Enhanced K-Edge Radiography Using a High-Spatial-Resolution Cadmium Telluride Array Detector

Manabu Watanabe, E. Sato, Yasuyuki Oda, Hodaka Moriyama, Osahiko Hagiwara, H. Matsukiyo, T. Enomoto, S. Kusachi
{"title":"Enhanced K-Edge Radiography Using a High-Spatial-Resolution Cadmium Telluride Array Detector","authors":"Manabu Watanabe, E. Sato, Yasuyuki Oda, Hodaka Moriyama, Osahiko Hagiwara, H. Matsukiyo, T. Enomoto, S. Kusachi","doi":"10.4236/ijmpcero.2018.73024","DOIUrl":null,"url":null,"abstract":"To confirm the imaging effect of a dual-energy (DE) cadmium telluride (CdTe) array detector (XCounter, Actaeon) and to perform fundamental studies on DE computed tomography, we performed enhanced K-edge radiography using iodine (I) and gadolinium (Gd) media. DE radiography was performed using an X-ray generator with a 0.1-mm-diam-focus tube and a 0.5-mm-thick beryllium window, a 1.0-mm-thick aluminum filter for absorbing extremely low-energy photons, and the CdTe array detector with pixel dimensions of 0.1 × 0.1 mm2. Each pixel has a charge-sensitive amplifier and a dual-energy counter, and the event pulses from the amplifier are sent to the counter to determine two threshold energies. The tube current was a maximum value of 0.50 mA, and the tube voltages for I- and Gd-K-edge radiograms were 60 and 80 kV, respectively. In the I-K-edge radiography of a dog-heart phantom at an energy range of 33 - 60 keV, the muscle density increased, and fine coronary arteries were visible. Utilizing Gd-K-edge radiography of a rabbit head phantom at an energy range of 50 - 80 keV, the muscle density increased, and fine blood vessels in the nose were observed at high contrasts. Using the DE array detector, we confirmed the image-contrast variations with changes in the threshold energy.","PeriodicalId":14028,"journal":{"name":"International Journal of Medical Physics, Clinical Engineering and Radiation Oncology","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2018-07-18","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"4","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Medical Physics, Clinical Engineering and Radiation Oncology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4236/ijmpcero.2018.73024","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 4

Abstract

To confirm the imaging effect of a dual-energy (DE) cadmium telluride (CdTe) array detector (XCounter, Actaeon) and to perform fundamental studies on DE computed tomography, we performed enhanced K-edge radiography using iodine (I) and gadolinium (Gd) media. DE radiography was performed using an X-ray generator with a 0.1-mm-diam-focus tube and a 0.5-mm-thick beryllium window, a 1.0-mm-thick aluminum filter for absorbing extremely low-energy photons, and the CdTe array detector with pixel dimensions of 0.1 × 0.1 mm2. Each pixel has a charge-sensitive amplifier and a dual-energy counter, and the event pulses from the amplifier are sent to the counter to determine two threshold energies. The tube current was a maximum value of 0.50 mA, and the tube voltages for I- and Gd-K-edge radiograms were 60 and 80 kV, respectively. In the I-K-edge radiography of a dog-heart phantom at an energy range of 33 - 60 keV, the muscle density increased, and fine coronary arteries were visible. Utilizing Gd-K-edge radiography of a rabbit head phantom at an energy range of 50 - 80 keV, the muscle density increased, and fine blood vessels in the nose were observed at high contrasts. Using the DE array detector, we confirmed the image-contrast variations with changes in the threshold energy.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
使用高空间分辨率碲化镉阵列探测器的增强k边缘射线照相
为了确认双能(DE)碲化镉(CdTe)阵列探测器(XCounter, Actaeon)的成像效果,并对DE计算机断层扫描进行基础研究,我们使用碘(I)和钆(Gd)介质进行了增强的k边缘射线照相。采用直径为0.1 mm的聚焦管、0.5 mm厚的铍窗、1.0 mm厚的铝滤光片和像素尺寸为0.1 × 0.1 mm2的CdTe阵列探测器进行DE射线照相。每个像素具有电荷敏感放大器和双能量计数器,来自放大器的事件脉冲被发送到计数器以确定两个阈值能量。管电流最大值为0.50 mA, I边和gd - k边射线图的管电压分别为60和80 kV。在33 ~ 60kev能量范围内,狗心幻像的i - k边缘x线片显示肌肉密度增加,可见细小冠状动脉。利用50 ~ 80 keV能量范围的gd - k边缘x线摄影,兔头幻肢肌肉密度增加,高对比度下观察到鼻内细小血管。利用DE阵列探测器,我们证实了图像对比度随阈值能量变化的变化。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Assessment of Monitor Units and Gamma Pass Rate for 6 MV and Flattening Filter Free (FFF) Beams in Volumetric Modulated Arc Therapy (VMAT) The Journey of SRS: An Initial Experience of Starting Sterotactic Radiosurgery Facility at Purbanchal Cancer Hospital—First in Nepal Hereditary Multiple Exostoses (HME) with Peroneal Nerve Compresion: A Case Report Dosimetric Study of the Effect of Calypso-Compatible Couch Top for Spine Stereotactic Body Radiation Therapy Radiation-Associated Cardiotoxicity during Breast Cancer Treatment with Ionizing Radiation
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1