Towards the Reliability Check of Complementary Methods for the Precise Estimation of Peak Efficiency of a Long Run Used Germanium Detector in Low Activity Environmental Measurements
{"title":"Towards the Reliability Check of Complementary Methods for the Precise Estimation of Peak Efficiency of a Long Run Used Germanium Detector in Low Activity Environmental Measurements","authors":"","doi":"10.33140/jeee.01.01.03","DOIUrl":null,"url":null,"abstract":"The gamma-ray spectrometry using long run used hyper pure germanium (HPGe) detector needs a stringent full energy peak (FEP) efficiency calibration for environmental radioactivity measurements. The FEP efficiency calibration was carried out using the complementary methods like analytical estimations and Monte Carlo (MC) simulations. In our work, for the analytical estimations, the relative deviation of the theoretical results with that of experimental FEP efficiency was found to be less than 6%. However, in the case of MC methodologies, an overall estimated standard deviation of 5% could be achievable only with the optimized geometry of the long run used germanium detector for low-activity environmental measurements.","PeriodicalId":39047,"journal":{"name":"Journal of Electrical and Electronics Engineering","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2022-08-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Electrical and Electronics Engineering","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.33140/jeee.01.01.03","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Engineering","Score":null,"Total":0}
引用次数: 1
Abstract
The gamma-ray spectrometry using long run used hyper pure germanium (HPGe) detector needs a stringent full energy peak (FEP) efficiency calibration for environmental radioactivity measurements. The FEP efficiency calibration was carried out using the complementary methods like analytical estimations and Monte Carlo (MC) simulations. In our work, for the analytical estimations, the relative deviation of the theoretical results with that of experimental FEP efficiency was found to be less than 6%. However, in the case of MC methodologies, an overall estimated standard deviation of 5% could be achievable only with the optimized geometry of the long run used germanium detector for low-activity environmental measurements.
期刊介绍:
Journal of Electrical and Electronics Engineering is a scientific interdisciplinary, application-oriented publication that offer to the researchers and to the PhD students the possibility to disseminate their novel and original scientific and research contributions in the field of electrical and electronics engineering. The articles are reviewed by professionals and the selection of the papers is based only on the quality of their content and following the next criteria: the papers presents the research results of the authors, the papers / the content of the papers have not been submitted or published elsewhere, the paper must be written in English, as well as the fact that the papers should include in the reference list papers already published in recent years in the Journal of Electrical and Electronics Engineering that present similar research results. The topics and instructions for authors of this journal can be found to the appropiate sections.