Suppression of Impedimetric Baseline Drift for Stable Biosensing

Hilena F. Gezahagne, E. L. Brightbill, Decarle S. Jin, Siamalan Krishnathas, B. Brown, M. Mooney, A. O’Riordan, Niamh Creedon, Caoimhe Robinson, E. Vogel
{"title":"Suppression of Impedimetric Baseline Drift for Stable Biosensing","authors":"Hilena F. Gezahagne, E. L. Brightbill, Decarle S. Jin, Siamalan Krishnathas, B. Brown, M. Mooney, A. O’Riordan, Niamh Creedon, Caoimhe Robinson, E. Vogel","doi":"10.1149/2754-2726/ac8fa1","DOIUrl":null,"url":null,"abstract":"Biosensors based on Electrochemical Impedance Spectroscopy (EIS) detect the binding of an analyte to a receptor functionalized electrode by measuring the subsequent change in the extracted charge-transfer resistance (RCT). In this work, the stability of a long chain alkanethiol, 16-mercaptohexadecanoic acid was compared to that of a polymer-based surface linker, ortho-aminobenzoic acid (o-ABA). These two classes of surface linkers were selected due to the marked differences in their structural properties. The drift in RCT observed for the native SAM functionalized gold electrodes was observed to correlate to the drift in the subsequent receptor functionalized SAM. This indicates the importance of the gold-molecule interface for reliable biosensing. Additionally, the magnitude of the baseline drift correlated to the percentage of thiol molecules improperly bound to the gold electrode as evaluated using X-ray Photoelectron Spectroscopy (XPS). Alternatively, the o-ABA functionalized gold electrodes demonstrated negligible drift in the RCT. Furthermore, these polymer functionalized gold electrodes do not require a stabilization period in the buffer solution prior to receptor functionalization. This work emphasizes the importance of understanding and leveraging the structural properties of various classes of surface linkers to ensure the stability of impedimetric measurements.","PeriodicalId":72870,"journal":{"name":"ECS sensors plus","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2022-09-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"ECS sensors plus","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1149/2754-2726/ac8fa1","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

Abstract

Biosensors based on Electrochemical Impedance Spectroscopy (EIS) detect the binding of an analyte to a receptor functionalized electrode by measuring the subsequent change in the extracted charge-transfer resistance (RCT). In this work, the stability of a long chain alkanethiol, 16-mercaptohexadecanoic acid was compared to that of a polymer-based surface linker, ortho-aminobenzoic acid (o-ABA). These two classes of surface linkers were selected due to the marked differences in their structural properties. The drift in RCT observed for the native SAM functionalized gold electrodes was observed to correlate to the drift in the subsequent receptor functionalized SAM. This indicates the importance of the gold-molecule interface for reliable biosensing. Additionally, the magnitude of the baseline drift correlated to the percentage of thiol molecules improperly bound to the gold electrode as evaluated using X-ray Photoelectron Spectroscopy (XPS). Alternatively, the o-ABA functionalized gold electrodes demonstrated negligible drift in the RCT. Furthermore, these polymer functionalized gold electrodes do not require a stabilization period in the buffer solution prior to receptor functionalization. This work emphasizes the importance of understanding and leveraging the structural properties of various classes of surface linkers to ensure the stability of impedimetric measurements.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
抑制阻抗基线漂移用于稳定生物传感
基于电化学阻抗谱(EIS)的生物传感器通过测量随后提取的电荷转移电阻(RCT)的变化来检测分析物与受体功能化电极的结合。在这项工作中,比较了长链烷硫醇16-巯基十六烷酸与聚合物基表面连接剂邻氨基苯甲酸(o-ABA)的稳定性。选择这两类表面连接剂是因为它们的结构性能有明显的差异。在随机对照试验中观察到的原生SAM功能化金电极的漂移与随后受体功能化SAM的漂移相关。这表明了金分子界面对于可靠的生物传感的重要性。此外,基线漂移的大小与使用x射线光电子能谱(XPS)评估的不正确结合到金电极上的硫醇分子的百分比相关。另外,o-ABA功能化的金电极在随机对照试验中显示出可以忽略不计的漂移。此外,这些聚合物功能化金电极在受体功能化之前不需要在缓冲溶液中稳定一段时间。这项工作强调了理解和利用各种表面连接体的结构特性的重要性,以确保阻抗测量的稳定性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Review—Energy and Power Requirements for Wearable Sensors 3D Printed Carbon Nanotubes Reinforced Polydimethylsiloxane Flexible Sensors for Tactile Sensing Editors’ Choice—Review—Advances in Electrochemical Sensors: Improving Food Safety, Quality, and Traceability Field Testing of a Mixed Potential IoT Sensor Platform for Methane Quantification Automated Quantification of DNA Damage Using Deep Learning and Use of Synthetic Data Generated from Basic Geometric Shapes
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1