The role of third-generation dual-source dual-energy computed tomography in characterizing the composition of renal stones with infrared spectroscopy as the reference standard

S. Shalini, V. Arunachalam, R. Varatharajaperumal, P. Mehta, T. S., M. Cherian
{"title":"The role of third-generation dual-source dual-energy computed tomography in characterizing the composition of renal stones with infrared spectroscopy as the reference standard","authors":"S. Shalini, V. Arunachalam, R. Varatharajaperumal, P. Mehta, T. S., M. Cherian","doi":"10.5114/pjr.2022.114841","DOIUrl":null,"url":null,"abstract":"Purpose The aim of our study was to prospectively evaluate the role of third-generation, dual-source, dual-energy computed tomography (DECT) in the characterization of renal calculi, with ex vivo renal stone evaluation using Fourier transform infrared spectroscopy (IS) as the reference standard. Material and methods In our study 50 patients with history suggestive of renal calculi were subjected to DECT using 100 kVp and Sn150 kVp. With DECT, renal stone attenuation at low and high kVp was attained, and the attenuation ratios were measured. The result of DECT was compared with IS to identify the chemical composition of the extracted renal stones. IBM SPSS version 22 was used for statistical analysis. Results In our study, the mean attenuation ratio of the renal stone was 1.57 ± 0.25. Out of 50 patients, the stones of 39 patients were predicted as calcium-containing stones, in 4 patients as cystine stones, and in 7 as uric acid stones on DECT. In IS analysis, 43 patients had calcium-containing stones, and 7 patients had uric acid stones. The accuracy rate of DECT for detecting calcium and uric acid stones in our study were 90% and 100%, respectively. The positive predictive value for the DECT to assess the chemical composition of renal calculi was found to be 92%. Conclusions Third-generation DECT scan had 100% accuracy in differentiating uric acid stones from non-uric acid stones in our study. Because the treatment is different for different chemical compositions of stones, identification of specific chemical components is very important, and it can be accurately done by DECT.","PeriodicalId":94174,"journal":{"name":"Polish journal of radiology","volume":"87 1","pages":"e172 - e176"},"PeriodicalIF":0.0000,"publicationDate":"2022-03-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"2","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Polish journal of radiology","FirstCategoryId":"0","ListUrlMain":"https://doi.org/10.5114/pjr.2022.114841","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 2

Abstract

Purpose The aim of our study was to prospectively evaluate the role of third-generation, dual-source, dual-energy computed tomography (DECT) in the characterization of renal calculi, with ex vivo renal stone evaluation using Fourier transform infrared spectroscopy (IS) as the reference standard. Material and methods In our study 50 patients with history suggestive of renal calculi were subjected to DECT using 100 kVp and Sn150 kVp. With DECT, renal stone attenuation at low and high kVp was attained, and the attenuation ratios were measured. The result of DECT was compared with IS to identify the chemical composition of the extracted renal stones. IBM SPSS version 22 was used for statistical analysis. Results In our study, the mean attenuation ratio of the renal stone was 1.57 ± 0.25. Out of 50 patients, the stones of 39 patients were predicted as calcium-containing stones, in 4 patients as cystine stones, and in 7 as uric acid stones on DECT. In IS analysis, 43 patients had calcium-containing stones, and 7 patients had uric acid stones. The accuracy rate of DECT for detecting calcium and uric acid stones in our study were 90% and 100%, respectively. The positive predictive value for the DECT to assess the chemical composition of renal calculi was found to be 92%. Conclusions Third-generation DECT scan had 100% accuracy in differentiating uric acid stones from non-uric acid stones in our study. Because the treatment is different for different chemical compositions of stones, identification of specific chemical components is very important, and it can be accurately done by DECT.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
第三代双源双能计算机断层扫描在以红外光谱为参考标准表征肾结石成分中的作用
目的本研究的目的是前瞻性地评估第三代双源双能计算机断层扫描(DECT)在肾结石表征中的作用,并以傅里叶变换红外光谱(IS)作为参考标准对离体肾结石进行评估。材料和方法在我们的研究中,50名有肾结石病史的患者接受了100 kVp和Sn150 kVp的DECT。使用DECT,可以获得低kVp和高kVp下肾结石的衰减,并测量衰减率。将DECT的结果与IS进行比较,以确定提取的肾结石的化学成分。使用IBM SPSS 22版进行统计分析。结果肾结石的平均衰减率为1.57±0.25。在50名患者中,39名患者的结石在DECT上被预测为含钙结石,4名患者为胱氨酸结石,7名患者为尿酸结石。在IS分析中,43名患者患有含钙结石,7名患者患有尿酸结石。在我们的研究中,DECT检测钙结石和尿酸结石的准确率分别为90%和100%。DECT评估肾结石化学成分的阳性预测值为92%。结论在我们的研究中,第三代DECT扫描在区分尿酸结石和非尿酸结石方面具有100%的准确性。由于石头的化学成分不同,处理方法不同,因此识别特定的化学成分非常重要,DECT可以准确地进行识别。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
2.10
自引率
0.00%
发文量
0
期刊最新文献
Advancing radiology education for medical students: leveraging digital tools and resources. Application of ultrasound-guided intranodal lymphangiography in the diagnosis and treatment of chylous ascites after abdominal surgery. Medication-induced changes on magnetic resonance imaging of the brain. The peritumoral brain zone in glioblastoma: a review of the pretreatment approach. Comparative efficacy of contrast-enhanced ultrasound versus B-mode ultrasound in the diagnosis and monitoring of hepatic abscesses.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1