Blind spot in ultrasound central corneal thickness measurement – Central corneal thickness of apex versus central corneal thickness of vertex

P. Ramesh, S. Parthasarathi, Abhinay Ashok, R. John
{"title":"Blind spot in ultrasound central corneal thickness measurement – Central corneal thickness of apex versus central corneal thickness of vertex","authors":"P. Ramesh, S. Parthasarathi, Abhinay Ashok, R. John","doi":"10.4103/kjo.kjo_42_21","DOIUrl":null,"url":null,"abstract":"This study aimed to compare central corneal thickness (CCT) using ultrasound pachymetry (USP) (CCT apex) and specular microscopy (CCT vertex) and also to find out the intra-reading variability from the two instruments. A prospective, observational study was conducted in a tertiary eye care center in southern India. This study was conducted on 12 patients (24 eyes, 96 data set eyes, 768 data points in total) aged 20–50, in non-pathological corneas. Eight CCT measurements by specular microscopy and eight measurements by USP were taken by two different experienced observers on day 1, followed by repeating the same on day 2. The readings were averaged and compared by paired t-test. SPSS software was used for statistical analysis. Variability between the eight readings produced by the same instrument was calculated and coefficient of variation was plotted. Inter-examiner variability and intra-examiner variability of the two modalities were studied, and it was considered statistically significant if P < 0.05. The mean CCT by USP was 522.71 µm and the mean CCT by specular microscopy was 519.43 µm. The coefficient of variation of the eight readings varied significantly between the machines. On an average, the coefficient of variation was 0.4% compared to 0.8% in specular microscopy and USP, respectively. The intra-reading variability of ultrasound CCT is twice as that compared to specular microscopy. The clinical relevance of this spread is relevant and important in modern-day practice. Hence, it is better advisable that CCT readings from both the modalities, not be used interchangeably.","PeriodicalId":32483,"journal":{"name":"Kerala Journal of Ophthalmology","volume":"1 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2023-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Kerala Journal of Ophthalmology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/kjo.kjo_42_21","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

This study aimed to compare central corneal thickness (CCT) using ultrasound pachymetry (USP) (CCT apex) and specular microscopy (CCT vertex) and also to find out the intra-reading variability from the two instruments. A prospective, observational study was conducted in a tertiary eye care center in southern India. This study was conducted on 12 patients (24 eyes, 96 data set eyes, 768 data points in total) aged 20–50, in non-pathological corneas. Eight CCT measurements by specular microscopy and eight measurements by USP were taken by two different experienced observers on day 1, followed by repeating the same on day 2. The readings were averaged and compared by paired t-test. SPSS software was used for statistical analysis. Variability between the eight readings produced by the same instrument was calculated and coefficient of variation was plotted. Inter-examiner variability and intra-examiner variability of the two modalities were studied, and it was considered statistically significant if P < 0.05. The mean CCT by USP was 522.71 µm and the mean CCT by specular microscopy was 519.43 µm. The coefficient of variation of the eight readings varied significantly between the machines. On an average, the coefficient of variation was 0.4% compared to 0.8% in specular microscopy and USP, respectively. The intra-reading variability of ultrasound CCT is twice as that compared to specular microscopy. The clinical relevance of this spread is relevant and important in modern-day practice. Hence, it is better advisable that CCT readings from both the modalities, not be used interchangeably.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
超声角膜中央厚度测量中的盲点——顶点角膜中央厚度与顶点角膜中央厚度的对比
本研究旨在比较超声测厚仪(USP) (CCT顶点)和镜面显微镜(CCT顶点)的中央角膜厚度(CCT),并找出两种仪器的读数内变异性。在印度南部的一家三级眼科护理中心进行了一项前瞻性观察研究。本研究纳入12例20-50岁非病理性角膜患者(24只眼,96只数据集眼,共768个数据点)。第1天由两名不同的经验丰富的观察者通过镜面显微镜和USP分别进行8次CCT测量,然后在第2天重复相同的操作。读数取平均值,并用配对t检验进行比较。采用SPSS软件进行统计分析。计算同一仪器产生的8个读数之间的变异性,并绘制变异系数。对两种检查方式的检查者间变异性和检查者内变异性进行研究,P < 0.05认为差异有统计学意义。USP法平均CCT为522.71µm,镜下平均CCT为519.43µm。这8个读数的变异系数在不同的机器之间差别很大。平均而言,变异系数为0.4%,而镜面显微镜和USP分别为0.8%。超声CCT的读数内变异性是镜面显微镜的两倍。这种传播的临床相关性在现代实践中具有相关性和重要性。因此,最好建议两种方式的CCT读数不要互换使用。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
24
审稿时长
13 weeks
期刊最新文献
Aicardi syndrome Review of “A randomized trial evaluating efficacy of overminus lenses combined with prism in the children with intermittent exotropia” Bilateral proptosis in Marshall-Smith syndrome Clinical features, diagnosis, and treatment outcome of Rhino-orbito-cerebral mucormycosis: A 2-year experience at a tertiary care center A study to assess meibomian gland morphology using meibography in posterior blepharitis patients and normal population: A hospital-based observational study
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1