Analysis of an Anesthesia and Intensive Care Ventilators during Suctioning through a Fiber-optic Bronchoscope

P. Kremeier
{"title":"Analysis of an Anesthesia and Intensive Care Ventilators during Suctioning through a Fiber-optic Bronchoscope","authors":"P. Kremeier","doi":"10.24966/ACC-8879/100032","DOIUrl":null,"url":null,"abstract":"Background: The aim of this bench test was to compare the performance of an anesthesia and an intensive care ventilator during Fiber-optic Bronchoscopy (FOB) suctioning. Methods: In a lung model, a 5.9 mm OD FOB was inserted into an endotracheal tube through a tight-fitting connector. Pressure and flow sensors were placed at the airways opening (Pao) and distal to the bronchoscope at the trachea (Ptr). This model was tested under pressure and volume controlled mandatory ventilation and during Continuous Positive Airway Pressure (CPAP) delivered by both, anesthesia and intensive care ventilators. End-expiratory pressure was set at 10 cm H2O. Measurements were performed during suctioning with a mean negative flow of -12 ± 0.2 L/min at the tip of the FOB. Results: The anesthesia ventilator failed to keep end-expiratory pressure above zero at the trachea during FOB suctioning using both, continuous mandatory ventilation and CPAP. When using an intensive care ventilator, Ptr decreased below zero during FOB suctioning during volume controlled ventilation but remained slightly above zero under pressure controlled ventilation. When this ventilator delivered 10 cm H2O of CPAP, Pao remained constant while Ptr decreased by 2.5 cm H2O during suctioning. Conclusion: During FOB suctioning, targeted Ptr could neither be maintained anesthesia nor intensive care ventilators used in mandatory modes of ventilation. Only CPAP administered by the intensive care ventilator kept Ptr reasonably stable during suctioning.","PeriodicalId":73662,"journal":{"name":"Journal of clinical anesthesia and intensive care","volume":"1 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2019-05-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of clinical anesthesia and intensive care","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.24966/ACC-8879/100032","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Background: The aim of this bench test was to compare the performance of an anesthesia and an intensive care ventilator during Fiber-optic Bronchoscopy (FOB) suctioning. Methods: In a lung model, a 5.9 mm OD FOB was inserted into an endotracheal tube through a tight-fitting connector. Pressure and flow sensors were placed at the airways opening (Pao) and distal to the bronchoscope at the trachea (Ptr). This model was tested under pressure and volume controlled mandatory ventilation and during Continuous Positive Airway Pressure (CPAP) delivered by both, anesthesia and intensive care ventilators. End-expiratory pressure was set at 10 cm H2O. Measurements were performed during suctioning with a mean negative flow of -12 ± 0.2 L/min at the tip of the FOB. Results: The anesthesia ventilator failed to keep end-expiratory pressure above zero at the trachea during FOB suctioning using both, continuous mandatory ventilation and CPAP. When using an intensive care ventilator, Ptr decreased below zero during FOB suctioning during volume controlled ventilation but remained slightly above zero under pressure controlled ventilation. When this ventilator delivered 10 cm H2O of CPAP, Pao remained constant while Ptr decreased by 2.5 cm H2O during suctioning. Conclusion: During FOB suctioning, targeted Ptr could neither be maintained anesthesia nor intensive care ventilators used in mandatory modes of ventilation. Only CPAP administered by the intensive care ventilator kept Ptr reasonably stable during suctioning.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
纤维支气管镜下麻醉和重症监护呼吸机吸痰的分析
背景:本台架试验的目的是比较麻醉和重症监护呼吸机在光纤支气管镜(FOB)吸入过程中的性能。方法:在肺模型中,将外径5.9 mm的FOB通过紧密接头插入气管内管。压力和流量传感器放置在气道开口(Pao)和气管支气管镜远端(Ptr)。该模型在压力和容量控制的强制通气和麻醉和重症监护呼吸机持续气道正压通气(CPAP)期间进行了测试。呼气末压设定为10 cm H2O。在抽吸过程中进行测量,平均负流量为-12±0.2 L/min。结果:在连续强制通气和CPAP两种方法下,麻醉呼吸机在FOB吸痰时均未能使气管呼气末压保持在零以上。当使用重症监护呼吸机时,在容积控制通气时,FOB吸痰时Ptr降至零以下,而在压力控制通气时,Ptr略高于零。当该呼吸机输送10 cm H2O的CPAP时,吸痰时Pao保持不变,而Ptr下降2.5 cm H2O。结论:在FOB吸痰过程中,靶向Ptr既不能维持麻醉,也不能使用强制通气模式下的重症监护呼吸机。只有重症监护呼吸机给予CPAP才能使吸入过程中Ptr保持合理稳定。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
The role of hybrid arch repair with thoracic endovascular repair for contemporary management of complex arch and descending thoracoabdominal aortopathies Commentary on ‘Late presentation of right coronary artery stent infection as left empyema and pleuro-pericardial fistula’ Case Report of a Contained Ascending Aortic Rupture Mind-body approaches for reducing the need for post-operative opioids: Evidence and opportunities Mind-body approaches for reducing the need for post-operative opioids: Evidence and opportunities.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1