Ocena testów i wytycznych zawartych w Rozporządzeniu Ministra Zdrowia z dnia 18 lutego 2011 roku dotyczących kontroli jakości tomografu komputerowego

Piotr M. Knapp , Bartosz Pawałowski , Agata Jodda
{"title":"Ocena testów i wytycznych zawartych w Rozporządzeniu Ministra Zdrowia z dnia 18 lutego 2011 roku dotyczących kontroli jakości tomografu komputerowego","authors":"Piotr M. Knapp ,&nbsp;Bartosz Pawałowski ,&nbsp;Agata Jodda","doi":"10.1016/j.onko.2015.03.001","DOIUrl":null,"url":null,"abstract":"<div><h3>Aim</h3><p>Evaluation of tests and guidelines provided by the regulation of the Minister of Health on 18 February 2011 relating to quality assurance in computed tomography.</p></div><div><h3>Material and method</h3><p>Catphan 600 phantom, 2-part PMMA CT-Phantom IBA phantom and Raysafe Xi modules were used to carry out quality assurance tests on Siemens Sensation Open tomograph. For evaluation of the results Microsoft Excel and Varian Eclipse software were used. The range of the assay with use of Catphan phantom was extended for better assessment of tests provided in regulations.</p></div><div><h3>Results</h3><p>Performed tests revealed that according to the regulations, CT is working properly. On the other hand, it is easily observed that results vary for different scan settings in tests performed on Catphan phantom.</p></div><div><h3>Conclusions</h3><p>Performed tests showed that conditions given in regulations are easy to fulfil. Another conclusion is that there are no strict guidelines about scan settings in by the regulation of the Minister of Health on 18 February 2011. In order to maintain constant conditions of testing, the settings from standard workflow should be used.</p></div>","PeriodicalId":101295,"journal":{"name":"Zeszyty Naukowe WCO, Letters in Oncology Science","volume":"12 1","pages":"Pages 1-9"},"PeriodicalIF":0.0000,"publicationDate":"2015-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1016/j.onko.2015.03.001","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Zeszyty Naukowe WCO, Letters in Oncology Science","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1734048915000073","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Aim

Evaluation of tests and guidelines provided by the regulation of the Minister of Health on 18 February 2011 relating to quality assurance in computed tomography.

Material and method

Catphan 600 phantom, 2-part PMMA CT-Phantom IBA phantom and Raysafe Xi modules were used to carry out quality assurance tests on Siemens Sensation Open tomograph. For evaluation of the results Microsoft Excel and Varian Eclipse software were used. The range of the assay with use of Catphan phantom was extended for better assessment of tests provided in regulations.

Results

Performed tests revealed that according to the regulations, CT is working properly. On the other hand, it is easily observed that results vary for different scan settings in tests performed on Catphan phantom.

Conclusions

Performed tests showed that conditions given in regulations are easy to fulfil. Another conclusion is that there are no strict guidelines about scan settings in by the regulation of the Minister of Health on 18 February 2011. In order to maintain constant conditions of testing, the settings from standard workflow should be used.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
评估2011年2月18日卫生部长关于计算机断层扫描质量保证的条例提供的测试和准则。采用Microsoft Excel和Varian Eclipse软件对结果进行评价。为了更好地评估法规中规定的试验,扩大了使用Catphan模体的测定范围。结果经检测,CT按规定工作正常。另一方面,很容易观察到,在对Catphan幻影执行的测试中,不同的扫描设置会产生不同的结果。结论经实验证明,本规程规定的条件易于满足。另一个结论是,2011年2月18日卫生部长的条例中没有关于扫描设置的严格指导方针。为了保持测试的恒定条件,应该使用标准工作流中的设置。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Editorial Board Radioterapia w leczeniu raka sromu Co nowego w leczeniu napromienianiem raka stercza oraz raka pęcherza moczowego na podstawie doniesień z Konferencji Amerykańskiego Towarzystwa Radioterapii Onkologicznej (ASTRO), San Francisco 2014? Rola SIB-IMRT w radioterapii przerzutów do mózgu – doniesienie konferencyjne z ESTRO Forum 2015 Sprawozdanie z pobytu na 56. konferencji ASTRO
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1