{"title":"Improving Experimental Design through Uncertainty Analysis","authors":"Ian M. Hobbs, J. Charboneau, Todd L. Jacobsen","doi":"10.3390/metrology3030014","DOIUrl":null,"url":null,"abstract":"In this paper, the development of a fission-gas collecting and physical-analysis-enabling instrument was proposed for small-volume determination. Analysis specifications require a design capable of accurately and repeatably determining volumes in the range of 0.07–2.5 mL. This system relies on a series of gas expansions originating from a cylinder with known internal volume. The combined gas law is used to derive the unknown volumes from these expansions. Initial system designs included one of two known volumes, 11.85 ± 0.34 mL and 5.807 ± 0.078 mL, with a manifold volume of 32 mL. Results obtained from modeling this system’s operation showed that 0.07 mL can be determined with a relative expanded uncertainty greater than 300% (k = 2) for a single replicate, which was unacceptable for the proposed experimental design. Initial modeling showed that the volume connecting the known volume and rodlet, i.e., the manifold volume, and the sensitivity of the pressure sensor were key contributors to the expanded uncertainty of the measured rodlet volume. The system’s design limited the available options for pressure sensors, so emphasis was placed on the design of the manifold volume. The final system design reduced the manifold volume to 17 mL. These changes in design, combined with replicate analysis, were able to reduce the relative expanded uncertainty by ±12% (k = 2) for the 0.07 mL volume.","PeriodicalId":100666,"journal":{"name":"Industrial Metrology","volume":"11 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2023-06-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Industrial Metrology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3390/metrology3030014","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

In this paper, the development of a fission-gas collecting and physical-analysis-enabling instrument was proposed for small-volume determination. Analysis specifications require a design capable of accurately and repeatably determining volumes in the range of 0.07–2.5 mL. This system relies on a series of gas expansions originating from a cylinder with known internal volume. The combined gas law is used to derive the unknown volumes from these expansions. Initial system designs included one of two known volumes, 11.85 ± 0.34 mL and 5.807 ± 0.078 mL, with a manifold volume of 32 mL. Results obtained from modeling this system’s operation showed that 0.07 mL can be determined with a relative expanded uncertainty greater than 300% (k = 2) for a single replicate, which was unacceptable for the proposed experimental design. Initial modeling showed that the volume connecting the known volume and rodlet, i.e., the manifold volume, and the sensitivity of the pressure sensor were key contributors to the expanded uncertainty of the measured rodlet volume. The system’s design limited the available options for pressure sensors, so emphasis was placed on the design of the manifold volume. The final system design reduced the manifold volume to 17 mL. These changes in design, combined with replicate analysis, were able to reduce the relative expanded uncertainty by ±12% (k = 2) for the 0.07 mL volume.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
通过不确定性分析改进实验设计
本文提出了一种用于小体积测定的裂变气体收集和物理分析仪器的开发。分析规范要求设计能够准确和重复地测定0.07-2.5 mL范围内的体积。该系统依赖于一系列气体膨胀,源自已知内部体积的圆柱体。结合气体定律用于从这些膨胀中推导出未知体积。最初的系统设计包括两种已知体积中的一种,11.85±0.34 mL和5.807±0.078 mL,流形体积为32 mL。通过模拟该系统的操作获得的结果表明,对于单个重复,0.07 mL的相对扩展不确定度大于300% (k = 2),这对于拟议的实验设计是不可接受的。初始建模表明,连接已知体积和小棒的体积(即流形体积)和压力传感器的灵敏度是影响测量小棒体积扩展不确定度的关键因素。该系统的设计限制了压力传感器的可用选择,因此重点放在了歧管体积的设计上。最终的系统设计将歧管体积减少到17 mL。这些设计上的变化,结合重复分析,能够将0.07 mL体积的相对扩展不确定度降低±12% (k = 2)。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Predictive Modeling of Photovoltaic Panel Power Production through On-Site Environmental and Electrical Measurements Using Artificial Neural Networks A Two-Dimensional K-Shell X-ray Fluorescence (2D-KXRF) Model for Soft Tissue Attenuation Corrections of Strontium Measurements in a Cortical Lamb Bone Sample Report of the CCU/CCQM Workshop on “The Metrology of Quantities Which Can Be Counted” Time and Its Measure: Historical and Social Implications Updated Strategy and Scope of Metrology
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1