Accuracy and consistency of respiratory inductive plethysmography for overnight tidal volume measurement.

J Zhang, E W Ruch, K E Bloch
{"title":"Accuracy and consistency of respiratory inductive plethysmography for overnight tidal volume measurement.","authors":"J Zhang, E W Ruch, K E Bloch","doi":"10.1007/BF02888032","DOIUrl":null,"url":null,"abstract":"<p><p>To validate the accuracy and consistency of respiratory inductive plethysmography (RIP) in measuring tidal volume after an overnight sleep, tidal volumes of 18 patients with suspected sleep-disordered breathing and 8 normal volunteers were measured simultaneously with RIP (VTRIP) and with an ultrasonic airflow meter (VTUFM) before and after an unstrained overnight sleep on supine and lateral decubitus. The bias of the VTRIP was expressed as (VTRIP-VTUFM)/ VTUFM.100%, limits of agreement between VTRIP and VTUFM was measured by averaged bias +/- 2 s. Results showed that in normal subjects, the bias of RIP before and after overnight sleep was precise and consistent in both supine (0.7% and -1.6%) and lateral decubitus (3.7% and -0.56%). In these patients, the bias of RIP before and after sleep in supine also remained small (1.9% and 1.7%), but it became larger in lateral decubitus (24.5% and 20.4%) and 11.5% exceeded the limits of agreement observed in the evening. The patients' body mass indices (BMI) were higher than those of normal subjects (median 34.2 vs. 27.8 kg/m2). Pooled data showed that the bias of VTRIP in the morning on lateral decubitus but not on supine was correlated to BMI (Spearman R = 0.32, n = 52, P = 0.02). Thus, we were led to conclude that the accuracy of VTRIP overnight was precise and consistent in normal subjects, but the deviation of VTRIP measured on lateral decubitus in patients especially in those with excessive obesity was greater, thus, the method should not be used for quantitative determination.</p>","PeriodicalId":73995,"journal":{"name":"Journal of Tongji Medical University = Tong ji yi ke da xue xue bao","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2001-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Tongji Medical University = Tong ji yi ke da xue xue bao","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/BF02888032","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

To validate the accuracy and consistency of respiratory inductive plethysmography (RIP) in measuring tidal volume after an overnight sleep, tidal volumes of 18 patients with suspected sleep-disordered breathing and 8 normal volunteers were measured simultaneously with RIP (VTRIP) and with an ultrasonic airflow meter (VTUFM) before and after an unstrained overnight sleep on supine and lateral decubitus. The bias of the VTRIP was expressed as (VTRIP-VTUFM)/ VTUFM.100%, limits of agreement between VTRIP and VTUFM was measured by averaged bias +/- 2 s. Results showed that in normal subjects, the bias of RIP before and after overnight sleep was precise and consistent in both supine (0.7% and -1.6%) and lateral decubitus (3.7% and -0.56%). In these patients, the bias of RIP before and after sleep in supine also remained small (1.9% and 1.7%), but it became larger in lateral decubitus (24.5% and 20.4%) and 11.5% exceeded the limits of agreement observed in the evening. The patients' body mass indices (BMI) were higher than those of normal subjects (median 34.2 vs. 27.8 kg/m2). Pooled data showed that the bias of VTRIP in the morning on lateral decubitus but not on supine was correlated to BMI (Spearman R = 0.32, n = 52, P = 0.02). Thus, we were led to conclude that the accuracy of VTRIP overnight was precise and consistent in normal subjects, but the deviation of VTRIP measured on lateral decubitus in patients especially in those with excessive obesity was greater, thus, the method should not be used for quantitative determination.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
呼吸诱导式胸透法测量夜间潮气量的准确性和一致性。
为了验证呼吸感应式胸压计(RIP)在测量一夜睡眠后潮气量时的准确性和一致性,在仰卧和侧卧位一夜睡眠前后,用 RIP(VTRIP)和超声气流计(VTUFM)同时测量了 18 名疑似睡眠呼吸障碍患者和 8 名正常志愿者的潮气量。VTRIP 的偏差表示为 (VTRIP-VTUFM)/ VTUFM.100%,VTRIP 和 VTUFM 之间的一致性限制按平均偏差 +/- 2 秒来测量。结果显示,正常人在一夜睡眠前后,仰卧位(0.7% 和-1.6%)和侧卧位(3.7% 和-0.56%)的 RIP 偏差都是精确和一致的。在这些患者中,仰卧位睡眠前后的 RIP 偏差也保持较小(1.9% 和 1.7%),但侧卧位的偏差变大(24.5% 和 20.4%),11.5% 的偏差超过了晚间观察到的一致性极限。患者的体重指数(BMI)高于正常人(中位数分别为 34.2 和 27.8 kg/m2)。汇总数据显示,早晨侧卧位时 VTRIP 的偏差与体重指数相关,而仰卧位时则不相关(Spearman R = 0.32,n = 52,P = 0.02)。因此,我们得出结论,正常人夜间 VTRIP 的准确性是精确和一致的,但患者尤其是过度肥胖者在侧卧位上测量的 VTRIP 偏差较大,因此该方法不应被用于定量测定。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
A comparative study of the influences of pulmonary arterial pressure and pulmonary capillary wedge pressure on impedance rheopneumogram in patients with three different diseases Construction and identification of recombinant adenovirus vector containing the hVEGF165 gene Effects of prostaglandins and leukotrienes on hypoxic pulmonary vasoconstriction in rats Preliminary study on the relationship between cAMP level and gsp expression in cultured human pituitary somatotrophinomas Variation of perioperative blood cTnT levels in patients undergoing cardiopulmonary bypass and its clinical implication
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1