔扩张对生物特征测量和人工晶状体度数计算的影响。材料与方法:对52例白内障患者52只眼进行前瞻性观察研究。扩瞳前后用IOL Master 500检查。将眼轴长度(AL)、角膜度数、前房深度(ACD)及相应的IOL度数与Haigis and Sanders、Retzlaff、Kraff(理论公式)(SRK/T)公式进行比较。结果:AL (+0.005 mm;P = .467)和角膜度数(+0.001 D;P = .889)。扩张后ACD显著增高(+0.12±0.35 mm;P < 0.01)。基于SRK/T公式的IOL度数计算不受影响(- 0.02±0.16,P = 0.20),而使用Haigis公式的IOL度数计算有显著变化(+0.07±0.54,P = 0.03)。结论:瞳孔扩张对IOL Master 500的人工晶状体度数、角膜度数和使用SRK/T公式计算的IOL度数无显著影响。然而,使用Haigis公式,发现ACD和IOL功率计算显着增加。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
31
审稿时长
27 weeks
期刊最新文献
Comparative analysis of Dry Eye Questionnaire 5 and Ocular Surface Disease Index in assessing dry eye symptoms Modulation transfer function and higher order aberrations with and without corneo scleral lens in an eye with keratoconus Typical cytomegalovirus retinitis in non-human immunodeficiency virus partially immunosuppressed patients: A case series To study visual evoked potential in patients with optic nerve diseases Corneal morphology following use of Malyugin versus B-hex pupil expansion rings in small pupil phacoemulsification
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1